Investigation of a matrix sentence test in noise: Reproducibility and discrimination function in cochlear implant patients

被引:59
作者
Hey, Matthias [1 ]
Hocke, Thomas [2 ]
Hedderich, Juergen [3 ]
Mueller-Deile, Joachim [1 ]
机构
[1] KG Hanover, Cochlear Deutschland GmbH & Co, Hannover, Germany
[2] KG Hanover, Cochlear Deutschland GmbH & Co, Hannover, Germany
[3] Christian Albrechts Univ Kiel, Inst Med Informat & Stat, Kiel, Germany
关键词
Speech intelligibility test; speech reception threshold; cochlear implant; matrix sentence test; OLSA; Oldenburg sentence test; test-retest reproducibility; intelligibility in noise; SPEECH RECEPTION THRESHOLD; INTELLIGIBILITY; STIMULATION; HEARING; SYSTEM;
D O I
10.3109/14992027.2014.938368
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: The aim of this study was to describe common properties (reproducibility, discrimination function, and its steepness) of matrix tests used for cochlear implant (CI) users and to obtain data for the German-language version matrix test, the Oldenburg sentence test (OLSA), presented in noise. Design: The speech reception thresholds (SRT) in noise were measured by means of an adaptive test procedure, and by measurement at various signal-to-noise ratios to determine the course of the entire discrimination function per subject. Study sample: The measurements were performed on 38 CI users fitted with a Cochlear (TM) Freedom (R) or a Cochlear (TM) Nucleus (R) 5 CI system. Results: The test-retest reproducibility showed a significant dependence on the SRT in noise. For the better performers, the test-retest difference was found to be smaller, while for the poorer performers the difference increased. For the better performers, the slope of the discrimination function at SRT (s50) was comparable to that for individuals with normal hearing, while for the poorer performers the s50 tended to be significantly reduced. Conclusions: As the CI users differed significantly in their SRT and their s50, a unified discrimination function for CI users must not be used. Further tailoring of the procedure may be required, especially for poorer CI performers.
引用
收藏
页码:895 / 902
页数:8
相关论文
共 29 条
  • [1] Spectrum of Hearing Disorders and Their Management in Children With CHARGE Syndrome
    Arndt, Susan
    Laszig, Roland
    Beck, Rainer
    Schild, Christian
    Maier, Wolfgang
    Birkenhaeger, Ralf
    Kroeger, Stefanie
    Wesarg, Thomas
    Aschendorff, Antje
    [J]. OTOLOGY & NEUROTOLOGY, 2010, 31 (01) : 67 - 73
  • [2] Statistical methods for assessing agreement between two methods of clinical measurement
    Bland, J. Martin
    Altman, Douglas G.
    [J]. INTERNATIONAL JOURNAL OF NURSING STUDIES, 2010, 47 (08) : 931 - 936
  • [3] Efficient adaptive procedures for threshold and concurrent slope estimates for psychophysics and speech intelligibility tests
    Brand, T
    Kollmeier, B
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2002, 111 (06) : 2801 - 2810
  • [4] EFFICIENT ADAPTIVE METHODS FOR MEASURING SPEECH RECEPTION THRESHOLD IN QUIET AND IN NOISE
    HAGERMAN, B
    KINNEFORS, C
    [J]. SCANDINAVIAN AUDIOLOGY, 1995, 24 (01): : 71 - 77
  • [5] Hagerman B, 1984, Scand Audiol Suppl, V21, P1
  • [6] SENTENCES FOR TESTING SPEECH-INTELLIGIBILITY IN NOISE
    HAGERMAN, B
    [J]. SCANDINAVIAN AUDIOLOGY, 1982, 11 (02): : 79 - 87
  • [7] MEASUREMENT OF SPEECH RECEPTION THRESHOLD - A COMPARISON BETWEEN 2 METHODS
    HAGERMAN, B
    [J]. SCANDINAVIAN AUDIOLOGY, 1982, 11 (03): : 191 - 193
  • [8] HAHLBROCK K H, 1953, Arch Ohren Nasen Kehlkopfheilkd, V162, P394, DOI 10.1007/BF02105664
  • [9] Hanschmann H, 2010, HNO, V58, P674, DOI 10.1007/s00106-010-2086-x
  • [10] Hedderich J., 2011, ANGEW STAT, P98