Direct charge sharing observation in single-photon-counting pixel detector

被引:17
作者
Pellegrini, G. [1 ]
Maiorino, M.
Blanchot, G.
Chmeissani, M.
Garcia, J.
Lozano, M.
Martinez, R.
Puigdengoles, C.
Ullan, M.
机构
[1] Ctr Nacl Microelect, CSIC, IMB CNM, E-08193 Barcelona, Spain
[2] Univ Autonoma Barcelona, IFAE, E-08193 Bellaterra, Spain
关键词
pixel detectors; medical imaging; charge sharing;
D O I
10.1016/j.nima.2006.11.010
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In photon-counting imaging devices, charge sharing can limit the detector spatial resolution and contrast, as multiple counts can be induced in adjacent pixels as a result of the spread of the charge cloud generated from a single X-ray photon of high energy in the detector bulk. Although debated for a long time, the full impact of charge sharing has not been completely assessed. In this work, the importance of charge sharing in pixellated CdTe and silicon detectors is studied by exposing imaging devices to different low activity sources. These devices are made of Si and CdTe pixel detector bump-bonded to Medipix2 single-photon-counting chips with a 55 mu m pixel pitch. We will show how charge sharing affects the spatial detector resolution depending on incident particle type (alpha, beta and gamma), detector bias voltage and read-out chip threshold. This study will give an insight on the impact on the design and operation of pixel detectors coupled to photon-counting devices for imaging applications. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:137 / 140
页数:4
相关论文
共 4 条
  • [1] CHMEISSANI M, P 21 IEEE INSTR MEAS, V1, P787
  • [2] LLOPART X, 2001, P IEEE NUCL SCI S M, V3, P1484
  • [3] Newman, 1991, ELECTROCHEMICAL SYST
  • [4] Performance limits of a 55-μm pixel CdTe detector
    Pellegrini, G
    Chmeissani, M
    Maiorino, M
    Blanchot, G
    Garcia, J
    Lozano, M
    Martinez, R
    Puigdengoles, C
    Ullan, M
    Casado, P
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 361 - 366