A novel tetrachlorothallate (III)-PVC membrane sensor for the potentiometric determination of thallium (III)

被引:53
作者
Hassanien, MM
Abou-El-Sherbini, KS
Mostafa, GAE [1 ]
机构
[1] Natl Res Ctr, Appl Organ Chem Dept, Microanalyt Lab, Cairo, Egypt
[2] Ind Educ Coll, Bani Suwayf, Egypt
[3] Natl Res Ctr, Dept Inorgan Chem, Cairo, Egypt
关键词
PVC membrane electrode; tetrachlorothalate (III); 2,3,5-triphenyl-2-H-tetrazolium;
D O I
10.1016/S0039-9140(02)00530-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel tetrachlorothallate (III) (TCT)-selective membrane sensor consisting of tetrachlorothallate (III)-2,3,5-triphenyl-2-H-tetrazolium ion pair dispersed in a PVC matrix plasticized with dioctylphthatate is described. The electrode shows a stable, near-Nernstian response for 1 x 10(-3) -4 x 10(-6) M thallium (III) at 25degreesC with an anionic slope of 56.5 +/- 0.5 over the pH range 3-6. The lower detection limit and the response time are 2 x 10(-6) M and 30-60 s, respectively. Selectivity coefficients for Tl(III) relative to a number of interfering substances were investigated. There is negligible interference from many cations and anions; however, iodide and bromide are significantly interfere. The determination of 0.5-200 pg ml(-1) of Tl(III) in aqueous solutions shows an average recovery of 99.0% and a mean relative standard deviation of 1.4% at 50.0 mug ml(-1). The direct determination of Tl(III) in spiked wastewater gave results that compare favorably with those obtained by the atomic absorption spectrometric method. The electrode was successfully applied for the determination of thallium in zinc concentrate. Also the tetrachlorothallate electrode has been utilized as an end point indicator electrode for the determination of thallium using potentiometric titration. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:383 / 392
页数:10
相关论文
共 25 条
[1]   PVC membrane ion selective electrode for the determination of pentachlorophenol in water, wood and soil using tetrazolium pentachlorophenolate [J].
Abbas, MN ;
Mostafa, GAE ;
Homoda, AMA .
TALANTA, 2001, 55 (03) :647-656
[2]  
ABRUTIS AA, 1987, ZH ANAL KHIM, V42, P287
[3]  
ASMAI T, 1996, FRESEN J ANAL CHEM, V356, P348
[4]   RECOMMENDATIONS FOR NOMENCLATURE OF ION-SELECTIVE ELECTRODES - (IUPAC RECOMMENDATIONS 1994) [J].
BUCK, RP ;
LINDNER, E .
PURE AND APPLIED CHEMISTRY, 1994, 66 (12) :2527-2536
[5]  
Cai Q., 1995, ELECTROANAL, V7, P1040
[6]   DETERMINATION OF TRACE THALLIUM AFTER ACCUMULATION OF THALLIUM(III) AT A 8-HYDROXYQUINOLINE-MODIFIED CARBON-PASTE ELECTRODE [J].
CAI, QT ;
KHOO, SB .
ANALYST, 1995, 120 (04) :1047-1053
[7]   A PVC-COATED CARBON ROD ION-SELECTIVE ELECTRODE FOR THALLIUM AND ITS APPLICATION TO THE ANALYSIS OF ROCKS AND MINERALS [J].
DAN, DH ;
DONG, Y .
TALANTA, 1988, 35 (07) :589-590
[8]  
DAVIES JEW, 1973, LAB PRACT, V22, P20
[9]   VOLTAMMETRIC BEHAVIOR OF THALLIUM(III) ON CARBON PASTE ELECTRODES CHEMICALLY MODIFIED WITH AN ANION-EXCHANGER [J].
DIEWALD, W ;
KALCHER, K ;
NEUHOLD, C ;
CAI, X ;
MAGEE, RJ .
ANALYTICA CHIMICA ACTA, 1993, 273 (1-2) :237-244
[10]   VOLTAMMETRIC BEHAVIOR OF THALLIUM(III) AT A SOLID HETEROGENEOUS CARBON ELECTRODE USING ION-PAIR FORMATION [J].
DIEWALD, W ;
KALCHER, K ;
NEUHOLD, C ;
SVANCARA, I ;
CAI, XH .
ANALYST, 1994, 119 (02) :299-304