Techniques for pneumocephalus and brain shift reduction in DBS surgery: a review of the literature

被引:17
作者
Beggio, Giacomo [1 ]
Raneri, Fabio [1 ]
Rustemi, Oriela [1 ]
Scerrati, Alba [1 ]
Zambon, Giampaolo [1 ]
Piacentino, Massimo [1 ]
机构
[1] San Bortolo Hosp AULSS 8 Berica, Dept Neurosurg, Viale Rodolfi 37, I-36100 Vicenza, VI, Italy
关键词
Brain shift; Pneumocephalus; Deep brain stimulation; Parkinson's disease; SUBTHALAMIC NUCLEUS; STIMULATION SURGERY; INTRACRANIAL AIR; IMPLANTATION; POSITION; ERROR;
D O I
10.1007/s10143-019-01220-2
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Deep brain stimulation has become an established therapeutic choice to manage the symptoms of medically refractory Parkinson's disease. Its efficacy is highly dependent on the accuracy of electrodes' positioning in the correct anatomical target. During DBS procedure, the opening of the dura mater induces the displacement of neural structures. This effect mainly depends on the loss of the physiological negative intracranial pressure, air inflow, and loss of cerebrospinal fluid. Several studies concentrated on correcting surgical techniques for DBS electrodes' positioning in order to reduce pneumocephalus which may result in therapeutic failure. The authors focused in particular on reducing the brain air window and maintaining the pressure gradient between intra- and extracranial compartments. A significant reduction of pneumocephalus and brain shift was obtained by excluding the opening of the subarachnoid space, by covering the dura mater opening with tissue sealant and by reducing the intracranial pressure in general anesthesia. Smaller burr hole diameters were not statistically relevant for reducing air inflow and displacement of anatomical targets. The review of the literature showed that conserving a physiological intra-extracranial pressure gradient plays a fundamental role in avoiding pneumocephalus and consequent displacement of brain structures, which improves surgical accuracy and DBS long-term results.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 25 条
[1]   Intracranial Air Correlates with Preoperative Cerebral Atrophy and Stereotactic Error during Bilateral STN DBS Surgery for Parkinson's Disease [J].
Azmi, Hooman ;
Machado, Andre ;
Deogaonkar, Milind ;
Rezai, Ali .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2011, 89 (04) :246-252
[2]   Influence of Intracranial Air on Electrode Position and Clinical Outcomes following Deep Brain Stimulation for Parkinson's Disease [J].
Bentley, J. Nicole ;
Guan, Zhe ;
Cummings, Karen S. ;
Chou, Kelvin L. ;
Patil, Parag G. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2017, 95 (01) :6-12
[3]   Impact of brain shift on subcallosal cingulate deep brain stimulation [J].
Choi, Ki Sueng ;
Noecker, Angela M. ;
Riva-Posse, Patricio ;
Rajendra, Justin K. ;
Gross, Robert E. ;
Mayberg, Helen S. ;
McIntyre, Cameron C. .
BRAIN STIMULATION, 2018, 11 (02) :445-453
[4]   Minimizing Brain Shift during Functional Neurosurgical Procedures - A Simple Burr Hole Technique that can Decrease CSF Loss and Intracranial Air [J].
Coenen, V. A. ;
Abdel-Rahman, A. ;
McMaster, J. ;
Bogod, N. ;
Honey, C. R. .
CENTRAL EUROPEAN NEUROSURGERY, 2011, 72 (04) :181-185
[5]   Simultaneous bilateral stereotactic procedure for deep brain stimulation implants: a significant step for reducing operation time [J].
Fonoff, Erich Talamoni ;
Azevedo, Angelo ;
dos Angelos, Jairo Silva ;
Ruiz Martinez, Raquel Chacon ;
Navarro, Jessie ;
Reis, Paul Rodrigo ;
San Martin Sepulveda, Miguel Ernesto ;
Cury, Rubens Gisbert ;
dos Santos Ghilardi, Maria Gabriela ;
Teixeira, Manoel Jacobsen ;
Contreras Lopez, William Omar .
JOURNAL OF NEUROSURGERY, 2016, 125 (01) :85-89
[6]  
Hakim S, 1976, Surg Neurol, V5, P187
[7]   Asleep Deep Brain Stimulation Reduces Incidence of Intracranial Air during Electrode Implantation [J].
Ko, Andrew L. ;
Magown, Philippe ;
Ozpinar, Alp ;
Hamzaoglu, Vural ;
Burchiel, Kim J. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2018, 96 (02) :83-90
[8]   Five-year follow-up of bilateral stimulation of the subthalamic nucleus in advanced Parkinson's disease [J].
Krack, P ;
Batir, A ;
Van Blercom, N ;
Chabardes, S ;
Fraix, V ;
Ardouin, C ;
Koudsie, A ;
Limousin, PD ;
Benazzouz, A ;
LeBas, JF ;
Benabid, AL ;
Pollak, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (20) :1925-1934
[9]   Review on Factors Affecting Targeting Accuracy of Deep Brain Stimulation Electrode Implantation between 2001 and 2015 [J].
Li, Zhe ;
Zhang, Jian-Guo ;
Ye, Yan ;
Li, Xiaoping .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2016, 94 (06) :351-362
[10]   Electrical stimulation of the subthalamic nucleus in advanced Parkinson's disease [J].
Limousin, P ;
Krack, P ;
Pollak, P ;
Benazzouz, A ;
Ardouin, C ;
Hoffmann, D ;
Benabid, AL .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (16) :1105-1111