PATIENT DOSIMETRY DURING INTERVENTIONAL CARDIAC PROCEDURES IN A DEDICATED CATHETERIZATION LABORATORY

被引:15
|
作者
Uniyal, Satish C. [1 ]
Chaturvedi, Vineet [1 ]
Sharma, Sunil D. [2 ]
Rawat, Anurag [3 ]
机构
[1] Swami Rama Himalayan Univ, Himalayan Inst Med Sci, Dept Radiol, Dehra Dun 248016, Uttar Pradesh, India
[2] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Bombay 400094, Maharashtra, India
[3] Swami Rama Himalayan Univ, Himalayan Inst Med Sci, Dept Cardiol, Dehra Dun 248016, Uttar Pradesh, India
关键词
TRANSLUMINAL CORONARY ANGIOPLASTY; DOSE AREA PRODUCT; CARDIOLOGY PROCEDURES; SKIN INJURIES; VALUES; EXPOSURE; BELGIUM;
D O I
10.1093/rpd/ncw286
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Cardiac interventions often result in high radiation dose to patient's skin, so a reliable indicator in terms of a commonly used dose descriptor is required to monitor skin exposures. In the present study, Gafchromic XR-RV3 film was used to measure the peak skin dose (PSD) during 40 coronary angiography (CA) and 50 percutaneous transluminal coronary angioplasty (PTCA) procedures. Corresponding values of kerma-area product (P-KA), fluoroscopy time (FT) and reference air-kerma (K-a,K-r) were recorded and correlated with PSD. Doses to patient's eyes and thyroid were also measured by using thermoluminescent dosimeters (TLDs) during PTCA procedures. The average dose to thyroid was about six times higher than the average dose to eyes. The mean values of PSD, P-KA and FT were 1140 mGy, 97 Gy cm(2) and 15.7 min for PTCA and 290 mGy, 21.1 Gy cm(2) and 2.4 min for CA procedures, respectively. One in seven patients of PTCA procedure received PSD > 2 Gy. With respect to FT, P-KA may be used as a better predictor of skin exposures because the correlation of PSD with P-KA was found better than with FT for both CA and PTCA procedures.
引用
收藏
页码:201 / 208
页数:8
相关论文
共 50 条
  • [41] RADIATION DOSES TO ANAESTHETISTS DURING PROSTATIC ARTERY EMBOLIZATION INTERVENTIONAL PROCEDURES
    Garzon, W. J.
    Khoury, H. J.
    RADIATION PROTECTION DOSIMETRY, 2019, 185 (02) : 196 - 200
  • [42] Particulate production during orthodontic production laboratory procedures
    Ireland, Anthony J.
    Wilson, Andrew A.
    Blythe, Luke
    Johnston, Nicola J.
    Price, Robert
    Sandy, Jonathan R.
    JOURNAL OF EXPOSURE SCIENCE AND ENVIRONMENTAL EPIDEMIOLOGY, 2011, 21 (05) : 536 - 540
  • [43] Patient dosimetry survey of pediatric diagnostic and therapeutic cardiac catheterisation in Japan
    Ishibashi, Toru
    Takei, Yasutaka
    Kato, Mamoru
    Yamashita, Yukari
    Tsukamoto, Atsuko
    Matsumoto, Kazuma
    Sakamoto, Hajime
    Masuda, Takanori
    Miyazaki, Osamu
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (10) : 1082 - 1089
  • [44] Radiation Dose Monitoring for Fluoroscopically Guided Interventional Procedures: Effect on Patient Radiation Exposure
    Liu, Bob
    Hirsch, Joshua A.
    Li, Xinhua
    Sheridan, Robert M.
    Rehani, Madan M.
    Zheng, Hui
    Rabinov, James D.
    RADIOLOGY, 2019, 290 (03) : 744 - 749
  • [45] Fluoroscopy Pulse Rate Reduction during Diagnostic and Therapeutic Imaging in the Cardiac Catheterization Laboratory: An Evaluation of Radiation Dose, Procedure Complications and Outcomes
    Hansen, James W.
    Foy, Andrew
    Schmidt, Torrey
    Ghahramani, Mehrdad
    Chambers, Charles E.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 89 (04) : 665 - 670
  • [46] Reduction of Exposure of Patients and Staff to Radiation During Fluoroscopically Guided Interventional Procedures
    Vano E.
    Current Radiology Reports, 2013, 1 (1) : 11 - 22
  • [47] Coaching Reduced the Radiation Dose of Pain Physicians by Half during Interventional Procedures
    Slegers, A. S.
    Gultuna, I.
    Aukes, J. A.
    van Gorp, E. J. J. A. A.
    Blommers, F. M. N.
    Niehof, S. P.
    Bosman, J.
    PAIN PRACTICE, 2015, 15 (05) : 400 - 406
  • [48] Patient Radiation Dose in Diagnostic and Interventional. Procedures for IntracraniaL Aneurysms: Experience at a Single Center
    Chun, Chang Woo
    Kim, Bum-Soo
    Lee, Cheol Hyoun
    Ihn, Yon Kwon
    Shin, Yong-Sam
    KOREAN JOURNAL OF RADIOLOGY, 2014, 15 (06) : 844 - 849
  • [49] Evaluation of patients' and occupational radiation risk dose during conventional and interventional radiology procedures
    Tamam, Nissren
    Salah, H.
    Almogren, Kholoud S.
    Mahgoub, Omer
    Saeed, Mohammed Khalil
    Abdullah, Yousef
    Tai, Duong Thanh
    Omer, Hiba
    Sulieman, Abdelmoneim
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 207
  • [50] Patient Body Mass Index and Physician Radiation Dose During Coronary Angiography Is the Obesity Epidemic Impacting the Occupational Risk of Physicians in the Catheterization Laboratory?
    Madder, Ryan D.
    VanOosterhout, Stacie
    Mulder, Abbey
    Ten Brock, Taylor
    Clarey, Austin T.
    Parker, Jessica L.
    Jacoby, Mark E.
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2019, 12 (01)