A magnet built on bronchoscopic suction for extraction of tracheobronchial headscarf pins: a novel technique and review of a tertiary centre experience

被引:8
作者
Elsayed, Hany H. [1 ]
Mostafa, Ahmed M. [1 ]
Soliman, Saleh [1 ]
El-Bawab, Hatem Y. [1 ]
Moharram, Adel A. [2 ]
El-Nori, Ahmed A. [1 ]
机构
[1] Ain Shams Univ, Dept Thorac Surg, Cairo 11566, Egypt
[2] Cairo Univ, Dept Med Engn, Cairo, Egypt
关键词
Magnets; Airway pins; FOREIGN-BODIES; REMOVAL; BODY;
D O I
10.1093/icvts/ivw006
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: Airway metal pins are one of the most commonly inhaled foreign bodies in Eastern societies in young females wearing headscarves. We innovated a modified bronchoscopic technique to extract tracheobronchial headscarf pins by the insertion of a magnet to allow an easy and non-traumatic extraction of the pins. The aim of this study was to assess the feasibility and safety of our new technique and compare it with our large previous experience with the classic bronchoscopic method of extraction of tracheobronchial headscarf pins. METHODS: We performed a study comparing our retrospective experience of classic bronchoscopic extraction from February 2004 to January 2014 and prospective experience with our modified technique using the magnet from January 2014 to June 2015. An institutional review board and new device approval were obtained. RESULTS: Three hundred and twenty-six procedures on 315 patients were performed during our initial 10-year experience. Of them, 304 patients were females. The median age of our group was 13 (0-62). The median time from inhalation to procedure was 1 day (0-1022). After introducing our modified new technique using the magnet, 20 procedures were performed. Nineteen were females. The median time of the procedure and the need to forcefully bend the pin for extraction were in favour of the new technique in comparison with our classic approach (2 vs 6 min; P < 0.001) (2 patients = 20% vs 192 = 58%; P < 0.001). The conversion rate to surgery was also in favour of the modified technique but did not reach statistical significance (0 = 0% vs 15 = 4.8%; P = 0.32). All patients who underwent the modified technique were discharged home on the same day of the procedure. No procedural complications were recorded. All remain well on a follow-up period of up to 14 months. CONCLUSIONS: Bronchoscopic extraction of tracheobronchial inhaled headscarf pins using a novel technique using homemade magnets was safer and simpler in comparison with our large experience with the classic approach. We advise the use of this device (or concept) in selected patients in centres dealing with this problem.
引用
收藏
页码:531 / 536
页数:6
相关论文
共 13 条
[1]  
Al-Sarraf Nael, 2009, Interact Cardiovasc Thorac Surg, V9, P187, DOI 10.1510/icvts.2009.207548
[2]   Tracheobronchial foreign bodies - Presentation and management in children and adults [J].
Baharloo, F ;
Veyckemans, F ;
Francis, C ;
Biettlot, MP ;
Rodenstein, DO .
CHEST, 1999, 115 (05) :1357-1362
[3]   REMOVAL OF METALLIC FOREIGN BODY BY MAGNETIC FORCE [J].
BROD, RC ;
SCHLANG, HA .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1962, 179 (02) :164-&
[4]   Surgery of intra-bronchic foreign bodies. [J].
Caidi, M ;
Kabiri, H ;
Lazrek, I ;
El Maslout, A ;
Ben Osman, A .
ANNALES DE CHIRURGIE, 2002, 127 (06) :456-460
[5]   Scarf pin inhalation: clinical characteristics and surgical treatment [J].
Fenane, Hicham ;
Bouchikh, Mohammed ;
Bouti, Khalid ;
El Maidi, Mehdi ;
Ouchen, Fahd ;
Mbola, Tchely-Oaly ;
Damessane, Lamboni ;
Achir, Abdellah ;
Benosman, Abdellatif .
JOURNAL OF CARDIOTHORACIC SURGERY, 2015, 10
[6]   Magnetic resonance imaging of oscillating electrical currents [J].
Halpern-Manners, Nicholas W. ;
Bajaj, Vikram S. ;
Teisseyre, Thomas Z. ;
Pines, Alexander .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2010, 107 (19) :8519-8524
[7]   Removal of nasal foreign bodies in the pediatric population [J].
Kadish, HA ;
Corneli, HM .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1997, 15 (01) :54-56
[8]   The heterodox nature of "Turban Pins" in foreign body aspiration; the central anatolian experience [J].
Kaptanoglu, Melih ;
Nadir, Aydin ;
Dogan, Kasim ;
Sahin, Ekber .
INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2007, 71 (04) :553-558
[9]  
Luo Yiwen, 2003, Yan Ke Xue Bao, V19, P142
[10]   Sharp foreign bodies in the tracheobronchial tree [J].
Murthy, PSN ;
Ingle, VS ;
George, E ;
Ramakrishna, S ;
Shah, FA .
AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2001, 22 (02) :154-156