Recent developments in electrolytic devices for ion chromatography

被引:54
作者
Liu, Y [1 ]
Srinivasan, K [1 ]
Pohl, C [1 ]
Avdalovic, N [1 ]
机构
[1] Dionex Chem Corp, Sunnyvale, CA 94086 USA
来源
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS | 2004年 / 60卷 / 03期
关键词
ion chromatography; on-line electrolytic eluent generation; electrolytic suppressors; continuously regenerated trap column; electrolysis of water; electrically polarized ion-exchange media;
D O I
10.1016/j.jbbm.2004.01.004
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Recent developments in new electrolytic devices that utilize the electrolysis of water and charge-selective electromigration of ions through ion-exchange media have significantly changed the routine operation of ion chromatographic methods. Examples of these new electrolytic devices include online eluent generators that produce high-purity electrolyte eluents using deionized water as the carrier stream, continuously regenerated trap columns that remove ionic contaminants in the eluents, and continuously regenerated suppressors that reduce eluent background conductance prior to conductivity detection. The combined use of these electrolytic devices has made it possible to per-form various ion chromatographic separations using only deionized water as the mobile phase. This paper reviews the operation principles of these electrolytic devices and their applications in the ion chromatographic determination of anionic and cationic analytes. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:205 / 232
页数:28
相关论文
共 57 条
[1]  
BAN T, 1983, Patent No. 4403039
[2]   2-DIMENSIONAL CONDUCTOMETRIC DETECTION IN ION CHROMATOGRAPHY - SEQUENTIAL SUPPRESSED AND SINGLE COLUMN DETECTION [J].
BERGLUND, I ;
DASGUPTA, PK ;
LOPEZ, JL ;
NARA, O .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1192-1198
[3]   Detection techniques in ion analysis: what are our choices? [J].
Buchberger, WW .
JOURNAL OF CHROMATOGRAPHY A, 2000, 884 (1-2) :3-22
[4]  
*DION CORP, 2002, 53 DION CORP
[5]   New suppressor technology improves trace level anion analysis with carbonate-hydrogencarbonate mobile phases [J].
Douglas, DR ;
Saari-Nordhaus, R ;
Despres, P ;
Anderson, JM .
JOURNAL OF CHROMATOGRAPHY A, 2002, 956 (1-2) :47-51
[6]  
HADDAD PR, 1990, J CHROMATOGRAPHY LIB, V46
[7]  
HENSALL A, 1992, JAM LAB NOV, pR20
[8]  
JANSEN KH, 1984, Patent No. 4459357
[9]  
KINDERMAN M, 2003, AM LAB FEB, P26
[10]   Determination of trace-level anions in high-purity water samples by ion chromatography with an automated on-line fluent generation system [J].
Liu, Y ;
Kaiser, E ;
Avdalovic, N .
MICROCHEMICAL JOURNAL, 1999, 62 (01) :164-173