Kinetic modelling using basis functions derived from two-tissue compartmental models with a plasma input function: General principle and application to [18F] fluorodeoxyglucose positron emission tomography

被引:36
作者
Hong, Young T. [1 ]
Fryer, Tim D. [1 ]
机构
[1] Univ Cambridge, Addenbrookes Hosp, Wolfson Brain Imaging Ctr, Dept Clin Neurosci, Cambridge CB2 2QQ, England
基金
英国医学研究理事会;
关键词
CEREBRAL GLUCOSE-UTILIZATION; EXPONENTIAL SPECTRAL-ANALYSIS; BLOOD-FLOW; METABOLIC-RATE; HETEROGENEOUS TISSUES; PARAMETRIC IMAGES; RIDGE-REGRESSION; DYNAMIC PET; BRAIN; DEOXYGLUCOSE;
D O I
10.1016/j.neuroimage.2010.02.013
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
A kinetic modelling method for the determination of influx constant, K-i is given that utilises basis functions derived from plasma input two-tissue compartmental models (BAFPIC). Two forms of the basis functions are given: BAFPIC(1) with k(4)=0 (no product loss) and BAFPIC(R) with k(4) non-zero. Simulations were performed using literature rate constant values for [F-18]fluorodeoxyglucose (FDG) in both normal and abnormal brain pathology. Both homogeneous and heterogeneous tissues were simulated and this data was also used as input for other methods commonly used to determine K-i: non-linear least squares compartmental modelling (NLLS), autoradiographic method and Patlak-Gjedde graphical analysis (PGA). The four methods were also compared for real FDG positron emission tomography (PET) data. For both k(4)=0 and k(4) non-zero simulated data BAFPIC had the best bias properties of the four methods. The autoradiographic method was always the best for variability but BAFPIC(1) had lower variability than PGA and NLLS. For non-zero k(4) data, the variance of BAFPIC(R) was inferior to PGA but still significantly superior to NLLS. K-i maps calculated from real data substantiate the simulation results, with BAFPIC(1) having lower noise than PGA. Voxel K-i values from BAFPIC(1) correlated well with those from PGA (r(2) = 0.989). BAFPIC is easy to implement and combines low bias with good noise properties for voxel-wise determination of K-i for FDG. BAFPIC is suitable for determining K-i for other tracers well characterised by a serial two-tissue compartment model and has the advantage of also producing values for individual kinetic constants and blood volume. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:164 / 172
页数:9
相关论文
共 41 条
[31]   MODEL OF KINETIC-BEHAVIOR OF DEOXYGLUCOSE IN HETEROGENEOUS TISSUES IN BRAIN - A REINTERPRETATION OF THE SIGNIFICANCE OF PARAMETERS FITTED TO HOMOGENEOUS TISSUE MODELS [J].
SCHMIDT, K ;
MIES, G ;
SOKOLOFF, L .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1991, 11 (01) :10-24
[32]   ERRORS INTRODUCED BY TISSUE HETEROGENEITY IN ESTIMATION OF LOCAL CEREBRAL GLUCOSE-UTILIZATION WITH CURRENT KINETIC-MODELS OF THE [F-18] FLUORODEOXYGLUCOSE METHOD [J].
SCHMIDT, K ;
LUCIGNANI, G ;
MORESCO, RM ;
RIZZO, G ;
GILARDI, MC ;
MESSA, C ;
COLOMBO, F ;
FAZIO, F ;
SOKOLOFF, L .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1992, 12 (05) :823-834
[33]   [DEOXYGLUCOSE-C-14 METHOD FOR MEASUREMENT OF LOCAL CEREBRAL GLUCOSE-UTILIZATION - THEORY, PROCEDURE, AND NORMAL VALUES IN CONSCIOUS AND ANESTHETIZED ALBINO-RAT [J].
SOKOLOFF, L ;
REIVICH, M ;
KENNEDY, C ;
DESROSIERS, MH ;
PATLAK, CS ;
PETTIGREW, KD ;
SAKURADA, O ;
SHINOHARA, M .
JOURNAL OF NEUROCHEMISTRY, 1977, 28 (05) :897-916
[34]   THE USE OF SPECTRAL-ANALYSIS TO DETERMINE REGIONAL CEREBRAL GLUCOSE-UTILIZATION WITH POSITRON EMISSION TOMOGRAPHY AND [F-18] FLUORODEOXYGLUCOSE - THEORY, IMPLEMENTATION, AND OPTIMIZATION PROCEDURES [J].
TURKHEIMER, F ;
MORESCO, RM ;
LUCIGNANI, G ;
SOKOLOFF, L ;
FAZIO, F ;
SCHMIDT, K .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1994, 14 (03) :406-422
[35]   Estimation of component and parameter distributions in spectral analysis [J].
Turkheimer, F ;
Sokoloff, L ;
Bertoldo, A ;
Lucignani, G ;
Reivich, M ;
Jaggi, JL ;
Schmidt, K .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1998, 18 (11) :1211-1222
[36]   Multi-resolution Bayesian regression in PET dynamic studies using wavelets [J].
Turkheimer, F. E. ;
Aston, J. A. D. ;
Asselin, M. -C. ;
Hinz, R. .
NEUROIMAGE, 2006, 32 (01) :111-121
[37]   Rank-shaping regularization of exponential spectral analysis for application to functional parametric mapping [J].
Turkheimer, FE ;
Hinz, R ;
Gunn, RN ;
Aston, JAD ;
Gunn, SR ;
Cunningham, VJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (23) :3819-3841
[38]  
Watabe H, 2005, J NUCL MED, V46, P1219
[39]  
WEINHARD K, 1985, J CEREB BLOOD FLOW M, V5, P115
[40]   Noise reduction in the simplified reference tissue model for neuroreceptor functional Imaging [J].
Wu, YJ ;
Carson, RE .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2002, 22 (12) :1440-1452