Factors influencing the accuracy of renal output efficiency

被引:12
作者
Piepsz, A [1 ]
Ham, H [1 ]
机构
[1] CHU St Pierre, Dept Radioisotopes, B-1000 Brussels, Belgium
关键词
D O I
10.1097/00006231-200011000-00004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Output efficiency (OE) has been proposed as an objective tool for the estimation of renal emptying. The aim of this paper was to evaluate some factors that may affect the value of this parameter obtained by a Tc-99(m) MAG3 renogram. Tn 22 patients, the effect of different types of renal background correction was calculated. In 10 patients, simultaneous gamma camera renography and multiple blood sampling allowed the error of using a heart curve instead of a plasma curve to be evaluated. Finally, an evaluation was carried out on the effect of neglecting the change of plasma activity between the end of the furosemide acquisition and the post-micturition view obtained after repositioning of the patient on the gamma camera: 142 patients who had had multiple blood sampling were retrospectively selected for that purpose. It has been shown that all these factors may significantly affect the value of OE, particularly in cases with poor renal drainage. Unless standardization of the procedure is introduced for some of these factors, the cut-off levels for good, fair and poor drainage may vary between centres. ((C) 2000 Lippincott Williams & Wilkins).
引用
收藏
页码:1009 / 1013
页数:5
相关论文
共 10 条
[1]   Assessment of drainage in PUJ dilatation: Pelvic excretion efficiency as an index of renal function [J].
Anderson, PJ ;
Rangarajan, V ;
Gordon, I .
NUCLEAR MEDICINE COMMUNICATIONS, 1997, 18 (09) :823-826
[2]  
BRITTON KE, 1971, CLIN RENOGRAPHY, P163
[3]  
CHAIWATANARAT T, 1993, J NUCL MED, V34, P845
[4]  
CONWAY J, 1992, J NUCL MED, V33, P2047
[5]  
KURUC A, 1987, J NUCL MED, V28, P1338
[6]   NORA: A simple and reliable parameter for estimating renal output with or without frusemide challenge [J].
Piepsz, A ;
Tondeur, M ;
Ham, H .
NUCLEAR MEDICINE COMMUNICATIONS, 2000, 21 (04) :317-323
[7]  
PIEPSZ A, 1990, J NUCL MED, V31, P430
[8]   Consensus report on quality control of quantitative measurements of renal function obtained from the renogram: International Consensus Committee from the Scientific Committee of Radionuclides in Nephrourology [J].
Prigent, A ;
Cosgriff, P ;
Gates, GF ;
Granerus, G ;
Fine, EJ ;
Itoh, K ;
Peters, M ;
Piepsz, A ;
Rehling, M ;
Rutland, M ;
Taylor, A .
SEMINARS IN NUCLEAR MEDICINE, 1999, 29 (02) :146-159
[9]  
Saunders CAB, 1997, J NUCL MED, V38, P1483
[10]   TECHNETIUM 99M MERCAPTOACETYLTRIGLYCINE GAMMA-CAMERA CLEARANCE CALCULATIONS - METHODOLOGICAL PROBLEMS [J].
TONDEUR, M ;
PIEPSZ, A ;
DOBBELEIR, A ;
HAM, H .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1991, 18 (02) :83-86