Safety and Efficacy of the Nit-Occlud® Coil for Percutaneous Closure of Various Sizes of PDA

被引:2
作者
Jung, Seyong [1 ]
Seol, Jaehee [2 ]
Choi, Jaeyoung [1 ]
Ha, Keesoo [3 ]
机构
[1] Yonsei Univ, Coll Med, Div Pediat Cardiol, Seoul 03722, South Korea
[2] Yonsei Univ, Dept Pediat, Wonju Coll Med, Wonju 26426, South Korea
[3] Korea Univ, Dept Pediat, Coll Med, Seoul 02841, South Korea
关键词
patent ductus arteriosus; therapeutic occlusion; procedure; PATENT DUCTUS-ARTERIOSUS; TRANSCATHETER OCCLUSION; DEVICE CLOSURE; ADULT;
D O I
10.3390/jcm11092469
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Most interventionalists use the Amplatzer Duct Occluder (ADO) or the Nit-Occlud (R) Coils (NOC) to close patent ductus arteriosus (PDA). Data regarding the success and effect of NOCs in the occlusion of large PDAs are insufficient. We aimed to investigate whether the PDA occlusion of large PDAs using NOC is safe and efficient for all ages. This was a retrospective study involving 361 pediatric and adult patients who underwent the transcatheter closure of PDA using NOC over the past 21 years for all PDA sizes and ages. The sizes of PDA were classified as small, moderate, and large. A comparison of the aortic pressure before and after PDA occlusion using NOC showed significant differences in terms of systolic and pulse pressures for all age groups (p < 0.05). The rate of the residual shunts of NOC was 2%, while the rate of complete occlusions of NOC was 98% at 12 months after occlusion regardless of the shape of PDA. The complication rate with PDA occlusion using NOC was 5%. PDA occlusion using NOC is as effective and safe as ADO for the occlusion of PDA of all sizes. Therefore, PDA occlusion using NOC can be a safe and feasible procedure to close various sizes and types of PDA without complications.
引用
收藏
页数:11
相关论文
共 18 条
[1]  
Ahmadi A, 2014, ARYA ATHEROSCLER, V10, P37
[2]   Percutaneous device closure of patent ductus arteriosus in adult patients with 10-year follow-up [J].
Alkashkari, Wail ;
Albugami, Saad ;
Alrahimi, Jamilah ;
Althobaiti, Mohammed ;
Kinsara, Abdulhalim ;
Abousa, Ahmed ;
Krimly, Ahmed ;
Alzahrani, Atif ;
Niazi, Akram ;
Aburemish, Hayam .
HEART VIEWS, 2019, 20 (04) :139-145
[3]   Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder [J].
Bilici, Meki ;
Demir, Fikri ;
Akin, Alper ;
Ture, Mehmet ;
Balik, Hasan ;
Kuyumcu, Mahir .
PEDIATRIC CARDIOLOGY, 2017, 38 (08) :1598-1605
[4]   Closure of persistently patent arterial duct and its impact on cerebral circulatory haemodynamics in children [J].
Bissonnette, B ;
Benson, LN .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1998, 45 (03) :199-205
[5]   Hypertension in Children and Adolescents: A Position Statement From a Panel of Multidisciplinary Experts Coordinated by the French Society of Hypertension [J].
Bouhanick, Beatrice ;
Sosner, Philippe ;
Brochard, Karine ;
Mounier-Vehier, Claire ;
Plu-Bureau, Genevieve ;
Hascoet, Sebastien ;
Ranchin, Bruno ;
Pietrement, Christine ;
Martinerie, Laetitia ;
Boivin, Jean Marc ;
Fauvel, Jean Pierre ;
Bacchetta, Justine .
FRONTIERS IN PEDIATRICS, 2021, 9
[6]   Transcatheter closure of patent ductus arteriosus with Nit-Occlud coils [J].
Celiker, A ;
Aypar, E ;
Karagöz, T ;
Dilber, E ;
Ceviz, N .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 65 (04) :569-576
[7]   Patent arterial duct [J].
Forsey, Jonathan T. ;
Elmasry, Ola A. ;
Martin, Robin P. .
ORPHANET JOURNAL OF RARE DISEASES, 2009, 4
[8]  
García-Montes José Antonio, 2006, Arch. Cardiol. Méx., V76, P163
[9]   Trans-Catheter Closure of Patent Ductus Arteriosus-What Is the Best Device? [J].
Ghasemi, Abdolrahim ;
Pandya, Swati ;
Reddy, Surendranath Veeram ;
Turner, Daniel R. ;
Du, Wei ;
Navabi, Mohamad Ali ;
Mirzaaghayan, Mohammad Reza ;
Kiani, Abdolrazagh ;
Sloan, Katherine ;
Forbes, Thomas J. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (05) :687-695
[10]   Effects of coil closure of patent ductus arteriosus on left anterior descending coronary artery blood flow using transthoracic Doppler echocardiography [J].
Harada, K ;
Toyono, M ;
Tamura, M .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2004, 17 (06) :659-663