Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs

被引:50
作者
Cooper, Edward S. [1 ]
Muir, William W. [1 ]
机构
[1] Ohio State Univ, Dept Vet Clin Sci, Coll Vet Med, Columbus, OH 43210 USA
关键词
pulse-contour analysis; arterial pressure waveform; noninvasive; beat-to-beat; lithium dilution; hemorrhagic shock;
D O I
10.1097/01.CCM.0000266590.25109.F2
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To determine agreement and correlation between cardiac output determined by arterial pressure waveform analysis (PulseCO) and the lithium dilution indicator technique (LiDCO) during severe hemorrhagic shock and after fluid resuscitation in dogs. Design: Prospective experimental study. Setting: University research laboratory. Subjects: Twelve adult mongrel dogs. Interventions: Dogs were anesthetized, and selected arteries and veins were catheterized. Baseline cardiac output was determined by LiDCO and used to calibrate the PulseCO. Hemorrhagic shock was induced by withdrawing blood to achieve and maintain a mean arterial pressure of 30-40 mm Hg for 60 mins, and cardiac output was measured again using both methods. All dogs were resuscitated by administering lactated Ringer's solution intravenously to achieve and maintain a mean arterial pressure between 60 and 70 mm Hg. PulseCO and LiDCO values were measured at 10 and 120 mins after resuscitation. Measurements and Main Results: Mean baseline cardiac output was 2.93 +/- 0.45 L/min. PulseCO values overestimated cardiac output compared with LiDCO during hemorrhagic shock (2.25 vs. 0.78 L/min). There were no differences in cardiac output determined by PulseCO and LiDCO at 10 and 120 mins after fluid resuscitation. Bland-Altman analysis suggested that PulseCO values were inaccurate after hemorrhage, producing significant bias with wide limits of agreement and percentage error (1.47 +/- 1.46 L/min; 97%). Bias was small but the limits of agreement and percentage error were large for cardiac output at 10 and 120 mins after resuscitation (-0.1 +/- 1.88 [98%] and -0.17 +/- 1.32 [71%] L/min, respectively). There appeared to be a negative but not significant correlation after hemorrhage (r = -.45; p = .15). Conclusions: PulseCO determination of cardiac output does not accurately predict rapid decreases in cardiac output or the effects of fluid resuscitation in dogs. Recalibration of PulseCO may be necessary after any apparent or suspected decrease in cardiac preload, afterload, or contractility.
引用
收藏
页码:1724 / 1729
页数:6
相关论文
共 35 条
[1]  
[Anonymous], 1904, EXPT STUDY BLOOD PRE
[2]   Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output [J].
Bein, B ;
Worthmann, F ;
Tonner, PH ;
Paris, A ;
Steinfath, M ;
Hedderich, J ;
Scholz, J .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2004, 18 (02) :185-189
[3]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[4]   Comparison of arterial pressure waveform analysis with the lithium dilution technique to monitor cardiac output in anesthetized dogs [J].
Chen, HC ;
Sinclair, MD ;
Dyson, DH ;
McDonell, WN ;
Sears, WC .
AMERICAN JOURNAL OF VETERINARY RESEARCH, 2005, 66 (08) :1430-1436
[5]   Noninvasive techniques for measurements of cardiac output [J].
Cholley, BP ;
Payen, D .
CURRENT OPINION IN CRITICAL CARE, 2005, 11 (05) :424-429
[6]   Comparison of lithium dilution and thermodilution cardiac output measurements in anaesthetised neonatal foals [J].
Corley, KTT ;
Donaldson, LL ;
Furr, MO .
EQUINE VETERINARY JOURNAL, 2002, 34 (06) :598-601
[7]   Accurate, noninvasive continuous monitoring of cardiac output by whole-body electrical bioimpedance [J].
Cotter, G ;
Moshkovitz, Y ;
Kaluski, E ;
Cohen, AJ ;
Miller, H ;
Goor, D ;
Vered, Z .
CHEST, 2004, 125 (04) :1431-1440
[8]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[9]   Impact of dopamine or dobutamine infusions on cardiovascular variables after rapid blood loss and volume replacement during isoflurane-induced anesthesia in dogs [J].
Dyson, DH ;
Sinclair, MD .
AMERICAN JOURNAL OF VETERINARY RESEARCH, 2006, 67 (07) :1121-1130
[10]   Lithium: Clinical considerations in internal medicine [J].
Freeman, Marlene P. ;
Freeman, Scott A. .
AMERICAN JOURNAL OF MEDICINE, 2006, 119 (06) :478-481