bone scanning;
dual-energy X-ray absorptiometry (DXA);
D O I:
10.1385/JCD:7:4:457
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
We present two patients in whom marked changes in dual-energy X-ray absorptiometry (DXA-)-measured bone mineral density were found to be the result of prior administration of technetium(-99m) methylene diphosphonate (MDP) for bone scanning. In patient 1, an artifactual 50% decrease in lumbar spine and 19% decrease in left total hip bone density over a 14-mo period was the result of technetium(-99m) injection 3 h before the second bone density measurement. In patient 2, a dramatic increase in bone density at 6 mo was the result of artifactual reduction in bone density on the baseline study as a result of technetium(-99m) injection 2 It earlier. Clinicians and technologists should be aware that technetium(-99m) used for bone scanning could cause artifactual reduction in DXA-measured bone mineral density.