Reliability analysis framework for computer-assisted medical decision systems

被引:7
作者
Habas, Piotr A. [1 ]
Zurada, Jacek M.
Elmaghraby, Adel S.
Tourassi, Georgia D.
机构
[1] Univ Louisville, Computat Intelligent Lab, Dept Elect & Comp Engn, Louisville, KY 40292 USA
[2] Univ Louisville, Dept Comp Engn & Comp Sci, Louisville, KY 40292 USA
[3] Duke Univ, Med Ctr, Dept Radiol, Digital Adv Imaging Labs, Durham, NC 27705 USA
关键词
mammography; computer-assisted detection (CAD); reliability analysis; artificial intelligence; receiver operating characteristic (ROC);
D O I
10.1118/1.2432409
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present a technique that enhances computer-assisted decision (CAD) systems with the ability to assess the reliability of each individual decision they make. Reliability assessment is achieved by measuring the accuracy of a CAD system with known cases similar to the one in question. The proposed technique analyzes the feature space neighborhood of the query case to dynamically select an input-dependent set of known cases relevant to the query. This set is used to assess the local (query-specific) accuracy of the CAD system. The estimated local accuracy is utilized as a reliability measure of the CAD response to the query case. The underlying hypothesis of the study is that CAD decisions with higher reliability are more accurate. The above hypothesis was tested using a mammographic database of 1337 regions of interest (ROIs) with biopsy-proven ground truth (681 with masses, 656 with normal parenchyma). Three types of decision models, (i) a back-propagation neural network (BPNN), (ii) a generalized regression neural network (GRNN), and (iii) a support vector machine (SVM), were developed to detect masses based on eight morphological features automatically extracted from each ROI The performance of all decision models was evaluated using the Receiver Operating Characteristic (ROC) analysis. The study showed that the proposed reliability measure is a strong predictor of the CAD system's case-specific accuracy. Specifically, the ROC area index for CAD predictions with high reliability was significantly better than for those with low reliability values. This result was consistent across all decision models investigated in the study. The proposed case-specific reliability analysis technique could be used to alert the CAD user when an opinion that is unlikely to be reliable is offered. The technique can be easily deployed in the clinical environment because it is applicable with a wide range of classifiers regardless of their structure and it requires neither additional training nor building multiple decision models to assess the case-specific CAD accuracy. 0 2007 American Association of Physicists in Medicine.
引用
收藏
页码:763 / 772
页数:10
相关论文
共 33 条
[1]   A COMPARISON BETWEEN CRITERION FUNCTIONS FOR LINEAR CLASSIFIERS, WITH AN APPLICATION TO NEURAL NETS [J].
BARNARD, E ;
CASASENT, D .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1989, 19 (05) :1030-1041
[2]   NOVELTY DETECTION AND NEURAL-NETWORK VALIDATION [J].
BISHOP, CM .
IEE PROCEEDINGS-VISION IMAGE AND SIGNAL PROCESSING, 1994, 141 (04) :217-222
[3]  
Bishop CM., 1995, Neural networks for pattern recognition
[4]   Classifier design for computer-aided diagnosis: Effects of finite sample size on the mean performance of classical and neural network classifiers [J].
Chan, HP ;
Sahiner, B ;
Wagner, RF ;
Petrick, N .
MEDICAL PHYSICS, 1999, 26 (12) :2654-2668
[5]   ORTHOGONAL LEAST-SQUARES LEARNING ALGORITHM FOR RADIAL BASIS FUNCTION NETWORKS [J].
CHEN, S ;
COWAN, CFN ;
GRANT, PM .
IEEE TRANSACTIONS ON NEURAL NETWORKS, 1991, 2 (02) :302-309
[6]   COMPUTATIONAL EXPERIENCE WITH CONFIDENCE-REGIONS AND CONFIDENCE-INTERVALS FOR NONLINEAR LEAST-SQUARES [J].
DONALDSON, JR ;
SCHNABEL, RB .
TECHNOMETRICS, 1987, 29 (01) :67-82
[7]   The application of neural networks to the papermaking industry [J].
Edwards, PJ ;
Murray, AF ;
Papadopoulos, G ;
Wallace, AR ;
Barnard, J ;
Smith, G .
IEEE TRANSACTIONS ON NEURAL NETWORKS, 1999, 10 (06) :1456-1464
[8]  
Efron B., 1993, INTRO BOOTSTRAP MONO, DOI DOI 10.1201/9780429246593
[9]   DNA: Directional neighborhood analysis for detection of breast masses in screening mammograms [J].
Eltonsy, NH ;
Tourassi, GD ;
Habas, PA ;
Elmaghraby, AS .
Medical Imaging 2005: Image Processing, Pt 1-3, 2005, 5747 :38-47
[10]  
HABAS PA, 2006, P SOC PHOTO-OPT INS, V6144, P1759