Magnetic resonance imaging in children with implants

被引:8
作者
Jaimes, Camilo [1 ]
Biaggotti, Diane [2 ]
Sreedher, Gayathri [3 ]
Chaturvedi, Apeksha [4 ]
Moore, Michael M. [5 ]
Danehy, Amy R. [1 ,6 ]
机构
[1] Boston Childrens Hosp, Div Neuroradiol, Dept Radiol, 300 Longwood Ave,2nd Floor,Main Bldg, Boston, MA 02115 USA
[2] Boston Childrens Hosp, Dept Radiol, Boston, MA USA
[3] Akron Childrens Hosp, Pediat Neuroradiol, Akron, OH 44308 USA
[4] Univ Rochester, Med Ctr, Pediat Radiol, Rochester, NY 14642 USA
[5] Penn State Univ, Milton S Hershey Med Ctr, Pediat Radiol, Hershey, PA 17033 USA
[6] Harvard Med Sch, Boston Childrens Hosp, MRI Safety Comm, Boston, MA 02115 USA
关键词
Children; Implants; Magnetic resonance imaging; Safety; CONTROLLED GROWING RODS; FIELD INTERACTIONS; MRI; SAFETY; DEVICES; IDENTIFICATION; ASSOCIATION; RADIOLOGY; STATEMENT; SEDATION;
D O I
10.1007/s00247-021-04965-5
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
As access to MRI in pediatrics increases, the radiologist needs to become acquainted with the basic principles of MRI safety. As part of the image acquisition, the static magnetic field, gradient system, and the radiofrequency transmit-receive coil interact with medical and non-medical implants and can result in serious injury. The main stage of risk triage is based on the determination of whether the implant is MRI-safe, conditional, unsafe or unknown. Guiding principles include the strict adherence to manufacturer specifications for MRI-conditional implants and the assumption that an unknown implant is MR-unsafe. In this article we review considerations for common medical implants encountered in pediatrics including ventriculoperitoneal shunts, orthopedic hardware, orthodontic hardware, pacemakers, vascular stents, vagal nerve stimulators and cochlear implants. Finally, we review a set of high-yield considerations, including the non-communicative patient (sedated or non-verbal), susceptibility artifacts from unclear source, and the approach to an unknown implant.
引用
收藏
页码:748 / 759
页数:12
相关论文
共 51 条
  • [1] Cardiac MRI of patients with implanted electrical cardiac devices
    Ainslie, Mark
    Miller, Christopher
    Brown, Benjamin
    Schmitt, Matthias
    [J]. HEART, 2014, 100 (05) : 363 - 369
  • [2] American College of Radiology, 2020, ACR MAN MRI SAF
  • [3] Tattoo-Based Noninvasive Glucose Monitoring: A Proof-of-Concept Study
    Bandodkar, Amay J.
    Jia, Wenzhao
    Yardimci, Ceren
    Wang, Xuan
    Ramirez, Julian
    Wang, Joseph
    [J]. ANALYTICAL CHEMISTRY, 2015, 87 (01) : 394 - 398
  • [4] Transcutaneous Vagus Nerve Stimulation (tVNS) for Treatment of Drug-Resistant Epilepsy: A Randomized, Double-Blind Clinical Trial (cMPsE02)
    Bauer, S.
    Baier, H.
    Baumgartner, C.
    Bohlmann, K.
    Fauser, S.
    Graf, W.
    Hillenbrand, B.
    Hirsch, M.
    Last, C.
    Lerche, H.
    Mayer, T.
    Schulze-Bonhage, A.
    Steinhoff, B. J.
    Weber, Y.
    Hartlep, A.
    Rosenow, F.
    Hamer, H. M.
    [J]. BRAIN STIMULATION, 2016, 9 (03) : 356 - 363
  • [5] Vagal Nerve Stimulation for Treatment-Resistant Depression
    Carreno, Flavia R.
    Frazer, Alan
    [J]. NEUROTHERAPEUTICS, 2017, 14 (03) : 716 - 727
  • [6] Cochlear, 2019, ONLINE DOCUMENT
  • [7] MR Imaging and Cochlear Implants with Retained Internal Magnets: Reducing Artifacts near Highly Inhomogeneous Magnetic Fields
    Edmonson, Heidi A.
    Carlson, Matthew L.
    Patton, Alice C.
    Watson, Robert E.
    [J]. RADIOGRAPHICS, 2018, 38 (01) : 94 - 106
  • [8] Update on Evaluation and Treatment of Scoliosis
    El-Hawary, Ron
    Chukwunyerenwa, Chukwudi
    [J]. PEDIATRIC CLINICS OF NORTH AMERICA, 2014, 61 (06) : 1223 - +
  • [9] Magnetic Resonance Imaging Safety of Magnetically Controlled Growing Rods in an In Vivo Animal Model
    Eroglu, Mehmet
    Demirkiran, Gokhan
    Kocyigit, Ismail Aykut
    Bilgili, Hasan
    Kaynar, Mehmet Burak
    Bumin, Ali
    Ozcan, Sadan
    Yazici, Muharrem
    [J]. SPINE, 2017, 42 (09) : E504 - E508
  • [10] International Geomagnetic Reference Field: the eleventh generation
    Finlay, C. C.
    Maus, S.
    Beggan, C. D.
    Bondar, T. N.
    Chambodut, A.
    Chernova, T. A.
    Chulliat, A.
    Golovkov, V. P.
    Hamilton, B.
    Hamoudi, M.
    Holme, R.
    Hulot, G.
    Kuang, W.
    Langlais, B.
    Lesur, V.
    Lowes, F. J.
    Luehr, H.
    Macmillan, S.
    Mandea, M.
    McLean, S.
    Manoj, C.
    Menvielle, M.
    Michaelis, I.
    Olsen, N.
    Rauberg, J.
    Rother, M.
    Sabaka, T. J.
    Tangborn, A.
    Toffner-Clausen, L.
    Thebault, E.
    Thomson, A. W. P.
    Wardinski, I.
    Wei, Z.
    Zvereva, T. I.
    [J]. GEOPHYSICAL JOURNAL INTERNATIONAL, 2010, 183 (03) : 1216 - 1230