A new 3-dimensional method for measuring precision in surgical navigation and methods to optimize navigation accuracy

被引:30
作者
Kleck, Christopher J. [1 ]
Cullilmore, Ian [2 ]
LaFleur, Matthew [1 ]
Lindley, Emily [1 ]
Rentschler, Mark E. [2 ]
Burger, Evalina L. [1 ]
Cain, Christopher M. J. [1 ]
Patel, Vikas V. [1 ]
机构
[1] Univ Colorado, Dept Orthoped, Anschutz Med Campus,12631 E 17th Ave, Aurora, CO 80045 USA
[2] Univ Colorado, Dept Mech Engn, Boulder, CO 80309 USA
关键词
Pedicle screw; Navigation; Free-hand; O-arm; PEDICLE SCREW PLACEMENT; THORACIC SPINE; MORPHOMETRIC-ANALYSIS; IDIOPATHIC SCOLIOSIS; RADIATION-EXPOSURE; LUMBAR VERTEBRAE; IMAGE GUIDANCE; SACRAL SPINE; IN-VIVO; O-ARM;
D O I
10.1007/s00586-015-4235-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Description of a novel method for evaluation of pedicle screws in 3 dimensions utilizing O-arm(A (R)) and StealthStation(A (R)) navigation; identifying sources of error, and pearls for more precise screw placement. O-arm and StealthStation navigation were utilized to place pedicle screws. Initial and final O-arm scans were performed, and the projected pedicle probe track, projected pedicle screw track, and final screw position were saved for evaluation. They were compared to evaluate the precision of the system as well as overall accuracy of final screw placement. Thoracolumbar deformity patients were analyzed, with 153 of 158 screws in adequate position. Only 5 screws were malpositioned, requiring replacement or removal. All 5 were breached laterally and no neurologic or other complications were noted in any of these patients. This resulted in 97 % accuracy using the navigation system, and no neurological injuries or deficits. The average distance of the screw tip and angle of separation for the predicted path versus the final pedicle screw position were analyzed for precision. The mean screw tip distance from the projected tip was 6.43 mm, with a standard deviation of 3.49 mm when utilizing a navigated probe alone and 5.92 mm with a standard deviation of 3.50 mm using a navigated probe and navigated screwdriver (p = 0.23). Mean angle differences were 4.02A degrees and 3.09A degrees respectively (p < 0.01), with standard deviations of 2.63A degrees and 2.12A degrees. This new technique evaluating precision of screw placement in 3 dimensions improves the ability to define screw placement. Pedicle screw position at final imaging showed the use of StealthStation navigation to be accurate and safe. As this is a preliminary evaluation, we have identified several factors affecting the precision of pedicle screw final position relative to that predicted with navigation.
引用
收藏
页码:1764 / 1774
页数:11
相关论文
共 43 条
  • [1] Radiation Exposure to the Spine Surgeon in Lumbar and Thoracolumbar Fusions With the Use of an Intraoperative Computed Tomographic 3-Dimensional Imaging System
    Abdullah, Kalil G.
    Bishop, Frank S.
    Lubelski, Daniel
    Steinmetz, Michael P.
    Benzel, Edward C.
    Mroz, Thomas E.
    [J]. SPINE, 2012, 37 (17) : E1074 - E1078
  • [2] Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine
    Amiot, LP
    Lang, K
    Putzier, M
    Zippel, H
    Labelle, H
    [J]. SPINE, 2000, 25 (05) : 606 - 614
  • [3] Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review
    Aoude, Ahmed A.
    Fortin, Maryse
    Figueiredo, Rainer
    Jarzem, Peter
    Ouellet, Jean
    Weber, Michael H.
    [J]. EUROPEAN SPINE JOURNAL, 2015, 24 (05) : 990 - 1004
  • [4] Use of CT-Based Intraoperative Spinal Navigation: Management of Radiation Exposure to Operator, Staff, and Patients
    Bandela, John R.
    Jacob, R. Patrick
    Arreola, Manuel
    Griglock, Thomas M.
    Bova, Frank
    Yang, Mu
    [J]. WORLD NEUROSURGERY, 2013, 79 (02) : 390 - 394
  • [5] Accuracy of thoracic pedicle screws in patients with and without coronal plane spinal deformities
    Belmont, PJ
    Klemme, WR
    Robinson, M
    Polly, DW
    [J]. SPINE, 2002, 27 (14) : 1558 - 1566
  • [6] In vivo accuracy of thoracic pedicle screws
    Belmont, PJ
    Klemme, WR
    Dhawan, A
    Polly, DW
    [J]. SPINE, 2001, 26 (21) : 2340 - 2346
  • [7] BOUCHER HH, 1959, J BONE JOINT SURG BR, V41, P248
  • [8] Accuracy of pedicle screw placement in lumbar vertebrae
    Castro, WHM
    Halm, H
    Jerosch, J
    Malms, J
    Steinbeck, J
    Blasius, S
    [J]. SPINE, 1996, 21 (11) : 1320 - 1324
  • [9] Pedicle instrumentation in the thoracic spine - A morphometric and cadaveric study for placement of screws
    Cinotti, G
    Gumina, S
    Ripani, M
    Postacchini, F
    [J]. SPINE, 1999, 24 (02) : 114 - 119
  • [10] COMPLICATIONS ASSOCIATED WITH THE TECHNIQUE OF PEDICLE SCREW FIXATION - A SELECTED SURVEY OF ABS MEMBERS
    ESSES, SI
    SACHS, BL
    DREYZIN, V
    [J]. SPINE, 1993, 18 (15) : 2231 - 2239