Experimental Validation of a Pin less Femoral Reference Array for Computer-Assisted Hip Arthroplasty

被引:10
作者
Renkawitz, Tobias [1 ]
Wegner, Melanie [2 ]
Gneiting, Sabine [2 ]
Haimerl, Martin [2 ]
Sendtner, Ernst [1 ]
Kalteis, Thomas [3 ]
Grifka, Joachim [1 ]
机构
[1] Univ Regensburg, Med Ctr, Dept Orthopaed Surg, Asklepios Klinikum Bad Abbach, D-93077 Bad Abbach, Germany
[2] BrainLAB AG, Feldkirchen, Germany
[3] Orthopaed Joint Practice, Munich, Germany
关键词
leg situation algorithm; hip arthroplasty; total joint replacement; computer-assisted surgery; total hip replacement; REGISTRATION; FRACTURE; HEAD;
D O I
10.1002/jor.21039
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The use of computer navigation systems during total hip arthroplasty requires the femoral fixation of a reflective dynamic reference base (DRB), which theoretically involves the risk of bony fracture, infection, and pin loosening. The first objective of this study was to evaluate the relative movements between a novel, noninvasive external femoral DRB system and the femur. Secondly, the maximum effects of these 3D movements on intraoperative, computer-assisted leg length and offset measures were evaluated. An imageless navigation system was used to track the positions of the soft tissue attached, pinless DRB relative to an invasive reference marker on the femur during a less-invasive, anterior surgical hip approach. Relative translatory movements up to 8.2 mm mediolaterally and up to 8.8 degrees in rotation were measured. Using a measurement technique in which the calculation of leg length and offset changes is primarily based on a specific realignment of the leg, maximum differences of 1.3 mm for leg length and 1.2 mm for offset were found when comparing the pin-based and pinless methods. Thus, invasive fixation techniques with screws or pins are still the method of choice when standard measurement algorithms for intraoperative leg length and offset measures are used. Though direct translatory and rotational variations between the pinless array and the femoral bone were detected, the pinless array can be used to assess leg length and offset when used with a specific measurement technique that compensates for such variations. (C) 2009 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 28:583-588, 2010
引用
收藏
页码:583 / 588
页数:6
相关论文
共 20 条
  • [1] Duwelius Paul J, 2008, Instr Course Lect, V57, P215
  • [2] Image-to-patient registration techniques in head surgery
    Eggers, G.
    Muehling, J.
    Marmulla, R.
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 35 (12) : 1081 - 1095
  • [3] Jenny JY, 2006, ORTHOPADE, V35, P1038, DOI 10.1007/s00132-006-0994-y
  • [4] Fractures associated with computer-navigated total knee arthroplasty
    Jung, Ho-Joong
    Jung, Young-Bok
    Song, Kwang-Sup
    Park, Se-Jin
    Lee, Jong-Seok
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (10) : 2280 - 2284
  • [5] Experimental validation of noninvasive referencing in navigated procedures on long bones
    Kendoff, Daniel
    Bogojevic, Aleksander
    Citak, Musa
    Citak, Mustafa
    Maier, Christian
    Maier, Georg
    Krettek, Christian
    Huefner, Tobias
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2007, 25 (02) : 201 - 207
  • [6] Khadem R, 2000, Comput Aided Surg, V5, P98, DOI 10.3109/10929080009148876
  • [7] Li Q, 1999, Comput Aided Surg, V4, P314, DOI 10.1002/(SICI)1097-0150(1999)4:6<314::AID-IGS3>3.0.CO
  • [8] 2-G
  • [9] Majdani O, 2003, LARYNGO RHINO OTOL, V82, P632
  • [10] Advanced surface-recording techniques for computer-assisted oral and maxillofacial surgery
    Marmulla, R
    Mühling, J
    Lüth, T
    Eggers, G
    Hassfeld, S
    [J]. BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2004, 42 (06) : 511 - 519