Is ventricular sensing always right, when it is left?

被引:10
作者
Biffi, Mauro [1 ]
de Zan, Giulia [1 ]
Massaro, Giulia [1 ]
Angeletti, Andrea [1 ]
Martignani, Cristian [1 ]
Boriani, Giuseppe [2 ]
Diemberger, Igor [1 ]
Ziacchi, Matteo [1 ]
机构
[1] Azienda Osped Univ Bologna, Via Massarenti 9, I-40138 Bologna, Italy
[2] Azienda Osped Univ Modena, Modena, Italy
关键词
Arrhythmia detection; Cardiac stimulation; Left ventricular lead; Sensing; Tricuspid regurgitation; CARDIAC RESYNCHRONIZATION THERAPY; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; GENERATOR REPLACEMENT; DEVICE ADVISORIES; ACTIVE FIXATION; FIDELIS LEADS; LONG-TERM; PACEMAKER; COMPLICATIONS; MULTICENTER;
D O I
10.1002/clc.23033
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Ventricular sensing in transvenous cardiac implantable electronic devices (CIEDs) occurs conventionally from the right ventricular (RV) channel, though it evolved from epicardial sensing both in pacemakers and implantable cardioverter-defibrillators (ICDs). Hypothesis: The objective of this study was to observe the reliability of left ventricular (LV) sensing by transvenous leads placed in coronary veins. Methods: LV leads were used for sensing and arrhythmia detection in clinical situations where placement of an RV lead across the tricuspid valve was either not preferred or not feasible, or RV signal was unsuitable for arrhythmia detection, or in the event of sensing failure of an RV lead under advisory in cardiac resynchronization therapy defibrillator (CRTD) recipients. Results: Thirty-seven patients had an IS-1 LV lead connected to the RV port of CIEDs (17 pacemakers, 5 cardiac resynchronization therapy pacemaker [CRTP], 2 ICDs, and 13 CRTDs). Along a median 41 (25-67) months follow-up, lead performance remained stable; there were neither undersensing nor oversensing of non-cardiac signals. VT/VF were correctly detected and terminated by ATP and shocks (one and three patients, respectively); no inappropriate arrhythmia detection. Device reprogramming occurred in four CRTD recipients because of transient counting the QRS (short intervals) when paced in LV-only, and in two with T-wave oversensing. Conclusions: Ventricular sensing by an LV lead is feasible in transvenous devices. Sensing programmability is an unmet need: to fix RV lead sensing issues in cardiac resynchronization therapy (CRT) recipients at no risk of infection (no pocket opening); to avoid interaction with the tricuspid valve; to avoid lead redundancy in the vasculature. Moreover, it will be mandatory owing to the loss of lead interchangeability due to the adoption of DF-4 and quadripolar leads.
引用
收藏
页码:1238 / 1245
页数:8
相关论文
共 36 条
[1]   European Heart Rhythm Association Guidance Document on cardiac rhythm management product performance [J].
Auricchio, Angelo ;
Gropp, Michael ;
Ludgate, Susanne ;
Vardas, Panos ;
Brugada, Josep ;
Priori, Silvia G. .
EUROPACE, 2006, 8 (05) :313-322
[2]   Left ventricular pacing by automatic capture verification [J].
Biffi, Mauro ;
Bertini, Matteo ;
Ziacchi, Matteo ;
Boriani, Giuseppe .
EUROPACE, 2007, 9 (12) :1177-1181
[3]   Transvenous cardioverter-defibrillator implantation in a patient with tricuspid mechanical prosthesis [J].
Biffi, Mauro ;
Bertini, Matteo ;
Ziacchi, Matteo ;
Boriani, Giuseppe .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 (03) :329-331
[4]   Left ventricular lead stabilization to retain cardiac resynchronization therapy at long term: when is it advisable? [J].
Biffi, Mauro ;
Bertini, Matteo ;
Ziacchi, Matteo ;
Diemberger, Igor ;
Martignani, Cristian ;
Boriani, Giuseppe .
EUROPACE, 2014, 16 (04) :533-540
[5]   Implant and Midterm Outcomes of the Subcutaneous Implantable Cardioverter-Defibrillator Registry The EFFORTLESS Study [J].
Boersma, Lucas ;
Barr, Craig ;
Knops, Reinoud ;
Theuns, Dominic ;
Eckardt, Lars ;
Neuzil, Petr ;
Scholten, Marcoen ;
Hood, Margaret ;
Kuschyk, Juergen ;
Jones, Paul ;
Duffy, Elizabeth ;
Husby, Michael ;
Stein, Kenneth ;
Lambiase, Pier D. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (07) :830-841
[6]   Recurrent Implantable Cardioverter-Defibrillator Replacement Is Associated with an Increasing Risk of Pocket-Related Complications [J].
Borleffs, C. Jan Willem ;
Thijssen, Joep ;
de Bie, Mihaly K. ;
van Rees, Johannes B. ;
van Welsenes, Guido H. ;
van Erven, Lieselot ;
Bax, Jeroen J. ;
Cannegieter, Suzanne C. ;
Schalij, Martin J. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (08) :1013-1019
[7]   Management of recalled implantable cardioverter-defibrillator leads at generator replacement: a decision analysis model for Fidelis leads [J].
Burri, Haran ;
Combescure, Christophe .
EUROPACE, 2014, 16 (08) :1210-1217
[8]   Sudden cardiac death: New approaches for implantable cardioverter-defibrillators (ICDs) [J].
Cappato, Riccardo ;
Ali, Hussam .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 237 :38-41
[9]   Tricuspid Valve Dysfunction Following Pacemaker or Cardioverter-Defibrillator Implantation [J].
Chang, James D. ;
Manning, Warren J. ;
Ebrille, Elisa ;
Zimetbaum, Peter J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (18) :2331-2341
[10]   Performance of a novel left ventricular lead with short bipolar spacing for cardiac resynchronization therapy: Primary results of the Attain Performa Quadripolar Left Ventricular Lead Study [J].
Crossley, George H. ;
Biffi, Mauro ;
Johnson, Ben ;
Lin, Albert ;
Gras, Daniel ;
Hussin, Azlan ;
Cuffio, Allen ;
Collier, Jack L. ;
El-Chami, Mikhael ;
Li, Shelby ;
Holloman, Keith ;
Exner, Derek V. .
HEART RHYTHM, 2015, 12 (04) :751-758