Acoustic wave immunosensing of Escherichia coli in water

被引:22
作者
Deobagkar, DD [1 ]
Limaye, V
Sinha, S
Yadava, RDS
机构
[1] Univ Poona, Dept Zool, Pune 411007, Maharashtra, India
[2] Solid State Phys Lab, Delhi 110054, India
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2005年 / 104卷 / 01期
关键词
acoustic wave biosensor; E. coli detection; potability of water;
D O I
10.1016/j.snb.2004.04.106
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper reports an experimental study of E. coli sensing based on surface-skimming-bulk-wave delay line device fabricated on ST-cut quartz. The sensor configuration consisted of an 87.7 MHz oscillator with the delay line as control element in its feedback loop. The antigen-antibody immunochemical reaction was employed on the acoustic wave device surface to generate mass loading and frequency shift. The sensor output was monitored by an RF spectrum analyzer. Highly specific sensing is demonstrated for E. coli cells up to 0.4 cells/mul in water. The methodology offers possibility for developing highly sensitive and fast E. coli sensors for testing of water potability. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:85 / 89
页数:5
相关论文
共 17 条