Inner Filter Effect Correction for Fluorescence Measurements in Microplates Using Variable Vertical Axis Focus

被引:60
作者
Weitner, Tin [1 ]
Friganovic, Tomislav [1 ]
Sakic, Davor [1 ]
机构
[1] Univ Zagreb, Fac Pharm & Biochem, Zagreb 10000, Croatia
关键词
D O I
10.1021/acs.analchem.2c01031
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The inner filter effect (IFE) hinders fluorescence measurements, limiting linear dependence of fluorescence signals to low sample concentrations. Modern microplate readers allow movement of the optical element in the vertical axis, changing the relative position of the focus and thus the sample geometry. The proposed Z-position IFE correction method requires only two fluorescence measurements at different known vertical axis positions (z-positions) of the optical element for the same sample. Samples of quinine sulfate, both pure and in mixtures with potassium dichromate, showed a linear dependence of corrected fluorescence on fluorophore concentration (R-2 > 0.999), up to A(ex) approximate to 2 and A(em) approximate to 0.5. The correction extended linear fluorescence response over approximate to 98% of the concentration range with approximate to 1% deviation of the calibration slope, effectively eliminating the need for sample dilution or separate absorbance measurements to account for IFE. The companion numerical IFE correction method further eliminates the need for any gematric parameters with similar results. Both methods are available online at https://ninfe.science .
引用
收藏
页码:7107 / 7114
页数:8
相关论文
共 21 条
[1]  
Desimoni E., 2015, Pharmaceutica Analytica Acta, V6, P2153, DOI [DOI 10.4172/2153-2435.1000355, 10.4172/2153-2435.1000355]
[2]   Fluorescence of Dyes in Solutions with High Absorbance. Inner Filter Effect Correction [J].
Fonin, Alexander V. ;
Sulatskaya, Anna I. ;
Kuznetsova, Irina M. ;
Turoverov, Konstantin K. .
PLOS ONE, 2014, 9 (07)
[3]   Improvement of Inner Filter Effect Correction Based on Determination of Effective Geometric Parameters Using a Conventional Fluorimeter [J].
Gu, Qun ;
Kenny, Jonathan E. .
ANALYTICAL CHEMISTRY, 2009, 81 (01) :420-426
[4]   CORRECTION OF RIGHT-ANGLE FLUORESCENCE MEASUREMENTS FOR ABSORPTION OF EXCITATION RADIATION [J].
HOLLAND, JF ;
TEETS, RE ;
KELLY, PM ;
TIMNICK, A .
ANALYTICAL CHEMISTRY, 1977, 49 (06) :706-710
[5]   A new approach to the old problem: Inner filter effect type I and II in fluorescence [J].
Kasparek, Adam ;
Smyk, Bogdan .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2018, 198 :297-303
[6]   On the origin and correction for inner filter effects in fluorescence Part I: primary inner filter effect-the proper approach for sample absorbance correction [J].
Kimball, Joseph ;
Chavez, Jose ;
Ceresa, Luca ;
Kitchner, Emma ;
Nurekeyev, Zhangatay ;
Doan, Hung ;
Szabelski, Mariusz ;
Borejdo, Julian ;
Gryczynski, Ignacy ;
Gryczynski, Zygmunt .
METHODS AND APPLICATIONS IN FLUORESCENCE, 2020, 8 (03)
[7]   Inner filter correction of dissolved organic matter fluorescence [J].
Kothawala, Dolly N. ;
Murphy, Kathleen R. ;
Stedmon, Colin A. ;
Weyhenmeyer, Gesa A. ;
Tranvik, Lars J. .
LIMNOLOGY AND OCEANOGRAPHY-METHODS, 2013, 11 :616-630
[8]   EXPERIMENTAL CORRECTION FOR THE INNER-FILTER EFFECT IN FLUORESCENCE-SPECTRA [J].
KUBISTA, M ;
SJOBACK, R ;
ERIKSSON, S ;
ALBINSSON, B .
ANALYST, 1994, 119 (03) :417-419
[9]  
Lakowicz J.R., 2006, PRINCIPLES FLUORESCE, VThird, P27, DOI [10.1007/978-0-387-46312-42, DOI 10.1007/978-0-387-46312-4_2]
[10]   CORRECTION FOR INNER FILTER EFFECTS IN FLUORESCENCE SPECTROSCOPY [J].
LUTZ, HP ;
LUISI, PL .
HELVETICA CHIMICA ACTA, 1983, 66 (07) :1929-1935