In situ Nd isotope analyses in geological materials with signal enhancement and non-linear mass dependent fractionation reduction using laser ablation MC-ICP-MS

被引:96
|
作者
Xu, Lei [1 ]
Hu, Zhaochu [1 ]
Zhang, Wen [1 ]
Yang, Lu [2 ]
Liu, Yongsheng [1 ]
Gao, Shan [1 ]
Luo, Tao [1 ]
Hu, Shenghong [1 ]
机构
[1] China Univ Geosci, State Key Lab Geol Proc & Mineral Resources, Wuhan 430074, Peoples R China
[2] Natl Res Council Canada, Measurement Sci & Stand, Ottawa, ON K1A 0R6, Canada
关键词
U-PB GEOCHRONOLOGY; SM-ND; RATIO ANALYSIS; NEODYMIUM; PRECISE; ZIRCON; HF; SPECTROMETRY; SYSTEMATICS; NITROGEN;
D O I
10.1039/c4ja00243a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To improve the precision and accuracy of in situ Nd isotope analyses in geological samples, high analytical sensitivity and minimized non-linear mass dependent fractionation are desirable. In this study, we investigated the effects of three different cone combinations and the addition of N-2 with the guard electrode, either grounded or floating, on the performance of Nd isotope analyses. Signal enhancements (2.5-3 folds) were observed for the Jet sample cone + X skimmer cone and for the standard sample cone + X skimmer cone, which were compared to those of the standard sample cone + H skimmer cone with the addition of N-2 at 3 or 6 ml min(-1). The influence of a guard electrode on the LA-MC-ICP-MS sensitivity was minimal for dry plasma conditions but became significant in the presence of N-2. The addition of 3 or 6 ml min(-1) N-2 into the carrier gas increased the sensitivity for Nd by a factor of 1.6 with the GE-on mode for the Jet sample cone + X skimmer cone. However, in the GE-off mode, the addition of 2-3 ml min(-1) N-2 decreased the signal intensity of Nd by a factor of two to three. The makeup gas flow rate was found to have a significant effect on the mass bias stability of Nd isotopes, and this effect was different for different cone combinations. To stabilize the mass bias, measurements at elevated makeup gas flow rates was required for the combination of the X skimmer cone with the standard sample cone or Jet sample cone, at the expense of sensitivity. In notable contrast, the mass bias was stable for the combination of the standard sample cone and H skimmer cone under their corresponding optimum makeup gas flow rates. However, we found that the neodymium isotopes exhibited a large non-linear component at elevated makeup gas flow rates that could not be corrected by the accepted mass fractionation laws. The addition of 3-6 ml min(-1) nitrogen to the central gas flow in laser ablation MC-ICP-MS was found to not only significantly enlarge the mass bias stability zone, but also to suppress the non-linear mass dependent fractionation for the combination of X skimmer cone with Jet sample cone or standard sample cone. The accuracy and precision of the developed analytical method for Nd isotope using Jet sample cone + X skimmer cone with the addition of nitrogen has been demonstrated by analyzing a series of reference glasses (JNdi-1 glass, LREE glass, NIST 610) and minerals (apatite, monazite and titanite). Results obtained for Nd-143/Nd-144 ratios in these reference materials are in excellent agreement with published values, demonstrating the capability of the developed analytical method to be an important tool for providing high-quality in situ Nd isotope data in geological samples.
引用
收藏
页码:232 / 244
页数:13
相关论文
共 14 条
  • [1] In Situ Carbon Isotope Analysis by Laser Ablation MC-ICP-MS
    Chen, Wei
    Lu, Jue
    Jiang, Shao-Yong
    Zhao, Kui-Dong
    Duan, Deng-Fei
    ANALYTICAL CHEMISTRY, 2017, 89 (24) : 13415 - 13421
  • [2] Effects of the sampling interface in MC-ICP-MS: Relative elemental sensitivities and non-linear mass dependent fractionation of Nd isotopes
    Newman, Karla
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (01) : 63 - 70
  • [3] Determination of boron isotope compositions of geological materials by laser ablation MC-ICP-MS using newly designed high sensitivity skimmer and sample cones
    Lin, Lin
    Hu, Zhaochu
    Yang, Lu
    Zhang, Wen
    Liu, Yongsheng
    Gao, Shan
    Hu, Shenghong
    CHEMICAL GEOLOGY, 2014, 386 : 22 - 30
  • [4] Gd matrix effects on Eu isotope fractionation using MC-ICP-MS: Optimizing europium isotope ratio measurements in geological rock samples
    Lee, Seung-Gu
    Tanaka, Tsuyoshi
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2021, 469
  • [5] High sensitivity skimmers and non-linear mass dependent fractionation in ICP-MS
    Newman, Karla
    Freedman, Philip A.
    Williams, Jamie
    Belshaw, Nick S.
    Halliday, Alex N.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (06) : 742 - 751
  • [6] In situ Sn isotope analysis of cassiterite (SnO2) by nanosecond laser ablation MC-ICP-MS
    She, Jia-Xin
    Li, Weiqiang
    An, Shichao
    Yang, Tao
    Zhang, Rongqing
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (05) : 1043 - 1056
  • [7] Combined U-Pb and Lu-Hf isotope analyses by laser ablation MC-ICP-MS: methodology and applications
    Matteini, Massimo
    Dantas, Elton L.
    Pimentel, Marcio M.
    Buehn, Bernhard
    ANAIS DA ACADEMIA BRASILEIRA DE CIENCIAS, 2010, 82 (02): : 479 - 491
  • [8] In situ Rb-Sr Dating of Lunar Meteorites Using Laser Ablation MC-ICP-MS
    Zhang, Wen
    Wang, Zaicong
    Hu, Zhaochu
    He, Qi
    Li, Yiheng
    Feng, Yangtong
    Liu, Hong
    Zhang, Xiaojing
    ATOMIC SPECTROSCOPY, 2022, 43 (01) : 60 - 69
  • [9] A single-column and efficient procedure for separating Fe and Mg from geological materials for isotopic analyses using MC-ICP-MS
    Zhu, Guanhong
    Ge, Zhenmin
    Zhang, Le
    Wei, Gangjian
    Ma, Jinlong
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (11) : 2783 - 2790
  • [10] Characterisation of δ142Ce/140Ce Isotope Ratios in Geological Reference Materials using an SSB plus Sm Doping and MC-ICP-MS Method
    Wu, Jiaojiao
    An, Yajun
    Li, Xin
    Liu, Fang
    Xu, Jianbing
    Ling, Mingxing
    Zhang, Zhaofeng
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2024,