Posture-related overdrainage: Comparison of the performance of 10 hydrocephalus shunts in vitro

被引:80
作者
Czosnyka, Z [1 ]
Czosnyka, M [1 ]
Richards, HK [1 ]
Pickard, JD [1 ]
机构
[1] Addenbrookes Hosp, Acad Neurosurg Unit, United Kingdom Shunt Evaluat Lab, Cambridge CB2 2QQ, England
基金
英国医学研究理事会;
关键词
hydrocephalus; overdrainage; shunts; testing;
D O I
10.1097/00006123-199802000-00069
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: Approximately 10 to 30% of shunt revisions may be attributed to posture-related overdrainage. The susceptibility of various hydrocephalus shunts to overdrainage of cerebrospinal fluid requires independent laboratory evaluation. METHODS: Shunts were tested in vitro by using precise computer-controlled equipment that was able to evaluate pressure-flow performance curves under various conditions. Hydrodynamic resistance and opening, closing, and operational pressures were evaluated for at least 28 days with normal (atmospheric) and decreased (-23 mmHg, based on the International Standard Organization/Draft International Standard 7197 standard, which simulates conditions in upright body positions) outlet pressures. RESULTS: Ten different models of valves have been tested to date (Medtronic PS Medical Delta valve, flow-control valve, and lumboperitoneal shunt, Heyer-Schulte in-line, low-profile, and Pudenz flushing valves, Codman-Medos programmable and nonprogrammable valves, Sophy programmable valve, and Cordis Orbis-Sigma valve). The majority of these valves produced significantly negative (less than -10 mm Hg) average intracranial pressures in vertical body positions. In conjunction with nonphysiologically low hydrodynamic resistance (with the exception of the Orbis-Sigma valve, Medtronic PS Medical lumboperitonial shunt, and Heyer-Schulte in-line valve), this may result in overdrainage related to body posture. The clinically reported rate of complications related to overdrainage is probably reduced by the long distal catheter, which increases the resistance of these valves by 100 to 200%. A few shunts (the Delta valve, low-profile valve, and Pudenz flushing valve with anti-siphon devices) offer reasonable resistance to negative outlet pressure, preventing complications related to overdrainage, but all valves with siphon-preventing devices may be blocked by increased subcutaneous pressure. CONCLUSIONS: Shunts without mechanisms preventing ver low intracranial pressure in vertical body positions should be identified and avoided for patients likely to develop complications related to cerebrospinal fluid drainage.
引用
收藏
页码:327 / 333
页数:7
相关论文
共 29 条
  • [1] CHRONIC HEADACHE ASSOCIATED WITH A FUNCTIONING SHUNT - USEFULNESS OF PRESSURE MONITORING
    ABBOTT, R
    EPSTEIN, FJ
    WISOFF, JH
    [J]. NEUROSURGERY, 1991, 28 (01) : 72 - 77
  • [2] INTRACRANIAL-PRESSURE AND CEREBROSPINAL-FLUID OUTFLOW CONDUCTANCE IN HEALTHY-SUBJECTS
    ALBECK, MJ
    BORGESEN, SE
    GJERRIS, F
    SCHMIDT, JF
    SORENSEN, PS
    [J]. JOURNAL OF NEUROSURGERY, 1991, 74 (04) : 597 - 600
  • [3] OVERDRAINAGE AND SHUNT TECHNOLOGY - A CRITICAL COMPARISON OF PROGRAMMABLE, HYDROSTATIC AND VARIABLE RESISTANCE VALVES AND FLOW-REDUCING DEVICES
    ASCHOFF, A
    KREMER, P
    BENESCH, C
    FRUH, K
    KLANK, A
    KUNZE, S
    [J]. CHILDS NERVOUS SYSTEM, 1995, 11 (04) : 193 - 202
  • [4] THE RELATIONSHIP BETWEEN VENTRICULAR FLUID PRESSURE AND BODY POSITION IN NORMAL SUBJECTS AND SUBJECTS WITH SHUNTS - A TELEMETRIC STUDY
    CHAPMAN, PH
    COSMAN, ER
    ARNOLD, MA
    [J]. NEUROSURGERY, 1990, 26 (02) : 181 - 189
  • [5] COLES RA, 1971, LANCET, V1, P1052
  • [6] Hydrodynamic properties of hydrocephalus shunts: United Kingdom shunt evaluation laboratory
    Czosnyka, M
    Czosnyka, Z
    Whitehouse, H
    Pickard, JD
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1997, 62 (01) : 43 - 50
  • [7] COMPARISON BETWEEN CLASSIC-DIFFERENTIAL AND AUTOMATIC SHUNT FUNCTIONING ON THE BASIS OF INFUSION TESTS
    CZOSNYKA, M
    MAKSYMOWICZ, W
    BATORSKI, L
    KOSZEWSKI, W
    CZOSNYKA, Z
    [J]. ACTA NEUROCHIRURGICA, 1990, 106 (1-2) : 1 - 8
  • [8] CZOSNYKA M, 1966, J NEUROL NEUROSUR PS, V60, P549
  • [9] Davson H., 1987, PHYSL PATHOPHYSIOLOG, P189
  • [10] COMPUTER MODELING OF SIPHONING FOR CSF SHUNT DESIGN EVALUATION
    DRAKE, JM
    TENTI, G
    SIVALSGANATHAN, S
    [J]. PEDIATRIC NEUROSURGERY, 1994, 21 (01) : 6 - 15