Validation of Analytical Method for Quantification of Egg Cholesterol Using Reversed Phase-High Performance Liquid Chromatography-Multiwavelength Detector
被引:3
作者:
Maslukhah, Y. L.
[1
]
Faridah, D. N.
[1
]
Lioe, H. N.
[1
]
机构:
[1] Bogor Agr Univ, IPB Univ, Fac Agr Technol, Dept Food Sci & Technol, Kampus IPB Darmaga, Bogor 16680, Indonesia
关键词:
cholesterol;
chromatography criteria;
eggs;
method development;
reversed phase;
OXIDES;
FOODS;
HPLC;
D O I:
10.5398/tasj.2019.42.3.230
中图分类号:
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号:
0905 ;
摘要:
In this research, analytical method of cholesterol content in eggs by Reversed Phase-High Performance Liquid Chromatography-Multiwavelength Detector (RP-HPLC-MWD) was validated. Our experiment validated the modified method of AOAC 994.10:2012 to get a more simple and efficient analytical method of cholesterol content. The sample was saponified using 10% KOH concentration for 15 min at 80 degrees C, then this analytical method was validated. RP-HPLC-MWD condition was at 100% MeOH as a mobile phase, flow rate of 1.0 mL/min, detection UV at 205 nm, cholesterol was detected at 10.38 +/- 0.13 min. As a result, the coefficients of determination for instrument and method linearities reached 0.9991 and 0.9912, respectively. The limits of detection and quantification of RP-HPLC-MWD instrument were found at 5 and 10 mu g/mL, respectively, while the method-detection limit and quantification limit were 250 and 500 mu g/g sample, respectively. Recovery values for the cholesterol analysis ranged from 98.62% to 112.26%, with a precision of 1.05%-3.90%. Additionally, intralab reproducibility was known to reach 3.27%. This validated method can be applied for the analysis of cholesterol in various eggs available in the market.
引用
收藏
页码:230 / 236
页数:7
相关论文