Scanning Electron Microscopy Observations of Failures of Implant Overdenture Bars: A Case Series Report

被引:14
|
作者
Waddell, J. Neil [1 ]
Payne, Alan G. T. [1 ]
Swain, Michael V. [1 ]
Kieser, Jules A. [1 ]
机构
[1] Univ Otago, Sch Dent, Sir John Walsh Res Inst, Oral Implantol Res Grp, Dunedin, New Zealand
关键词
bar; corrosion; creep; failure; fatigue; overdenture; overload; OF-THE-LITERATURE; EDENTULOUS MAXILLA; SUPPORTED OVERDENTURES; RETAINED OVERDENTURES; CORROSION; ALLOYS; COMPLICATIONS; EXTENSIONS; JOINTS; FORCE;
D O I
10.1111/j.1708-8208.2008.00127.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: Soldered or cast bars are used as a standard of care in attachment systems supporting maxillary and mandibular implant overdentures. When failures of these bars occur, currently there is a lack of evidence in relation to their specific etiology, location, or nature. Purpose: To investigate the failure process of a case series of six failed soldered bars, four intact soldered bars, and one intact cast milled bar, which had been supporting implant overdentures. Materials and Methods: A total of 11 different overdenture bars were removed from patients with different configuration of opposing arches. A failed bar (FB) group (n = 6) had failed soldered overdenture bars, which were recovered from patients following up to 2 years of wear before requiring prosthodontic maintenance and repair. An intact bar (IB) group (n = 5) had both soldered bars and a single cast milled bar, which had been worn by patients for 2 to 5 years prior to receiving other aspects of prosthodontic maintenance. All bars were examined using scanning electron microscopy to establish the possible mode of failure (FB) or to identify evidence of potential failure in the future (IB). Results: Evidence of a progressive failure mode of corrosion fatigue and creep were observed on all the FB and IB usually around the solder areas and nonoxidizing gold cylinder. Fatigue and creep were also observed in all the IB. Where the level of corrosion was substantial, there was no evidence of wear from the matrices of the attachment system. Evidence of an instantaneous failure mode, ductile and brittle overload, was observed on the fracture surfaces of all the FB, within the solder and the nonoxidizing gold cylinders, at the solder/cylinder interface. Conclusion: Corrosion, followed by corrosion fatigue, appears to be a key factor in the onset of the failure process for overdenture bars in this case series of both maxillary and mandibular overdentures. Limited sample size and lack of standardization identify trends only but prevent broad interpretation of the findings.
引用
收藏
页码:26 / 38
页数:13
相关论文
共 21 条
  • [1] IMPLANT-RETAINED OVERDENTURE AND O-RING ATTACHMENTS - CASE REPORT
    Bulegon, Andressa
    Rinaldi, Isadora
    Miyagaki, Daniela Cristina
    Sandini Linden, Maria Salete
    de Carli, Joao Paulo
    BIOSCIENCE JOURNAL, 2018, 34 (06): : 1796 - 1804
  • [2] Implant-Supported Overdenture Using Ball Attachments in Maxilla and Mandible: A Case Report
    Krishna, Sai
    Kumar, Rohan
    Sailasri, Kunchala
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (03)
  • [4] Transmission scanning electron microscopy: Defect observations and image simulations
    Callahan, Patrick G.
    Stinville, Jean-Charles
    Yao, Eric R.
    Echlin, McLean P.
    Titus, Michael S.
    De Graef, Marc
    Gianola, Daniel S.
    Pollock, Tresa M.
    ULTRAMICROSCOPY, 2018, 186 : 49 - 61
  • [5] Maxillary implant overdenture retained by four unsplinted attachments and opposed by a natural or fixed dentition: Five-year clinical outcomes. A prospective case series
    Bouhy, Alice
    Lamy, Marc
    Altaep, Yaman
    Lambert, France
    CLINICAL ORAL IMPLANTS RESEARCH, 2023, 34 (04) : 285 - 296
  • [6] Fractured zirconia implants and related implant designs: scanning electron microscopy analysis
    Osman, Reham B.
    Ma, Sunyoung
    Duncan, Warwick
    De Silva, Rohana K.
    Siddiqi, Allauddin
    Swain, Michael V.
    CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (05) : 592 - 597
  • [7] Implant-retained overdenture following hemiglossectomy: a 10-year clinical case report
    Taira, Y
    Sekine, J
    Sawase, T
    Atsuta, M
    JOURNAL OF ORAL REHABILITATION, 2006, 33 (04) : 313 - 315
  • [8] PROSTHODONTIC TREATMENT OF AN EDENTULOUS EPILEPTIC PATIENT WITH AN IMPLANT-RETAINED OVERDENTURE. A CASE REPORT
    Karolyhazy, Katalin
    Schmidt, Peter
    Bogdan, Sandor
    Hermann, Peter
    Aranyi, Zsuzsanna
    IDEGGYOGYASZATI SZEMLE-CLINICAL NEUROSCIENCE, 2014, 67 (9-10): : 342 - 346
  • [9] Dental rehabilitation of a cleft lip and palate patient by implant-supported overdenture: A case report
    Arshad, Mahnaz
    Ameri, Narges
    Heidari, Abdolreza
    Shirani, Gholamreza
    CLINICAL CASE REPORTS, 2020, 8 (10): : 1932 - 1936
  • [10] Use of a single implant to retain a mandibular complete overdenture on the compromised atrophic alveolar ridge: a case report
    Schneider, Galen B.
    Synan, William J.
    SPECIAL CARE IN DENTISTRY, 2011, 31 (04) : 138 - 142