Design and Implementation of a Hypothermic Machine Perfusion Device for Clinical Preservation of Isolated Organs

被引:3
作者
Shen, Fei [1 ]
Yan, Ruqiang [1 ]
机构
[1] Southeast Univ, Sch Instrument Sci & Engn, Nanjing 210096, Jiangsu, Peoples R China
关键词
HMP device; physiology parameters monitoring; modified Bayes estimation; fuzzy-PID controller; EXPANDED CRITERIA DONOR; PULSATILE PRESERVATION; COLD-STORAGE; TRANSPLANTATION; KIDNEY; CONTROLLER;
D O I
10.3390/s17061256
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The imbalance between limited organ supply and huge potential need has hindered the development of organ-graft techniques. In this paper a low-cost hypothermic machine perfusion (HMP) device is designed and implemented to maintain suitable preservation surroundings and extend the survival life of isolated organs. Four necessary elements (the machine perfusion, the physiological parameter monitoring, the thermostatic control and the oxygenation apparatus) involved in this HMP device are introduced. Especially during the thermostatic control process, a modified Bayes estimation, which introduces the concept of improvement factor, is realized to recognize and reduce the possible measurement errors resulting from sensor faults and noise interference. Also, a fuzzy-PID controller contributes to improve the accuracy and reduces the computational load using the DSP. Our experiments indicate that the reliability of the instrument meets the design requirements, thus being appealing for potential clinical preservation applications.
引用
收藏
页数:19
相关论文
共 36 条
[1]  
BARBER WH, 1988, TRANSPLANT P, V20, P865
[2]  
BARUA SN, 1954, P ROY SOC A, V227, P133, DOI DOI 10.1098/RSPA.1954.0285
[3]   Determination of the refrigerating capacity of cryogenic gas machines for cooling current leads containing high-temperature superconductors [J].
Buyanov Yu.L. ;
Veselovskii A.S. ;
Baev V.P. .
Journal of Engineering Physics and Thermophysics, 2011, 84 (3) :678-688
[4]   Analysis and design for a class of complex control systems .2. Fuzzy controller design [J].
Cao, SG ;
Rees, NW ;
Feng, G .
AUTOMATICA, 1997, 33 (06) :1029-1039
[5]   Fuzzy PID controller: Design, performance evaluation, and stability analysis [J].
Carvajal, J ;
Chen, GR ;
Ogmen, H .
INFORMATION SCIENCES, 2000, 123 (3-4) :249-270
[6]  
CHOTAI A, 1991, IFAC WORK S, V1991, P41
[7]   Machine perfusion for cardiac allograft preservation [J].
Cobert, Michael L. ;
West, LaShondra M. ;
Jessen, Michael E. .
CURRENT OPINION IN ORGAN TRANSPLANTATION, 2008, 13 (05) :526-530
[8]   A RANDOMIZED PROSPECTIVE TRIAL OF COLD-STORAGE VERSUS PULSATILE PERFUSION FOR CADAVER KIDNEY-PRESERVATION [J].
HALLORAN, P ;
APRILE, M .
TRANSPLANTATION, 1987, 43 (06) :827-832
[9]  
Hu X., 2013, CHIN MED J, V35, P1105
[10]   FUNCTIONAL PRESERVATION OF MAMMALIAN KIDNEY .4. FUNCTIONAL EFFECTS OF PERFUSION WITH DIMETHYL-SULFOXIDE (DMSO) AT NORMOTHERMIA [J].
KAROW, AM ;
JESKE, AH .
CRYOBIOLOGY, 1976, 13 (04) :448-454