Attenuation parameter and liver stiffness measurement using FibroTouch vs Fibroscan in patients with chronic liver disease

被引:20
作者
Ng, Ying Zhuang [1 ]
Lai, Lee Lee [1 ,2 ]
Wong, Sui Weng [1 ]
Mohamad, Siti Yatimah [1 ]
Chuah, Kee Huat [1 ]
Chan, Wah Kheong [1 ]
机构
[1] Univ Malaya, Fac Med, Dept Med, Gastroenterol & Hepatol Unit, Kuala Lumpur, Malaysia
[2] Univ Malaya, Dept Nursing Sci, Kuala Lumpur, Malaysia
关键词
HEPATIC STEATOSIS; XL PROBE; QUANTIFICATION; DIAGNOSIS; CRITERIA;
D O I
10.1371/journal.pone.0250300
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background & aim We studied FibroTouch (FT) and Fibroscan (FS) examination results and their repeatability when performed by healthcare personnel of different background. Methods FT and FS examinations were performed on patients with chronic liver disease by two operators, a doctor and a nurse, twice on each patient, at two different time points, independent of each other. Results The data for 163 patients with 1304 examinations was analyzed. There was strong correlation between FT and FS for attenuation parameter (Spearman's rho 0.76, p<0.001) and liver stiffness measurement (LSM) (Spearman's rho 0.70, p<0.001). However, FT produced higher value at lower attenuation parameter and LSM, and lower value at higher attenuation parameter and LSM. There was substantial agreement when using 15kPa LSM cut-off, but only moderate agreement when using 10kPa and 20kPa LSM cut-offs and 248dB/m, 268dB/m and 280dB/m attenuation parameter cut-offs. The IQR for attenuation parameter and IQR/median for LSM were significantly lower for FT compared with FS (4dB/m vs 27dB/m, p<0.001, and 10 vs 12, p<0.001, respectively). The intra- and inter-observer reliability of attenuation parameter and LSM using FT and FS were good to excellent with intraclass correlation coefficients 0.89-0.99. FT had shorter examination time (33s vs 47s, p<0.001) and less invalid measurements (0 vs 2, p<0.001). Conclusion Measurements obtained with FT and FS strongly correlated, but significant differences in their absolute values, consistency, examination time and number of invalid measurements were observed. Either device can be used by healthcare personnel of different backgrounds when sufficiently trained.
引用
收藏
页数:13
相关论文
共 27 条
[1]   Hepatitis C Guidance 2019 Update: American Association for the Study of Liver Diseases-Infectious Diseases Society of America Recommendations for Testing, Managing, and Treating Hepatitis C Virus Infection [J].
Ghany M.G. ;
Morgan T.R. .
HEPATOLOGY, 2020, 71 (02) :686-721
[2]   The metabolic syndrome - a new worldwide definition [J].
Alberti, KGMM ;
Zimmet, P ;
Shaw, J .
LANCET, 2005, 366 (9491) :1059-1062
[3]   The new BMI criteria for Asians by the regional office for the Western Pacific Region of WHO are suitable for screening of overweight to prevent metabolic syndrome in elder Japanese workers [J].
Anuurad, E ;
Shiwaku, K ;
Nogi, A ;
Kitajima, K ;
Enkhmaa, B ;
Shimono, K ;
Yamane, Y .
JOURNAL OF OCCUPATIONAL HEALTH, 2003, 45 (06) :335-343
[4]   Determination of reliability criteria for liver stiffness evaluation by transient elastography [J].
Boursier, Jerome ;
Zarski, Jean-Pierre ;
de Ledinghen, Victor ;
Rousselet, Marie-Christine ;
Sturm, Nathalie ;
Lebail, Brigitte ;
Fouchard-Hubert, Isabelle ;
Gallois, Yves ;
Oberti, Frederic ;
Bertrais, Sandrine ;
Cales, Paul .
HEPATOLOGY, 2013, 57 (03) :1182-1191
[5]   Can the same controlled attenuation parameter cut-offs be used for M and XL probes for diagnosing hepatic steatosis? [J].
Chan, Wah-Kheong ;
Mustapha, Nik Raihan Nik ;
Mahadeva, Sanjiv ;
Wong, Vincent Wai-Sun ;
Cheng, Jenny Yeuk-Ki ;
Wong, Grace Lai-Hung .
JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2018, 33 (10) :1787-1794
[6]   Controlled attenuation parameter using the FibroScan® XL probe for quantification of hepatic steatosis for non-alcoholic fatty liver disease in an Asian population [J].
Chan, Wah-Kheong ;
Mustapha, Nik Raihan Nik ;
Wong, Grace Lai-Hung ;
Wong, Vincent Wai-Sun ;
Mahadeva, Sanjiv .
UNITED EUROPEAN GASTROENTEROLOGY JOURNAL, 2017, 5 (01) :76-85
[7]   Controlled attenuation parameter for the detection and quantification of hepatic steatosis in nonalcoholic fatty liver disease [J].
Chan, Wah-Kheong ;
Mustapha, Nik Raihan Nik ;
Mahadeva, Sanjiv .
JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2014, 29 (07) :1470-1476
[8]  
Chen G F, 2017, Zhonghua Gan Zang Bing Za Zhi, V25, P145, DOI 10.3760/cma.j.issn.1007-3418.2017.02.013
[10]   Diagnosis of liver fibrosis and cirrhosis using liver stiffness measurement: Comparison between M and XL probe of FibroScan® [J].
de Ledinghen, Victor ;
Wong, Vincent Wai-Sun ;
Vergniol, Julien ;
Wong, Grace Lai-Hung ;
Foucher, Juliette ;
Chu, Shirley Ho-Ting ;
Le Bail, Brigitte ;
Choi, Paul Cheung-Lung ;
Chermak, Faiza ;
Yiu, Karen Kar-Lum ;
Merrouche, Wassil ;
Chan, Henry Lik-Yuen .
JOURNAL OF HEPATOLOGY, 2012, 56 (04) :833-839