Coherent Raman Scattering Microscopy for Evaluation of Head and Neck Carcinoma

被引:15
作者
Hoesli, Rebecca C. [1 ]
Orringer, Daniel A. [2 ]
McHugh, Jonathan B. [3 ]
Spector, Matthew E. [1 ]
机构
[1] Univ Michigan Hlth Syst, Dept Otolaryngol Head & Neck Surg, Ann Arbor, MI USA
[2] Univ Michigan Hlth Syst, Dept Neurosurg, Ann Arbor, MI USA
[3] Univ Michigan Hlth Syst, Dept Pathol, Ann Arbor, MI USA
关键词
coherent Raman scattering microscopy; head and neck carcinoma; H&E stain; histopathology; SQUAMOUS-CELL CARCINOMA; FROZEN-SECTION; ORAL-CANCER; SPECTROSCOPY; UTILITY; TISSUE; IMPACT;
D O I
10.1177/0194599817700388
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objective. We aim to describe a novel, label-free, real-time imaging technique, coherent Raman scattering (CRS) microscopy, for histopathological evaluation of head and neck cancer. We evaluated the ability of CRS microscopy to delineate between tumor and nonneoplastic tissue in tissue samples from patients with head and neck cancer. Study Design. Prospective case series. Setting. Tertiary care medical center. Subjects and Methods. Patients eligible were surgical candidates with biopsy-proven, previously untreated head and neck carcinoma and were consented preoperatively for participation in this study. Tissue was collected from 50 patients, and after confirmation of tumor and normal specimens by hematoxylin and eosin (H&E), there were 42 tumor samples and 42 normal adjacent controls. Results. There were 42 confirmed carcinoma specimens on H&E, and CRS microscopy identified 37 as carcinoma. Of the 42 normal specimens, CRS microscopy identified 40 as normal. This resulted in a sensitivity of 88.1% and specificity of 95.2% in distinguishing between neoplastic and nonneoplastic images. Conclusion. CRS microscopy is a unique label-free imaging technique that can provide rapid, high-resolution images and can accurately determine the presence of head and neck carcinoma. This holds potential for implementation into standard practice, allowing frozen margin evaluation even at institutions without a histopathology laboratory.
引用
收藏
页码:448 / 453
页数:6
相关论文
共 19 条
[1]  
[Anonymous], 2015, CANC FACTS FIG 2015
[2]   Discrimination between Oral Cancer and Healthy Tissue Based on Water Content Determined by Raman Spectroscopy [J].
Barroso, E. M. ;
Smits, R. W. H. ;
Schut, T. C. Bakker ;
ten Hove, I. ;
Hardillo, J. A. ;
Wolvius, E. B. ;
de Jong, R. J. Baatenburg ;
Koljenovic, S. ;
Puppels, G. J. .
ANALYTICAL CHEMISTRY, 2015, 87 (04) :2419-2426
[3]   Imaging of human brain tumor tissue by near-infrared laser coherence tomography [J].
Boehringer, H. J. ;
Lankenau, E. ;
Stellmacher, F. ;
Reusche, E. ;
Huettmann, G. ;
Giese, A. .
ACTA NEUROCHIRURGICA, 2009, 151 (05) :507-517
[4]   Investigation of the potential of Raman spectroscopy for oral cancer detection in surgical margins [J].
Cals, Froukje L. J. ;
Schut, Tom C. Bakker ;
Hardillo, Jose A. ;
de Jong, Robert J. Baatenburg ;
Koljenovic, Senada ;
Puppels, Gerwin J. .
LABORATORY INVESTIGATION, 2015, 95 (10) :1186-1196
[5]   The impact of pathologic close margin on the survival of patients with early stage oral squamous cell carcinoma [J].
Chen, Tseng-Cheng ;
Wang, Cheng-Ping ;
Ko, Jenq-Yuh ;
Yang, Tsung-Lin ;
Lou, Pei-Jen .
ORAL ONCOLOGY, 2012, 48 (07) :623-628
[6]   Accuracy, utility, and cost of frozen section margins in head and neck cancer surgery [J].
DiNardo, LJ ;
Lin, J ;
Karageorge, LS ;
Powers, CN .
LARYNGOSCOPE, 2000, 110 (10) :1773-1776
[7]   Refining the utility and role of Frozen section in head and neck squamous cell carcinoma resection [J].
Du, Eugenie ;
Ow, Thomas J. ;
Lo, Yung-Tai ;
Gersten, Adam ;
Schiff, Bradley A. ;
Tassler, Andrew B. ;
Smith, Richard V. .
LARYNGOSCOPE, 2016, 126 (08) :1768-1775
[8]  
Freudiger CW, 2014, NAT PHOTONICS, V8, P153, DOI [10.1038/NPHOTON.2013.360, 10.1038/nphoton.2013.360]
[9]   Multicolored stain-free histopathology with coherent Raman imaging [J].
Freudiger, Christian W. ;
Pfannl, Rolf ;
Orringer, Daniel A. ;
Saar, Brian G. ;
Ji, Minbiao ;
Zeng, Qing ;
Ottoboni, Linda ;
Ying, Wei ;
Waeber, Christian ;
Sims, John R. ;
De Jager, Philip L. ;
Sagher, Oren ;
Philbert, Martin A. ;
Xu, Xiaoyin ;
Kesari, Santosh ;
Xie, X. Sunney ;
Young, Geoffrey S. .
LABORATORY INVESTIGATION, 2012, 92 (10) :1492-1502
[10]   Label-Free Biomedical Imaging with High Sensitivity by Stimulated Raman Scattering Microscopy [J].
Freudiger, Christian W. ;
Min, Wei ;
Saar, Brian G. ;
Lu, Sijia ;
Holtom, Gary R. ;
He, Chengwei ;
Tsai, Jason C. ;
Kang, Jing X. ;
Xie, X. Sunney .
SCIENCE, 2008, 322 (5909) :1857-1861