Carbonized blood deposited on fibres during 810, 940 and 1,470 nm endovenous laser ablation: thickness and absorption by optical coherence tomography

被引:47
作者
Amzayyb, Mustafa [1 ,2 ]
van den Bos, Renate R. [3 ]
Kodach, Vitali M. [1 ]
de Bruin, D. Martijn [1 ]
Nijsten, Tamar [3 ]
Neumann, H. A. Martino [3 ]
van Gemert, Martin J. C. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Biomed Engn & Phys, NL-1105 AZ Amsterdam, Netherlands
[2] Hague Univ Appl Sci, Delft, Netherlands
[3] Erasmus MC, Dept Dermatol, Rotterdam, Netherlands
关键词
Phlebology; Endovenous laser ablation; 810, 940, 1,470 nm; Deposited layer on fibre tip; Transmission; OCT; Layer thickness; Layer absorption; Fibre tip damage; DIODE-LASER; TISSUE; OCCLUSION; MECHANISM; DAMAGE; RANGE; VEINS; MODEL; VIVO;
D O I
10.1007/s10103-009-0749-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Endovenous laser ablation (EVLA) is commonly used to treat saphenous varicosities. Very high temperatures at the laser fibre tip have been reported during EVLA. We hypothesized that the laser irradiation deposits a layer of strongly absorbing carbonized blood of very high temperature on the fibre tip. We sought to prove the existence of these layers and study their properties by optical transmission, optical coherence tomography (OCT) and microscopy. We analysed 23 EVLA fibres, 8 used at 810 nm, 7 at 940 nm and 8 at 1,470 nm. We measured the transmission of these fibres in two wavelength bands (450-950 nm; 950-1,650 nm). We used 1,310 nm OCT to assess the thickness of the layers and the attenuation as a function of depth to determine the absorption coefficient. Microscopy was used to view the tip surface. All fibres showed a slightly increasing transmission with wavelength in the 450-950 nm band, and a virtually wavelength-independent transmission in the 950-1,650 nm band. OCT scans showed a thin layer deposited on all 13 fibres investigated, 6 used at 810 nm, 4 at 940 nm and 3 at 1,470 nm, some with inhomogeneities over the tip area. The average absorption coefficient of the 13 layers was 72 +/- 16 mm(-1). The average layer thickness estimated from the transmission and absorption measurements was 8.0 +/- 2.7 A mu m. From the OCT data, the average maximal thickness was 26 A +/- 6 A mu m. Microscopy of three fibre tips, one for each EVLA wavelength, showed rough, cracked and sometimes seriously damaged tip surfaces. There was no clear correlation between the properties of the layers and the EVLA parameters such as wavelength, except for a positive correlation between layer thickness and total delivered energy. In conclusion, we found strong evidence that all EVLA procedures in blood filled veins deposit a heavily absorbing hot layer of carbonized blood on the fibre tip, with concomitant tip damage. This major EVLA mechanism is unlikely to have much wavelength dependence at similar delivered energies per centimetre of vein. Optical-thermal interaction between the vein wall and the transmitted laser light depends on wavelength.
引用
收藏
页码:439 / 447
页数:9
相关论文
共 18 条
  • [1] REFRACTIVE-INDEX OF SOME MAMMALIAN-TISSUES USING A FIBER OPTIC CLADDING METHOD
    BOLIN, FP
    PREUSS, LE
    TAYLOR, RC
    FERENCE, RJ
    [J]. APPLIED OPTICS, 1989, 28 (12): : 2297 - 2303
  • [2] In vivo three-dimensional imaging of neovascular age-related macular degeneration using optical frequency domain imaging at 1050 nm
    de Bruin, Daniel M.
    Burnes, Daina L.
    Loewenstein, John
    Chen, Yueli
    Chang, Susie
    Chen, Teresa C.
    Esmaili, Daniel D.
    de Boer, Johannes F.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (10) : 4545 - 4552
  • [3] Endovenous laser ablation: an experimental study on the mechanism of action
    Disselhoff, B. C. V. M.
    Rem, A. I.
    Verdaasdonk, R. M.
    Kinderen, D. J. Der
    Moll, F. L.
    [J]. PHLEBOLOGY, 2008, 23 (02) : 69 - 76
  • [4] Endovenous laser ablation: mechanism of action
    Fan, C-M
    Rox-Anderson, R.
    [J]. PHLEBOLOGY, 2008, 23 (05) : 206 - 213
  • [5] Germer CT, 1998, LASER SURG MED, V23, P194, DOI 10.1002/(SICI)1096-9101(1998)23:4<194::AID-LSM2>3.3.CO
  • [6] 2-Y
  • [7] A 3-ZONE MODEL OF SOFT-TISSUE DAMAGE BY A CO2-LASER
    MCKENZIE, AL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (09) : 967 - 983
  • [8] Endovascular optical coherence tomography ex vivo: venous wall anatomy and tissue alterations after endovenous therapy
    Meissner, Oliver A.
    Schmedt, Claus-Georg
    Hunger, Kathrin
    Hetterich, Holger
    Sroka, Ronald
    Rieber, Johannes
    Babaryka, Gregor
    Steckmeier, Bernd Manfred
    Reiser, Maximilian
    Siebert, Uwe
    Mueller-Lisse, Ullrich
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (09) : 2384 - 2393
  • [9] Is a Vein Filled With Blood a Good Model for Studying Endovenous Laser Ablation?
    Mordon, Serge
    Wassmer, Benjamin
    Servell, Pascal
    Desmyttere, Jacques
    Grard, Christophe
    Stalnikiewicz, Guillaume
    [J]. LASERS IN SURGERY AND MEDICINE, 2009, 41 (08) : 543 - 544
  • [10] Mathematical modeling of 980-nm and 1320-nm endovenous laser treatment
    Mordon, Serge R.
    Wassmer, Benjamin
    Zemmouri, Jaouad
    [J]. LASERS IN SURGERY AND MEDICINE, 2007, 39 (03) : 256 - 265