Potentialities of High-Resolution 3-D CZT Drift Strip Detectors for Prompt Gamma-Ray Measurements in BNCT

被引:15
|
作者
Abbene, Leonardo [1 ]
Principato, Fabio [1 ]
Buttacavoli, Antonino [1 ]
Gerardi, Gaetano [1 ]
Bettelli, Manuele [2 ]
Zappettini, Andrea [2 ]
Altieri, Saverio [3 ,4 ]
Auricchio, Natalia [5 ]
Caroli, Ezio [5 ]
Zanettini, Silvia [6 ]
Protti, Nicoletta [3 ,4 ]
机构
[1] Univ Palermo, Dept Phys & Chem DiFC Emilio Segre, Viale Sci,Edificio 18, I-90128 Palermo, Italy
[2] CNR, IMEM, Parco Area Sci 37-A, I-43100 Parma, Italy
[3] Univ Pavia, Dept Phys, Via Agostino Bassi 6, I-27100 Pavia, Italy
[4] Nucl Phys Natl Inst INFN, Unit Pavia, Via Agostino Bassi 6, I-27100 Pavia, Italy
[5] OAS Bologna, INAF, I-40129 Bologna, Italy
[6] Due2lab srl, Via Paolo Borsellino 2, I-42019 Scandiano, Italy
关键词
CZT detectors; X-ray and gamma-ray detectors; BNCT; NEUTRON-CAPTURE THERAPY; THICK CDTE DETECTOR; COMPTON CAMERA; PG-SPECT; SYSTEM; ARRAYS;
D O I
10.3390/s22041502
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Recently, new high-resolution cadmium-zinc-telluride (CZT) drift strip detectors for room temperature gamma-ray spectroscopic imaging were developed by our group. The CZT detectors equipped with orthogonal anode/cathode collecting strips, drift strips and dedicated pulse processing allow a detection area of 6 x 20 mm(2) and excellent room temperature spectroscopic performance (0.82% FWHM at 661.7 keV). In this work, we investigated the potentialities of these detectors for prompt gamma-ray spectroscopy (PGS) in boron neutron capture therapy (BNCT). The detectors, exploiting the measurement of the 478 keV prompt gamma rays emitted by 94% Li-7 nuclides from the B-10(n, alpha)Li-7 reaction, are very appealing for the development of single-photon emission computed tomography (SPECT) systems and Compton cameras in BNCT. High-resolution gamma-ray spectra from B-10 samples under thermal neutrons were measured at the T.R.I.G.A. Mark II research nuclear reactor of the University of Pavia (Italy).
引用
收藏
页数:11
相关论文
共 23 条
  • [1] A monolithic array of silicon drift detectors for high-resolution gamma-ray imaging
    Fiorini, C.
    Longoni, A.
    Perotti, F.
    Labanti, C.
    Rossi, E.
    Lechner, P.
    Soltau, H.
    Strüder, Lothar
    IEEE Transactions on Nuclear Science, 2002, 49 II (03) : 995 - 1000
  • [2] Vertex detection in a stack of Si-drift detectors for high resolution gamma-ray imaging
    Walenta, AH
    Brill, AB
    Castoldi, A
    Çonka-Nurdan, T
    Guazzoni, C
    Hartmann, K
    Longoni, A
    Nurdan, K
    Strüder, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 1434 - 1438
  • [3] Recent advances in the development of high-resolution 3D cadmium-zinc-telluride drift strip detectors
    Abbene, Leonardo
    Gerardi, Gaetano
    Principato, Fabio
    Buttacavoli, Antonino
    Altieri, Saverio
    Protti, Nicoletta
    Tomarchio, Elio
    Del Sordo, Stefano
    Auricchio, Natalia
    Bettelli, Manuele
    Amade, Nicola Sarzi
    Zanettini, Silvia
    Zappettini, Andrea
    Caroli, Ezio
    JOURNAL OF SYNCHROTRON RADIATION, 2020, 27 : 1564 - 1576
  • [4] Gamma-Ray imaging for nuclear security and safety: Towards 3-D gamma-ray vision
    Vetter, Kai
    Barnowksi, Ross
    Haefner, Andrew
    Joshi, Tenzing H. Y.
    Pavlovsky, Ryan
    Quiter, Brian J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 878 : 159 - 168
  • [5] Background radiation reduction for a high-resolution gamma-ray spectrometer used for environmental radioactivity measurements
    Radulescu, I.
    Blebea-Apostu, A. M.
    Margineanu, R. M.
    Mocanu, N.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 715 : 112 - 118
  • [6] Evaluation of the energy resolution of a prompt gamma-ray imaging detector using LaBr3(Ce) scintillator and 8 x 8 array MPPC for an animal study of BNCT
    Okazaki, Keita
    Tanaka, Hiroki
    Takata, Takushi
    Kawabata, Shinji
    Akabori, Kiyotaka
    Sakurai, Yoshinori
    APPLIED RADIATION AND ISOTOPES, 2020, 163
  • [7] Temperature control design for a high-resolution gamma-ray tomography detector
    Bieberle, Andre
    Schleicher, Eckhard
    Hampel, Uwe
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (01)
  • [8] Mapping the Minimum Detectable Activities of Gamma-Ray Sources in a 3-D Scene
    Bandstra, M. S.
    Hellfeld, D.
    Lee, J.
    Quiter, B. J.
    Salathe, M.
    Vavrek, J. R.
    Joshi, T. H. Y.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2023, 70 (01) : 64 - 75
  • [9] Handheld real-time volumetric 3-D gamma-ray imaging
    Haefner, Andrew
    Barnowski, Ross
    Luke, Paul
    Amman, Mark
    Vetter, Kai
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 857 : 42 - 49
  • [10] Large-Area Microcalorimeter Detectors for Ultra-High-Resolution X-Ray and Gamma-Ray Spectroscopy
    Bacrania, M. K.
    Hoover, A. S.
    Karpius, P. J.
    Rabin, M. W.
    Rudy, C. R.
    Vo, D. T.
    Beall, J. A.
    Bennett, D. A.
    Doriese, W. B.
    Hilton, G. C.
    Horansky, R. D.
    Irwin, K. D.
    Jethava, N.
    Sassi, E.
    Ullom, J. N.
    Vale, L. R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (04) : 2299 - 2302