Optical Monitoring During Bioreactor Conditioning of Tissue-Engineered Heart Valves

被引:8
作者
Ziegelmueller, Johannes A. [1 ]
Zaenkert, Eva K. [1 ]
Schams, Rahmin [1 ]
Lackermair, Stephan [1 ]
Schmitz, Christoph [1 ]
Reichart, Bruno [1 ]
Sodian, Ralf [1 ]
机构
[1] Univ Munich, Lab Tissue Engn & Cell Transplantat, Dept Cardiac Surg, D-81377 Munich, Germany
关键词
IN-VITRO; PULSATILE BIOREACTOR; FABRICATION; SCAFFOLDS; CELLS;
D O I
10.1097/MAT.0b013e3181cf3bdd
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Currently, one approach to tissue engineering has been to develop in vitro conditions to fabricate functional cardiovascular structures such as heart valves before final implantation. In vivo conditions are simulated using a bioreactor system that supplies cells with oxygen and culture media while providing mechanical stimulation to promote tissue maturation. In our experiment, we developed a novel combined optical monitoring and conditioning device. The entire system is made of acrylic glass and is completely transparent. The bioreactor is connected to an air-driven respirator pump, and the cell culture medium continuously circulates through a closed-loop system. By adjusting stroke volume, stroke rate, and inspiration/expiration time of the ventilator, the system allows various pulsatile flows and different levels of pressure. Our optical monitoring and conditioning device provides a sterile environment, mechanical stimulation, and optical monitoring for the in vitro maturation of a tissue-engineered heart valve. With the camera module attached, tissue-engineered valves can be observed during the entire in vitro phase. This setting helps to find the optimal dynamic conditions for tissue-engineered heart valves to mature by adjusting flow and pressure conditions to provide physiological opening and closing behavior of the heart valve construct. ASAIO Journal 2010; 56:228-231.
引用
收藏
页码:228 / 231
页数:4
相关论文
共 22 条
[1]   Effects of fluid flow on the in vitro degradation kinetics of biodegradable scaffolds for tissue engineering [J].
Agrawal, CM ;
McKinney, JS ;
Lanctot, D ;
Athanasiou, KA .
BIOMATERIALS, 2000, 21 (23) :2443-2452
[2]   HEART-VALVES - 10 COMMANDMENTS AND STILL COUNTING [J].
HARKEN, DE .
ANNALS OF THORACIC SURGERY, 1989, 48 (03) :S18-S19
[3]  
Hoerstrup SP, 2000, CIRCULATION, V102, P44
[4]   New pulsatile bioreactor for in vitro formation of tissue engineered heart valves [J].
Hoerstrup, SP ;
Sodian, R ;
Sperling, JS ;
Vacanti, JP ;
Mayer, JE .
TISSUE ENGINEERING, 2000, 6 (01) :75-79
[5]   Current Developments and Future Prospects for Heart Valve Replacement Therapy [J].
Kidane, Asmeret G. ;
Burriesci, Gaetano ;
Cornejo, Patricia ;
Dooley, Audrey ;
Sarkar, Sandip ;
Bonhoeffer, Philipp ;
Edirisinghe, Mohan ;
Seifalian, Alexander M. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 88B (01) :290-303
[6]  
L'Heureux N, 1999, SCIENCE, V284, P1621
[7]   Tissue engineering [J].
Langer, R .
MOLECULAR THERAPY, 2000, 1 (01) :12-15
[8]   The role of bioreactors in tissue engineering [J].
Martin, I ;
Wendt, D ;
Heberer, M .
TRENDS IN BIOTECHNOLOGY, 2004, 22 (02) :80-86
[9]  
Mayer J E Jr, 1995, Semin Thorac Cardiovasc Surg, V7, P130
[10]   Tissue-engineered heart valve prostheses: 'state of the heart' [J].
Migneco, Francesco ;
Hollister, Scott J. ;
Birla, Ravi K. .
REGENERATIVE MEDICINE, 2008, 3 (03) :399-419