On-line ion-exchange pretoncentration and determination of traces of platinum by electrothermal atomic absorption spectrometry

被引:22
作者
García, MMG [1 ]
Rojas, FS [1 ]
Ojeda, CB [1 ]
de Torres, AG [1 ]
Pavón, JMC [1 ]
机构
[1] Univ Malaga, Fac Sci, Dept Analyt Chem, Malaga 29071, Spain
关键词
platinum; electrothermal atomic absorption spectrometry; preconcentration; microcolumn; flow injection; vegetation; soil; waters; and catalyst samples;
D O I
10.1007/s00216-002-1679-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method to determine trace amounts of platinum in different samples based on electrothermal atomic absorption spectrometry is described. The preconcentration step is performed on a chelating resin microcolumn [1,5-bis(2-pyridyl)-3-sulfophenyl methylene thiocarbonohydrazide (PSTH) immobilized on an anion-exchange resin (Dowex 1x8-200)] placed in the autosampler arm. The combination of a peristaltic pump for sample loading and the atomic absorption spectrometer pumps for elution through a selection valve simplifies the hardware. The peristaltic pump and the selection valve are easily controlled electronically with two switches placed in the autosampler, which are activated when the autosampler arm is down. Thus, the process is fully automated without any modification of the software of the atomic absorption spectrometer. Under the optimum conditions with a 60-s preconcentration time, a sample flow rate of 2.4 mL min(-1), and an injection volume of eluent of 40 muL, a linear calibration graph was obtained in the range 0-100ngmL(-1). The enrichment factor was 14. The detection limit under these conditions is 1 ng mL(-1), and the relative standard deviation (RSD) is 1.6% for 10 ng mL(-1) of Pt. The method has been applied to the determination of platinum in catalyst, vegetation, soil, and natural water samples. The results showed good agreement with the certified value and the recoveries of Pt added to samples were 98-105%.
引用
收藏
页码:1229 / 1233
页数:5
相关论文
共 27 条
[1]   Analytical methods for the determination of platinum in biological and environmental materials [J].
Balcerzak, M .
ANALYST, 1997, 122 (05) :R67-R74
[2]   DETERMINATION OF PLATINUM IN BIOTIC AND ENVIRONMENTAL-SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AFTER ITS ELECTRODEPOSITION INTO A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON [J].
BEINROHR, E ;
LEE, ML ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9) :689-692
[3]   ONLINE PRECONCENTRATION AND DETERMINATION OF TRACE PLATINUM BY FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
CANTARERO, A ;
GOMEZ, MM ;
CAMARA, C ;
PALACIOS, MA .
ANALYTICA CHIMICA ACTA, 1994, 296 (02) :205-211
[4]   Preconcentration and matrix separation of precious metals in geological and related materials using metalfix-chelamine resin prior to inductively coupled plasma mass spectrometry [J].
Coedo, AG ;
Dorado, MT ;
Padilla, I ;
Alguacil, F .
ANALYTICA CHIMICA ACTA, 1997, 340 (1-3) :31-40
[5]   ONLINE PRECONCENTRATION AND SEPARATION OF PALLADIUM, PLATINUM AND IRIDIUM USING ALPHA-AMINO PYRIDINE RESIN WITH FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
DI, P ;
DAVEY, DE .
TALANTA, 1995, 42 (05) :685-692
[6]   AN EFFICIENT COMBINED PROCEDURE FOR THE EXTREME TRACE ANALYSIS OF GOLD, PLATINUM, PALLADIUM AND RHODIUM WITH THE AID OF GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AND TOTAL-REFLECTION X-RAY-FLUORESCENCE ANALYSIS [J].
ELLER, R ;
ALT, F ;
TOLG, G ;
TOBSCHALL, HJ .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 334 (08) :723-739
[7]   DETERMINATION OF PALLADIUM AND PLATINUM IN FRESH-WATERS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY AND ACTIVATED-CHARCOAL PRECONCENTRATION [J].
HALL, GEM ;
PELCHAT, JC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (08) :1059-1065
[8]  
HARTKY FR, 1991, CHEM PLATINUM GROUP
[9]  
Hidalgo MM, 1996, FRESEN J ANAL CHEM, V354, P420
[10]  
Keppler B.K., 1993, METAL COMPLEXES CANC