Evaluation of Gianturco coils for closure of large (>=3.5 mm) patent ductus arteriosus

被引:44
作者
Owada, CY
Teitel, DF
Moore, P
机构
[1] Division of Pediatric Cardiology, Univ. of California San Francisco, San Francisco, CA
[2] Division of Pediatric Cardiology, Box 0544, Univ. of California at San Francisco, San Francisco, CA 94143
关键词
D O I
10.1016/S0735-1097(97)00382-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. This report evaluates the use of Gianturco coils to close large patent ductus arteriosus (PDAs) (greater than or equal to 3.5 mm) and describes transvenous delivery of 0.052-in. (0.132-cm) Gianturco coils. Background. Coil closure of PDAs has become increasingly popular, However, the technique has significant limitations when used to close large PDAs. This report evaluates patient characteristics, PDA anatomy, hemodynamic variables, delivery technique and coil geometry to determine predictors of success. Methods. Between January 1995 and January 1997, 16 of 118 patients undergoing catheterization for PDA closure were found to have large PDAs. Their median age and weight were 14 months (range 3 months to 43 years) and 8.5 kg (range 3.5 to 73), respectively, The mean PDA diameter was 4.3 mm (range 3.5 to 5.9), Closure of PDAs was attempted using transcatheter delivery of 0.038-in. (0.096-cm) and 0.052-in. coils. Differences in clinical, anatomic, hemodynamic and technical variables between successes and failures were compared. Results. Eleven (69%) of 16 patients had successful closure of their PDA, Failures occurred only in patients <8 months of age with an indexed PDA diameter >7 mm/m and a pulmonary/systemic how ratio greater than or equal to 2.8:1. Use of 0.052-in. coils tended to reduce the incidence of embolization and the number of coils needed for closure, Conclusions. Patients >8 months of age can have successful closure of large PDAs with currently available Gianturco coils, The 0.052-in. Gianturco coils can be used safely to close large PDAs in infants as small as 6 kg. Increased experience and improved coil design may improve closure rates of large PDAs in infants. (C) 1997 by the American College of Cardiology.
引用
收藏
页码:1856 / 1862
页数:7
相关论文
共 16 条
  • [1] [Anonymous], GER MED MON
  • [2] TRANSCATHETER CLOSURE OF A LARGE PATENT DUCTUS-ARTERIOSUS WITH THE CLAMSHELL SEPTAL UMBRELLA
    BRIDGES, ND
    PERRY, SB
    PARNESS, I
    KEANE, JF
    LOCK, JE
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (05) : 1297 - 1302
  • [3] PERCUTANEOUS CLOSURE OF THE SMALL (LESS-THAN-2.5 MM) PATENT DUCTUS-ARTERIOSUS USING COIL EMBOLIZATION
    CAMBIER, PA
    KIRBY, WC
    WORTHAM, DC
    MOORE, JW
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (08) : 815 - 816
  • [4] NEW GIANTURCO-GRIFKA VASCULAR OCCLUSION DEVICE - INITIAL STUDIES IN A CANINE MODEL
    GRIFKA, RG
    MULLINS, CE
    GIANTURCO, C
    NIHILL, MR
    OLAUGHLIN, MP
    SLACK, MC
    CLUBB, FJ
    MYERS, TJ
    [J]. CIRCULATION, 1995, 91 (06) : 1840 - 1846
  • [5] GROWTH OF THE HUMAN HEART RELATIVE TO BODY-SURFACE AREA
    GUTGESELL, HP
    REMBOLD, CM
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (09) : 662 - 668
  • [6] RESULTS OF ANTEROGRADE TRANSCATHETER CLOSURE OF PATENT DUCTUS-ARTERIOSUS USING SINGLE OR MULTIPLE GIANTURCO COILS
    HIJAZI, ZM
    GEGGEL, RL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (09) : 925 - 929
  • [7] ICHIDA F, 1987, BRIT HEART J, V58, P627
  • [8] TRANSCATHETER EMBOLIZATION COIL CLOSURE OF PATENT DUCTUS-ARTERIOSUS - MODIFIED DELIVERY FOR ENHANCED CONTROL DURING COIL POSITIONING
    KUHN, MA
    LATSON, LA
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1995, 36 (03): : 288 - 290
  • [9] TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH GIANTURCO COILS
    LLOYD, TR
    FEDDERLY, R
    MENDELSOHN, AM
    SANDHU, SK
    BEEKMAN, RH
    [J]. CIRCULATION, 1993, 88 (04) : 1412 - 1420
  • [10] PERCUTANEOUS CLOSURE OF THE SMALL PATENT DUCTUS-ARTERIOSUS USING OCCLUDING SPRING COILS
    MOORE, JW
    GEORGE, L
    KIRKPATRICK, SE
    MATHEWSON, JW
    SPICER, RL
    UZARK, K
    ROTHMAN, A
    CAMBIER, PA
    SLACK, MC
    KIRBY, WC
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (03) : 759 - 765