Evaluation of Arterial Spin Labeling MRI-Comparison with 15O-Water PET on an Integrated PET/MR Scanner

被引:8
作者
Fahlstrom, Markus [1 ]
Appel, Lieuwe [1 ]
Kumlien, Eva [2 ]
Danfors, Torsten [1 ]
Engstrom, Mathias [3 ]
Wikstrom, Johan [1 ]
Antoni, Gunnar [4 ]
Larsson, Elna-Marie [1 ]
Lubberink, Mark [1 ]
机构
[1] Uppsala Univ, Dept Surg Sci, S-75185 Uppsala, Sweden
[2] Uppsala Univ, Dept Neurosci, S-75185 Uppsala, Sweden
[3] GE Healthcare, Appl Sci Lab, S-75185 Uppsala, Sweden
[4] Uppsala Univ, Dept Med Chem, S-75185 Uppsala, Sweden
关键词
O-15-water PET; ASL; CBF; PET/MR; validation; CEREBRAL-BLOOD-FLOW; POSITRON-EMISSION-TOMOGRAPHY; TIME ATTENUATION CORRECTION; PERFUSION; CBF; DELAY; HEAD; AGREEMENT; INVERSION; DISEASE;
D O I
10.3390/diagnostics11050821
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cerebral blood flow (CBF) measurements are of high clinical value and can be acquired non-invasively with no radiation exposure using pseudo-continuous arterial spin labeling (ASL). The aim of this study was to evaluate accordance in resting state CBF between ASL (CBFASL) and O-15-water positron emission tomography (PET) (CBFPET) acquired simultaneously on an integrated 3T PET/MR system. The data comprised ASL and dynamic O-15-water PET data with arterial blood sampling of eighteen subjects (eight patients with focal epilepsy and ten healthy controls, age 21 to 61 years). O-15-water PET parametric CBF images were generated using a basis function implementation of the single tissue compartment model. Cortical and subcortical regions were automatically segmented using Freesurfer. Average CBFASL and CBFPET in grey matter were 60 +/- 20 and 75 +/- 22 mL/100 g/min respectively, with a relatively high correlation (r = 0.78, p < 0.001). Bland-Altman analysis revealed poor agreement (bias = -15 mL/100 g/min, lower and upper limits of agreements = -16 and 45 mL/100 g/min, respectively) with a negative relationship. Accounting for the negative relationship, the width of the limits of agreement could be narrowed from 61 mL/100 g/min to 35 mL/100 g/min using regression-based limits of agreements. Although a high correlation between CBFASL and CBFPET was found, the agreement in absolute CBF values was not sufficient for ASL to be used interchangeably with O-15-water PET.
引用
收藏
页数:11
相关论文
共 46 条
[1]   Recommended Implementation of Arterial Spin-Labeled Perfusion MRI for Clinical Applications: A Consensus of the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia [J].
Alsop, David C. ;
Detre, John A. ;
Golay, Xavier ;
Guenther, Matthias ;
Hendrikse, Jeroen ;
Hernandez-Garcia, Luis ;
Lu, Hanzhang ;
MacIntosh, Bradley J. ;
Parkes, Laura M. ;
Smits, Marion ;
van Osch, Matthias J. P. ;
Wang, Danny J. J. ;
Wong, Eric C. ;
Zaharchuk, Greg .
MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (01) :102-116
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Evaluation of basis function and linear least squares methods for generating parametric blood flow images using 15O-water and positron emission tomography [J].
Boellaard, R ;
Knaapen, P ;
Rijbroek, A ;
Luurtsema, GJJ ;
Lammertsma, AA .
MOLECULAR IMAGING AND BIOLOGY, 2005, 7 (04) :273-285
[5]   Arterial spin labeling perfusion MRI at multiple delay times: a correlative study with H215O positron emission tomography in patients with symptomatic carotid artery occlusion [J].
Bokkers, Reinoud P. H. ;
Bremmer, Jochem P. ;
van Berckel, Bart N. M. ;
Lammertsma, Adriaan A. ;
Hendrikse, Jeroen ;
Pluim, Josien P. W. ;
Kappelle, L. Jaap ;
Boellaard, Ronald ;
Klijn, Catharina J. M. .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2010, 30 (01) :222-229
[6]   A general kinetic model for quantitative perfusion imaging with arterial spin labeling [J].
Buxton, RB ;
Frank, LR ;
Wong, EC ;
Siewert, B ;
Warach, S ;
Edelman, RR .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (03) :383-396
[7]   Intersubject variability and reproducibility of 15O PET studies [J].
Coles, JP ;
Fryer, TD ;
Bradley, PG ;
Nortje, J ;
Smielewski, P ;
Rice, K ;
Clark, JC ;
Pickard, JD ;
Menon, DK .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2006, 26 (01) :48-57
[8]   Reduced resolution transit delay prescan for quantitative continuous arterial spin labeling perfusion imaging [J].
Dai, Weiying ;
Robson, Philip M. ;
Shankaranarayanan, Ajit ;
Alsop, David C. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) :1252-1265
[9]   PERFUSION IMAGING [J].
DETRE, JA ;
LEIGH, JS ;
WILLIAMS, DS ;
KORETSKY, AP .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) :37-45
[10]   Long-Delay Arterial Spin Labeling Provides More Accurate Cerebral Blood Flow Measurements in Moyamoya Patients A Simultaneous Positron Emission Tomography/MRI Study [J].
Fan, Audrey P. ;
Guo, Jia ;
Khalighi, Mohammad M. ;
Gulaka, Praveen K. ;
Shen, Bin ;
Park, Jun Hyung ;
Gandhi, Harsh ;
Holley, Dawn ;
Rutledge, Omar ;
Singh, Prachi ;
Haywood, Tom ;
Steinberg, Gary K. ;
Chin, Frederick T. ;
Zaharchuk, Greg .
STROKE, 2017, 48 (09) :2441-2449