Comparison of Image Quality of Shoulder CT Arthrography Conducted Using 120 kVp and 140 kVp Protocols

被引:6
作者
Ahn, Se Jin [1 ]
Hong, Sung Hwan [1 ]
Chai, Jee Won [3 ]
Choi, Ja-Young [1 ]
Yoo, Hye Jin [1 ]
Kim, Sae Hoon [2 ]
Kang, Heung Sik [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul 110744, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Orthoped Surg, Seoul 110744, South Korea
[3] Boramae Med Ctr, Dept Radiol, Seoul 156707, South Korea
关键词
Shoulder CT arthrography; Radiation dose; Tube voltage; Image quality; THIN-SECTION CT; MR ARTHROGRAPHY; GLENOHUMERAL JOINT; CHEST CT; FEASIBILITY; ANGIOGRAPHY; LESIONS; TEARS; ACCURACY;
D O I
10.3348/kjr.2014.15.6.739
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To compare the image quality of shoulder CT arthrography performed using 120 kVp and 140 kVp protocols. Materials and Methods: Fifty-four CT examinations were prospectively included. CT scans were performed on each patient at 120 kVp and 140 kVp; other scanning parameters were kept constant. Image qualities were qualitatively and quantitatively compared with respect to noise, contrast, and diagnostic acceptability. Diagnostic acceptabilities were graded using a one to five scale as follows: 1, suboptimal; 2, below average; 3, acceptable; 4, above average; and 5, superior. Radiation doses were also compared. Results: Contrast was better at 120 kVp, but noise was greater. No significant differences were observed between the 120 kVp and 140 kVp protocols in terms of diagnostic acceptability, signal-to-noise ratio, or contrast-to-noise ratio. Lowering tube voltage from 140 kVp to 120 kVp reduced the radiation dose by 33%. Conclusion: The use of 120 kVp during shoulder CT arthrography reduces radiation dose versus 140 kVp without significant Loss of image quality.
引用
收藏
页码:739 / 745
页数:7
相关论文
共 24 条
[1]   MDCT of the coronary arteries: Feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose [J].
Abada, Hicham T. ;
Larchez, Christophe ;
Daoud, Beatrice ;
Sigal-Cinqualbre, Anne ;
Paul, Jean-Francois .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :S387-S390
[2]   FEASIBILITY OF HIGH-RESOLUTION, LOW-DOSE CHEST CT IN EVALUATING THE PEDIATRIC CHEST [J].
AMBROSINO, MM ;
GENIESER, NB ;
ROCHE, KJ ;
KAUL, A ;
LAWRENCE, RM .
PEDIATRIC RADIOLOGY, 1994, 24 (01) :6-10
[3]   CT arthrography of the glenohumeral joint: CT fluoroscopy versus conventional CT and fluoroscopy - Comparison of image-guidance techniques [J].
Binkert, CA ;
Verdun, FR ;
Zanetti, M ;
Pfirrmann, CW ;
Hodler, J .
RADIOLOGY, 2003, 229 (01) :153-158
[4]   Superior Labral Anterior-to-Posterior Lesions: Comparison of External Rotation and Active Supination CT Arthrography with Neutral CT Arthrography [J].
Choi, Ja-Young ;
Kim, Sae Hoon ;
Yoo, Hye Jin ;
Shin, Seung Han ;
Oh, Joo Han ;
Baek, Goo Hyun ;
Hong, Sung Hwan .
RADIOLOGY, 2012, 263 (01) :199-205
[5]   Optimization of the Contrast Mixture Ratio for Simultaneous Direct MR and CT Arthrography: an in Vitro Study [J].
Choi, Ja-Young ;
Kang, Heung Sik ;
Hong, Sung Hwan ;
Lee, Joon Woo ;
Kim, Na Ra ;
Jun, Woo Sun ;
Moon, Sung Gyu ;
Choi, Jung-Ah .
KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (06) :520-525
[6]   Rotator cuff tears: should abduction and external rotation (ABER) positioning be performed before image acquisition? A CT arthrography study [J].
Cochet, Hubert ;
Couderc, Stephane ;
Pele, Eric ;
Amoretti, Nicolas ;
Moreau-Durieux, Marie-Helene ;
Hauger, Olivier .
EUROPEAN RADIOLOGY, 2010, 20 (05) :1234-1241
[7]   Multidetector computed tomography arthrography of the shoulder: Diagnostic accuracy and indications [J].
De Filippo, M. ;
Bertellini, A. ;
Sverzellati, N. ;
Pogliacomi, F. ;
Costantino, C. ;
Vitale, M. ;
Zappia, M. ;
Corradi, D. ;
Garlaschi, G. ;
Zompatori, M. .
ACTA RADIOLOGICA, 2008, 49 (05) :540-549
[8]   Recurrent Superior Labral Anterior-to-Posterior Tears after Surgery: Detection and Grading with CT Arthrography [J].
De Filippo, Massimo ;
Araoz, Philip A. ;
Pogliacomi, Francesco ;
Castagna, Alex ;
Petriccioli, Dario ;
Sverzellati, Nicola ;
Zompatori, Maurizio .
RADIOLOGY, 2009, 252 (03) :781-788
[9]   Assessment of the long head of the biceps tendon of the shoulder with 3T magnetic resonance arthrography and CT arthrography [J].
De Maeseneer, Michel ;
Boulet, Cedric ;
Pouliart, Nicole ;
Kichouh, Mimoun ;
Buls, Nico ;
Verhelle, Filip ;
De Mey, Johan ;
Shahabpour, Maryam .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (05) :934-939
[10]   Radiation dose reduction by using 100-kV tube voltage in cardiac 64-slice computed tomography: A comparative study [J].
Feuchtner, Gudrun M. ;
Jodocy, Daniel ;
Klauser, Andrea ;
Haberfellner, Bernhard ;
Aglan, Iman ;
Spoeck, Alexander ;
Hiehs, Stefan ;
Soegner, Peter ;
Jaschke, Werner .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 75 (01) :E51-E56