Analysis of dioxins in contaminated soils with the calux and caflux bioassays, an immunoassay, and gas chromatography/high-resolution mass spectrometry

被引:17
作者
Nording, Malin [1 ]
Denison, Michael S.
Baston, David
Persson, Ylva
Spinnel, Erik
Haglund, Peter
机构
[1] Umea Univ, SE-90187 Umea, Sweden
[2] Swedish Def Res Agcy, SE-90182 Umea, Sweden
[3] Univ Calif Davis, Dept Environm Toxicol, Davis, CA 95616 USA
关键词
CALUX; CAFLUX; enzyme-linked immunosorbent assay; soil; dioxin;
D O I
10.1897/06-458R.1
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The chemically activated luciferase expression assay, the chemically activated fluorescence expression assay, and the enzyme-linked immunosorbent assay (ELISA) are all bioanalytical methods that have been used for the detection and quantification of polychlorinated dibenzo-p-dioxins and polychlorinated dibenzofurans (PCDD/Fs). However, no comparisons of the results obtained by these three methods have been published analyzing identical replicates of purified sample extracts. Therefore, we have evaluated the performance of each of these methods for analyzing PCDD/Fs in aliquots of extracts from aged-contaminated soil samples and compared the results with those obtained by gas chromatography/high-resolution mass spectrometry (GC/HRMS). The quantitative performance was assessed and the effects of sample purification and data interpretation on the quality of the bioassay results were investigated. Results from the bioanalytical techniques were, in principle, not significantly different from each other or from the GC/HRMS data (p = 0.05). Further-more, properly used, all of the bioanalytical techniques examined were found to be sufficiently sensitive, selective, and accurate to be used in connection with soil remediation activities when aiming at the remediation goal recommended by the U.S. Environmental Protection Agency (i.e., < 1,000 pg toxic equivalency/g). However, a site-specific correction factor should be applied with the use of the ELISA to account for differences between the toxic equivalency factors and the ELISA cross-reactivities of the various PCDD/F congeners, which otherwise might significantly underestimate the PCDD/F content.
引用
收藏
页码:1122 / 1129
页数:8
相关论文
共 39 条
[1]  
AARTS JMMJG, 1995, EUR J PHARM-ENVIRON, V293, P463
[2]  
[Anonymous], 1998, PERSISTENT ORGANIC P
[3]  
[Anonymous], [No title captured]
[4]   Screening of dioxin-like toxicity equivalents for various matrices with wildtype and recombinant rat hepatoma H4IIE cells [J].
Behnisch, PA ;
Hosoe, K ;
Brouwer, A ;
Sakai, S .
TOXICOLOGICAL SCIENCES, 2002, 69 (01) :125-130
[5]   Combinatorial bio/chemical analysis of dioxin and dioxin-like compounds in waste recycling, feed/food, humans/wildlife and the environment [J].
Behnisch, PA ;
Hosoe, K ;
Sakai, S .
ENVIRONMENT INTERNATIONAL, 2001, 27 (06) :495-519
[6]   Bioanalytical screening methods for dioxins and dioxin-like compounds - a review of bioassay/biomarker technology [J].
Behnisch, PA ;
Hosoe, K ;
Sakai, S .
ENVIRONMENT INTERNATIONAL, 2001, 27 (05) :413-439
[7]   Intra- and interlaboratory calibration of the DR Calux® bioassay for the analysis of dioxins and dioxin-like chemicals in sediments [J].
Besselink, HT ;
Schipper, C ;
Klamer, H ;
Leonards, P ;
Verhaar, H ;
Felzel, E ;
Murk, AJ ;
Thain, J ;
Hosoe, K ;
Schoeters, G ;
Legler, J ;
Brouwer, B .
ENVIRONMENTAL TOXICOLOGY AND CHEMISTRY, 2004, 23 (12) :2781-2789
[8]  
Denison M.S., 1999, MOL BIOL TOXIC RESPO, P393
[9]   The Ah receptor: A regulator of the biochemical and toxicological actions of structurally diverse chemicals [J].
Denison, MS ;
Heath-Pagliuso, S .
BULLETIN OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 1998, 61 (05) :557-568
[10]   Recombinant cell bioassay systems for the detection and relative quantitation of halogenated dioxins and related chemicals [J].
Denison, MS ;
Zhao, B ;
Baston, DS ;
Clark, GC ;
Murata, H ;
Han, D .
TALANTA, 2004, 63 (05) :1123-1133