Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model

被引:26
|
作者
Kamei, Shuzo [1 ]
Osada, Denju [1 ]
Tamai, Kazuya [1 ]
Kato, Nakayuki [1 ]
Takai, Morimitsu [1 ]
Kameda, Masahiro [1 ]
Nohara, Yutaka [1 ]
机构
[1] Dokkyo Med Univ, Sch Med, Dept Orthopaed Surg, Mibu, Tochigi 3210293, Japan
关键词
FRAGMENT-SPECIFIC FIXATION; COMPRESSION PLATE; EXTERNAL FIXATION; DORSAL; SCREW;
D O I
10.1007/s00776-010-1466-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of the present study was to compare the relative stability of five volar locking plates (all of which are available for the treatment of intraarticular fractures of the distal radius) under loading conditions simulating the physiological forces that occur during early active mobilization. Five plating techniques were applied to surgically simulated AO type C3.2 distal radius fractures in formalin-fixed cadavers. The specimens were tested with a servohydraulic materials testing machine with 250 N of axial compression load for 3000 cycles. After cyclic loading, the specimens were loaded until they demonstrated failure in axial compression. The five fixation systems studied included a DRV locking plate (group 1), a Stellar plate (group 2), an Acu-Loc plate (group 3), AO Locking Distal Radius System 2.4 (group 4); and a Matrix SmartLock plate (group 5). None of the plate fixations tested failed during the cyclic loading. Group 2 had a higher elastic limit than groups 4 and 5. There were no significant differences among the five groups for the failure load. Failure occurred at the distal portion of the fixation system, at the ulnar side locking pin, or the locking screw was bent (groups 1, 2, 3); the ulnar side locking screw was broken (groups 3, 4); the locking screw became loose (group 4); and the ulnar side locking screw was uncoupled from the screw hole (group 5). All of the five volar plate fixation systems provided sufficient stability to permit 3000 repeated motions of the digits after surgery for AO type C3 distal radius fractures.
引用
收藏
页码:357 / 364
页数:8
相关论文
共 50 条
  • [21] Biomechanical Stability of Volar Plate Only Versus Addition of Dorsal Ulnar Pin Plate: A Dorsal Ulnar Fragment, C-3-Type, Distal Radius, Cadaver Fracture Model
    Patrick, Nathan C.
    Lewis, Gregory S.
    Roush, Evan P.
    Black, Sarah S.
    Henderson, Stacy R.
    Taylor, Kenneth F.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2020, 34 (09) : E298 - E303
  • [22] Volar versus combined dorsal and volar plate fixation of complex intraarticular distal radius fractures with small dorsoulnar fragment - a biomechanical study
    Hadzhinikolova, Mariya
    Zderic, Ivan
    Ciric, Daniel
    Barcik, Jan P.
    Enchev, Dian
    Baltov, Asen
    Rusimov, Lyubomir
    Varga, Peter
    Stoffel, Karl
    Richards, Geoff
    Gueorguiev, Boyko
    Rashkov, Mihail
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [23] Volar locking plate (VLP) versus non-locking plate (NLP) in the treatment of die-punch fractures of the distal radius, an observational study
    Zhang, Xiong
    Hu, Chunhe
    Yu, Kunlun
    Bai, Jiangbo
    Tian, Dehu
    Xu, Yi
    Zhang, Bing
    INTERNATIONAL JOURNAL OF SURGERY, 2016, 34 : 142 - 147
  • [24] Volar locking plate fixation versus external fixation of distal radius fractures: a meta-analysis
    Gouk, Conor J. C.
    Bindra, Randip R.
    Tarrant, Drew J.
    Thomas, Michael J. E.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2018, 43 (09) : 954 - 960
  • [25] Is volar locking plate superior to external fixation for distal radius fractures? A comprehensive meta-analysis
    Wang, Jie
    Lu, Yandong
    Cui, Yujie
    Wei, Xuelei
    Sun, Jie
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2018, 52 (05) : 334 - 342
  • [26] Internal fixation of dorsally comminuted fractures of the distal part of the radius: a biomechanical analysis of volar plate and intramedullary nail fracture stability
    Konstantinidis, Lukas
    Helwig, Peter
    Seifert, Judith
    Hirschmueller, Anja
    Liodakis, Emmanouil
    Suedkamp, Norbert Paul
    Oberst, Michael
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (11) : 1529 - 1537
  • [27] Biomechanical Comparison of Locking versus Nonlocking Volar and Dorsal T-Plates for Fixation of Dorsally Comminuted Distal Radius Fractures
    Gondusky, Joseph S.
    Carney, Joseph
    Erpenbach, Jonathan
    Robertson, Claire
    Mahar, Andrew
    Oka, Richard
    Thompson, Michael
    Mazurek, Michael
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (01) : 44 - 50
  • [28] Evaluation of a novel, nonspanning external fixator for treatment of unstable extra-articular fractures of the distal radius: Biomechanical comparison with a volar locking plate
    Strauss, Eric J.
    Banerjee, Devraj
    Kummer, Frederick J.
    Tejwani, Nirmal C.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04): : 975 - 981
  • [29] More is not necessarily better. A biomechanical study on distal screw numbers in volar locking distal radius plates
    Drobetz, Herwig
    Weninger, Patrick
    Grant, Caroline
    Heal, Clare
    Muller, Reinhold
    Schuetz, Michael
    Pham, Minh
    Steck, Roland
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2013, 44 (04): : 535 - 539
  • [30] Small volar fragment in the lunate fossa leads to volar tilt loss after volar plate fixation for AO/OTA type-C distal radius fracture
    Seok, Hyun-Gyu
    Park, Wook Tae
    Park, Sung-jin
    Park, Sam-Guk
    HAND SURGERY & REHABILITATION, 2024, 43 (02)