Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model

被引:26
|
作者
Kamei, Shuzo [1 ]
Osada, Denju [1 ]
Tamai, Kazuya [1 ]
Kato, Nakayuki [1 ]
Takai, Morimitsu [1 ]
Kameda, Masahiro [1 ]
Nohara, Yutaka [1 ]
机构
[1] Dokkyo Med Univ, Sch Med, Dept Orthopaed Surg, Mibu, Tochigi 3210293, Japan
关键词
FRAGMENT-SPECIFIC FIXATION; COMPRESSION PLATE; EXTERNAL FIXATION; DORSAL; SCREW;
D O I
10.1007/s00776-010-1466-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of the present study was to compare the relative stability of five volar locking plates (all of which are available for the treatment of intraarticular fractures of the distal radius) under loading conditions simulating the physiological forces that occur during early active mobilization. Five plating techniques were applied to surgically simulated AO type C3.2 distal radius fractures in formalin-fixed cadavers. The specimens were tested with a servohydraulic materials testing machine with 250 N of axial compression load for 3000 cycles. After cyclic loading, the specimens were loaded until they demonstrated failure in axial compression. The five fixation systems studied included a DRV locking plate (group 1), a Stellar plate (group 2), an Acu-Loc plate (group 3), AO Locking Distal Radius System 2.4 (group 4); and a Matrix SmartLock plate (group 5). None of the plate fixations tested failed during the cyclic loading. Group 2 had a higher elastic limit than groups 4 and 5. There were no significant differences among the five groups for the failure load. Failure occurred at the distal portion of the fixation system, at the ulnar side locking pin, or the locking screw was bent (groups 1, 2, 3); the ulnar side locking screw was broken (groups 3, 4); the locking screw became loose (group 4); and the ulnar side locking screw was uncoupled from the screw hole (group 5). All of the five volar plate fixation systems provided sufficient stability to permit 3000 repeated motions of the digits after surgery for AO type C3 distal radius fractures.
引用
收藏
页码:357 / 364
页数:8
相关论文
共 50 条
  • [1] Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression
    Marshall, Tyler
    Momaya, Amit
    Eberhardt, Alan
    Chaudhari, Nilesh
    Hunt, Thomas R., III
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (10): : 2032 - 2038
  • [2] Breakage of volar locking plate system for distal radius fractures
    Kawanishi, Syohei
    Gato, Uasushi
    Nakai, Tsuyoshi
    ACTA ORTHOPAEDICA BELGICA, 2020, 86 (S 1): : 71 - 77
  • [3] Results of type C fractures of distal radius treated by volar locking plate (VLP), an observational comparative study
    Lv, Zhenmu
    Zhao, Yan
    Zhang, Yingze
    Wang, Qiusheng
    Ding, Wenyuan
    Shen, Yong
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (10): : 14721 - 14729
  • [4] Treatment of unstable distal radius fractures with the volar locking plate
    Minegishi, Hanae
    Dohi, Osamu
    An, Soukan
    Sato, Hidetsugu
    UPSALA JOURNAL OF MEDICAL SCIENCES, 2011, 116 (04) : 280 - 284
  • [5] Volar locking plate vs epibloc system for distal radius fractures in the elderly
    Solarino, Giuseppe
    Vicenti, Giovanni
    Abate, Antonella
    Carrozzo, Massimiliano
    Picca, Girolamo
    Colella, Antonio
    Moretti, Biagio
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 : S84 - S90
  • [6] Biomechanical Study on the Cyclic Stability of Distal Radius C2 Type Osteoporotic Fractures Using Locking Plates in a Cadaver Model
    Hsiao, Chih-Kun
    Tu, Yuan-Kun
    Lee, Cheng-Hung
    Yen, Cheng-Yo
    Kao, Feng-Chen
    Tsai, Yi-Jung
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2018, 38 (05) : 707 - 714
  • [7] Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model
    Hoegel, F.
    Mair, S.
    Eberle, S.
    Weninger, P.
    von Oldenburg, G.
    Augat, P.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2013, 133 (01) : 51 - 57
  • [8] Biomechanical Properties of Volar Hybrid and Locked Plate Fixation in Distal Radius Fractures
    Sokol, Shima C.
    Amanatullah, Derek F.
    Curtiss, Shane
    Szabo, Robert M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2011, 36A (04): : 591 - 597
  • [9] Volar locking plate versus external fixation with optional additional K-wire for treatment of AO type C2/C3 fractures: a retrospective comparative study
    Yu, Xiaofei
    Yu, Yadong
    Shao, Xinzhong
    Bai, Yanbin
    Zhou, Tong
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [10] Impact of double-tiered subchondral support procedure with a polyaxial locking plate on the stability of distal radius fractures using fresh cadaveric forearms: Biomechanical and radiographic analyses
    Tsutsui, Sadaaki
    Kawasaki, Keikichi
    Yamakoshi, Ken-ichi
    Uchiyama, Eiichi
    Aoki, Mitsuhiro
    Inagaki, Katsunori
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2016, 21 (05) : 603 - 608