Preparation of urine samples for trace metal determination: a study with aluminium analysis by inductively coupled plasma optical emission spectrometry

被引:20
作者
Burden, TJ [1 ]
Whitehead, MW [1 ]
Thompson, RPH [1 ]
Powell, JJ [1 ]
机构
[1] St Thomas Hosp, Rayne Inst, Gastrointestinal Lab, London SE1 7EH, England
基金
英国惠康基金;
关键词
X-ray microanalysis;
D O I
10.1177/000456329803500208
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Urinary analysis of trace metals forms a significant role in clinical chemistry, but the optimal preparation and analysis of urine samples has not been investigated. Human urine is generally supersaturated with dissolved solids. Therefore, samples often precipitate following collection. X-ray microanalysis showed that this precipitate was predominantly rich in calcium and phosphorus but could include some trace metals from urine, potentially lowering their concentrations in solution. Hence, the precipitate must be fully redissolved for accurate analysis of trace metals in urine. Methods are emphasized for the best collection and preparation of urine samples for subsequent trace metal analysis; in this work inductively coupled plasma optical emission spectrometry (ICPOES) was used for the analysis of aluminium. For optimal accuracy, peak profiles were collected over 396.147 nm-396.157 nm. Urinary aluminium levels were investigated from 10 healthy volunteers and concentrations were obtained using either aqueous, pooled or individual urine-based standard curves. Since urine has a highly variable matrix, individual sample-based standards, which are unique to that particular sample, gave the most accurate results. However, where sample size is small or sample numbers are unfeasibly large, pooled sample-based standards give good approximations to within 15% and, with appropriate validation, other elements as internal standards could also be used for approximations. Aqueous standards should be avoided. Spike-recovery experiments confirmed these data since individual sample based standards showed optimal recovery [99.3 (4.4)%], while pooled sample-based standards were a close proxy [101.6 (9.2)%] but aqueous standards were inappropriate [137.4 (12.8)%]. Postprandial urinary aluminium levels of the 10 volunteers were [7.2 (3.7)mu g/L] after analysis using individual sample-based standard curves.
引用
收藏
页码:245 / 253
页数:9
相关论文
共 21 条
[1]   SILICON AND ALUMINUM AND THEIR INTERRELATIONSHIP IN SERUM AND URINE AFTER RENAL-TRANSPLANTATION [J].
BELLIA, JP ;
NEWTON, K ;
DAVENPORT, A ;
BIRCHALL, JD ;
ROBERTS, NB .
EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 1994, 24 (10) :703-710
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   MATRIX EFFECTS OF POTASSIUM-CHLORIDE AND PHOSPHORIC-ACID IN ARGON INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BUDIC, B ;
HUDNIK, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (01) :53-57
[4]   OPTIMAL ACCURACY, PRECISION AND SENSITIVITY OF INDUCTIVELY-COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY - BIOANALYSIS OF ALUMINUM [J].
BURDEN, TJ ;
POWELL, JJ ;
THOMPSON, RPH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (03) :259-266
[5]  
Burden Trevor, 1994, Analytical Proceedings, V31, P153, DOI 10.1039/ai9943100153
[6]  
COMBER MHI, 1986, ANAL LETT, V19, P2285
[7]   PARTICLES IN PARAFFIN SECTIONS DEMONSTRATED IN THE BACKSCATTERED ELECTRON IMAGE (BEI) [J].
CROCKER, PR ;
TOULSON, E ;
LEVISON, DA .
MICRON, 1982, 13 (04) :437-446
[8]  
Hollis S, 1996, ANN CLIN BIOCHEM, V33, P1
[9]  
LENTNER C, 1981, GEIGY SCI TABLES, V1, P58
[10]   RENAL FILTRATION, REABSORPTION AND EXCRETION OF ALUMINUM IN THE RAT [J].
LOTE, CJ ;
WOOD, JA ;
SAUNDERS, HC .
CLINICAL SCIENCE, 1992, 82 (01) :13-18