The half-life of 18F

被引:16
作者
Garcia-Torano, Eduardo [1 ]
Peyres Medina, Virginia [1 ]
Roteta Ibarra, Miguel [1 ]
机构
[1] CIEMAT, Lab Metrol Radiac Ionizantes, E-28040 Madrid, Spain
关键词
F-18; Half-life; TIME;
D O I
10.1016/j.apradiso.2009.11.052
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The half-life of the positron-emitter F-18 has been measured by following the decay rate with three systems ionization chambers. Ge detectors and coincidence with fast scintillators The decay rate was measured for periods of time up to 9 half-lives The combination of the results obtained with the three measuring systems gives a value of T-1/2 = 1.82871 (18)h. in good agreement with recommended data and with an estimated uncertainty lower than any other previously reported value. (C) 2009 Elsevier Ltd All rights reserved.
引用
收藏
页码:1561 / 1565
页数:5
相关论文
共 12 条
[1]   Precision half-life measurement of 140La with Ge-detector [J].
Adam, J ;
Belov, AG ;
Brandt, R ;
Chaloun, P ;
Honusek, M ;
Kalinnikov, VG ;
Krivopustov, MI ;
Kulakov, BA ;
Langrock, EJ ;
Pronskikh, VS ;
Sosnin, AN ;
Stegailov, VI ;
Tsoupko-Sitnikov, VM ;
Wan, JS ;
Westmeier, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2002, 187 (03) :419-426
[2]   DEAD-TIME CORRECTIONS IN MEASUREMENT OF SHORT-LIVED RADIONUCLIDES [J].
AXTON, EJ ;
RYVES, TB .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1963, 14 (03) :159-&
[3]  
Be M.-M., 2004, Monographie BIPM-5 - Table of Radionuclides, V2
[4]  
CHISTE V, 2006, ATOMIC NUCL DATA REC
[5]  
GUM, 2008, EV MEAS DAT GUID EXP
[6]   DIGITAL SYNTHESIS OF PULSE SHAPES IN REAL-TIME FOR HIGH-RESOLUTION RADIATION SPECTROSCOPY [J].
JORDANOV, VT ;
KNOLL, GF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 345 (02) :337-345
[7]   Accurate γ-ray spectrometry measurements of the half-life of 92Sr [J].
Leconte, P. ;
Hudelot, J. P. ;
Antony, M. .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (10) :1450-1458
[8]   Protocol for uncertainty assessment of half-lives [J].
Pomme, S. ;
Camps, J. ;
Van Ammel, R. ;
Paepen, J. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2008, 276 (02) :335-339
[9]   Update of the BIPM comparison BIPM.RI(II)-K1.F-18 of activity measurements of the radionuclide 18F to include the CIEMAT [J].
Ratel, G. ;
Michotte, C. ;
Garcia-Torano, E. ;
Los Arcos, J-M .
METROLOGIA, 2004, 41
[10]   Half-life measurements with ionization chambers - A study of systematic effects and results [J].
Schrader, H .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (2-4) :317-323