Liquid chromatography-tandem mass spectrometry measurement of 26 steroid hormones in human serum and plasma samples

被引:8
|
作者
Jia, Yongjuan [1 ]
Liu, Xingli [1 ]
Xu, Li [1 ]
Zhao, Jinbao [1 ]
Ni, Junjun [1 ]
Zhang, Yuanyuan [2 ]
机构
[1] Beijing Harmony Hlth Med Diagnost Co Ltd, Beijing 100011, Peoples R China
[2] Beijing AB Sciex Analyt Instrument Trading Co Ltd, Beijing, Peoples R China
关键词
liquid chromatography; tandem mass spectrometry; plasma; serum; steroid hormones; SOLID-PHASE EXTRACTION; LC-MS/MS; DERIVATIZATION; IMMUNOASSAYS; PERFORMANCE; SALIVA; URINE;
D O I
10.1002/jssc.202100091
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The measurement of steroid hormones provided critical information in the clinical evaluation of endocrine disorders. In this study, we developed a high-throughput solid-phase extraction method for the analysis of 26 steroids in human serum and plasma samples by liquid chromatography-tandem mass spectrometry. Chromatographic conditions and sample preparation were optimized to achieve good separation and maximum sensitivity for these analytes. Under the optimum conditions, good linearities were achieved in the quantitative range for each steroid hormone with the correlation coefficients (r) larger than 0.99. The limits of quantitation of the method were in the range from 0.0005 to 0.7901 ng/mL. The recoveries were in the range of 87.2-114.2% with intra- and interday precision lower than 9.94%. This method has already been applied to series of samples from clinical trials, and there was no significant difference between serum and ethylenediaminetetraacetic acid plasma samples.
引用
收藏
页码:2358 / 2370
页数:14
相关论文
共 50 条
  • [1] Analysis of thyroid hormones in serum by liquid chromatography-tandem mass spectrometry
    Wang, Dongli
    Stapleton, Heather M.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 397 (05) : 1831 - 1839
  • [2] Simultaneous determination of endogenous steroid hormones in human and animal plasma and serum by liquid or gas chromatography coupled to tandem mass spectrometry
    Abdel-Khalik, Jonas
    Bjorklund, Erland
    Hansen, Martin
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2013, 928 : 58 - 77
  • [3] Measurement of steroid hormones by liquid chromatography-tandem mass spectrometry with small amounts of hair
    Hobo, Yoshitaka
    Nishikawa, Jun
    Miyashiro, Yoshimichi
    Fujikata, Akira
    STEROIDS, 2020, 164
  • [4] Determination of 19 Steroid Hormones in Human Serum and Urine Using Liquid Chromatography-Tandem Mass Spectrometry
    Li, Zhong-Min
    Kannan, Kurunthachalam
    TOXICS, 2022, 10 (11)
  • [5] Analysis of thyroid hormones in serum by liquid chromatography-tandem mass spectrometry
    Dongli Wang
    Heather M. Stapleton
    Analytical and Bioanalytical Chemistry, 2010, 397 : 1831 - 1839
  • [6] Highly sensitive and selective measurement of underivatized methylmalonic acid in serum and plasma by liquid chromatography-tandem mass spectrometry
    Yuan, Chao
    Gabler, Jessica
    El-Khoury, Joe M.
    Spatholt, Regina
    Wang, Sihe
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 404 (01) : 133 - 140
  • [7] Measurement of tamsulosin in human serum by liquid chromatography-tandem mass spectrometry
    Upreti, Rita
    Homer, Natalie Z. M.
    Naredo, Gregorio
    Cobice, Diego F.
    Hughes, Katherine A.
    Stewart, Laurence H.
    Walker, Brian R.
    Andrew, Ruth
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2013, 930 : 121 - 128
  • [8] Quantification of multiple steroid hormones in serum and human breast cancer tissue by liquid chromatography-tandem mass spectrometry analysis
    Wang, Feng
    Eikeland, Eline
    Reidunsdatter, Randi J.
    Hagen, Lars
    Engstrom, Monica J.
    Geisler, Jurgen
    Haanpaa, Mikko
    Hamalainen, Esa
    Giskeodegard, Guro F.
    Bathen, Tone F.
    FRONTIERS IN ONCOLOGY, 2024, 14
  • [9] Quantification of steroid hormones in human serum by liquid chromatography-high resolution tandem mass spectrometry
    Matysik, Silke
    Liebisch, Gerhard
    JOURNAL OF CHROMATOGRAPHY A, 2017, 1526 : 112 - 118
  • [10] Highly sensitive and selective measurement of underivatized methylmalonic acid in serum and plasma by liquid chromatography-tandem mass spectrometry
    Chao Yuan
    Jessica Gabler
    Joe M. El-Khoury
    Regina Spatholt
    Sihe Wang
    Analytical and Bioanalytical Chemistry, 2012, 404 : 133 - 140