Performance of Two Resin-Containing Blood Culture Media in Detection of Bloodstream Infections and in Direct Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS) Broth Assays for Isolate Identification: Clinical Comparison of the BacT/Alert Plus and Bactec Plus Systems

被引:38
作者
Fiori, Barbara [1 ]
D'Inzeo, Tiziana [1 ]
Di Florio, Viviana [1 ]
De Maio, Flavio [1 ]
De Angelis, Giulia [1 ]
Giaquinto, Alessia [1 ]
Campana, Lara [2 ]
Tanzarella, Eloisa [3 ]
Tumbarello, Mario [2 ]
Antonelli, Massimo [3 ]
Sanguinetti, Maurizio [1 ]
Spanu, Teresa [1 ]
机构
[1] Univ Cattolica Sacro Cuore, Gemelli Hosp, Inst Microbiol, I-00168 Rome, Italy
[2] Univ Cattolica Sacro Cuore, Gemelli Hosp, Inst Infect Dis, I-00168 Rome, Italy
[3] Univ Cattolica Sacro Cuore, Gemelli Hosp, Dept Clin Anesthesiol & Intens Care, I-00168 Rome, Italy
关键词
ANTIMICROBIAL THERAPY; RAPID IDENTIFICATION; MYCOSIS IC/F; FAN BOTTLES; ROUTINE USE; BACTEREMIA; FN; AEROBIC/F; CANDIDA; SENSITIVITY;
D O I
10.1128/JCM.01171-14
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
We compared the clinical performances of the BacT/Alert Plus (bioMerieux) and Bactec Plus (Becton Dickinson) aerobic and anaerobic blood culture (BC) media with adsorbent polymeric beads. Patients >= 16 years old with suspected bloodstream infections (BSIs) were enrolled in intensive care units and infectious disease wards. A single 40-ml blood sample was collected from each and used to inoculate (10 ml/bottle) one set of BacT/Alert Plus cultures and one set of Bactec Plus cultures, each set consisting of one aerobic and one anaerobic bottle. Cultures were incubated <= 5 days in the BacT/Alert 3D and Bactec FX instruments, respectively. A total of 128 unique BSI episodes were identified based on the recovery of clinically significant growth in 212 aerobic cultures (106 BacT/Alert and 106 Bactec) and 151 anaerobic cultures (82 BacT/Alert and 69 Bactec). The BacT/Alert aerobic medium had higher recovery rates for Gram-positive cocci (P = 0.024), whereas the Bactec aerobic medium was superior for recovery of Gram-negative bacilli (P = 0.006). BacT/Alert anaerobic medium recovery rates exceeded those of the Bactec anaerobic medium for total organisms (P = 0.003), Gram-positive cocci (P = 0.013), and Escherichia coli (P = 0.030). In terms of capacity for diagnosing the 128 septic episodes, the BacT/Alert and Bactec sets were comparable, although the former sets diagnosed more BSIs caused by Gram-positive cocci (P = 0.008). They also allowed earlier identification of coagulase-negative staphylococcal growth (mean, 2.8 h; P = 0.003) and growth in samples from patients not on antimicrobial therapy that yielded positive results (mean, 1.3 h; P < 0.001). Similarly high percentages of microorganisms in BacT/Alert and Bactec cultures (93.8% and 93.3%, respectively) were identified by direct matrix-assisted laser desorption ionization-time of flight mass spectrometry assay of BC broths. The BacT/Alert Plus media line appears to be a reliable, timesaving tool for routine detection of BSIs in the population we studied, although further studies are needed to evaluate their performance in other settings.
引用
收藏
页码:3558 / 3567
页数:10
相关论文
共 53 条
[1]   Reemergence of gram-negative health care-associated bloodstream infections [J].
Albrecht, Svenja J. ;
Fishman, Neil O. ;
Kitchen, Jennifer ;
Nachamkin, Irving ;
Bilker, Warren B. ;
Hoegg, Cindy ;
Samel, Carol ;
Barbagallo, Stephanie ;
Arentzen, Judy ;
Lautenbach, Ebbing .
ARCHIVES OF INTERNAL MEDICINE, 2006, 166 (12) :1289-1294
[2]  
[Anonymous], 2011, MANUAL CLIN MICROBIO
[3]   ENHANCED DETECTION OF BACTEREMIA WITH A NEW BACTEC RESIN BLOOD CULTURE-MEDIUM [J].
APPELBAUM, PC ;
BECKWITH, DG ;
DIPERSIO, JR ;
DYKE, JW ;
SALVENTI, JF ;
STONE, LL .
JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 17 (01) :48-51
[4]   Decreased Mortality Associated With Prompt Gram Staining of Blood Cultures [J].
Barenfanger, Joan ;
Graham, Donald R. ;
Kolluri, Lavanya ;
Sangwan, Gaurav ;
Lawhorn, Jerry ;
Drake, Cheryl A. ;
Verhulst, Steven J. ;
Peterson, Ryan ;
Moja, Lauren B. ;
Ertmoed, Matthew M. ;
Moja, Ashley B. ;
Shevlin, Douglas W. ;
Vautrain, Robert ;
Callahan, Charles D. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (06) :870-876
[5]   Is the volume of blood cultured still a significant factor in the diagnosis of bloodstream infections? [J].
Bouza, Emilio ;
Sousa, Dolores ;
Rodriguez-Creixems, Marta ;
Lechuz, Juan Garcia ;
Munoz, Patricia .
JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (09) :2765-2769
[6]   Impact of Matrix-Assisted Laser Desorption Ionization Time-of-Flight Mass Spectrometry on the Clinical Management of Patients With Gram-negative Bacteremia: A Prospective Observational Study [J].
Clerc, Olivier ;
Prod'hom, Guy ;
Vogne, Christelle ;
Bizzini, Alain ;
Calandra, Thierry ;
Greub, Gilbert .
CLINICAL INFECTIOUS DISEASES, 2013, 56 (08) :1101-1107
[7]  
Clinical Laboratory Standards Institute, 2007, M47A CLSI
[8]   Optimal testing parameters for blood cultures [J].
Cockerill, FR ;
Wilson, JW ;
Vetter, EA ;
Goodman, KM ;
Torgerson, CA ;
Harmsen, WS ;
Schleck, CD ;
Ilstrup, DM ;
Washington, JA ;
Wilson, WR .
CLINICAL INFECTIOUS DISEASES, 2004, 38 (12) :1724-1730
[9]   Controlled Clinical Comparison of New Pediatric Medium with Adsorbent Polymeric Beads (PF Plus) versus Charcoal-Containing PF Medium in the BacT/Alert Blood Culture System [J].
Doern, Christopher D. ;
Mirrett, Stanley ;
Halstead, Diane ;
Abid, Joan ;
Okada, Pamela ;
Reller, L. Barth .
JOURNAL OF CLINICAL MICROBIOLOGY, 2014, 52 (06) :1898-1900
[10]   DETECTION OF BACTEREMIA IN PATIENTS RECEIVING ANTI-MICROBIAL THERAPY - AN EVALUATION OF THE ANTI-MICROBIAL REMOVAL DEVICE AND B-16 MEDIUM [J].