The triangular wave test for electrocardiographic devices: A historical perspective

被引:13
作者
Bailey, JJ [1 ]
机构
[1] NIH, Ctr Informat Technol, Bethesda, MD 20892 USA
关键词
baseline wander; ST segment; triangular wave test;
D O I
10.1016/j.jelectrocard.2004.08.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Baseline wander makes interpretation ECG recordings difficult, especially the assessment of ST deviation. Eliminating baseline wander without distorting the ST segment is a problem. The traditional high pass filter with a 0.5 Hz low frequency cutoff effectively suppresses baseline but introduces considerable distortion in the level of the ST segment. This distortion results from phase nonlinearities that occur when frequency content and wave amplitude change abruptly, as occurs where the end of the QRS complex meets the ST segment. Since the 1980s nonlinear, digital filters have been designed that can increase the low frequency Cutoff without the introduction of phase distortion. The triangular wave test, first described in the 1990 AHA Recommendations, is an objective method for measuring the ability to suppress baseline wander without affecting the ST segment. This methodology was adopted in 3 American National Standards (Diagnostic ECG in1991; Ambulatory ECGs in 1998; and Cardiac Monitors in 2002).
引用
收藏
页码:71 / 73
页数:3
相关论文
共 15 条
[1]  
*ASS ADV MED INSTR, 2002, AM NAT STAND CARD MO
[2]  
*ASS ADV MED INSTR, 1998, AM NAT STAND AMB EL
[3]  
*ASS ADV MED INSTR, 1992, AM NAT STAND DIAGN E
[4]   RECOMMENDATIONS FOR STANDARDIZATION AND SPECIFICATIONS IN AUTOMATED ELECTROCARDIOGRAPHY - BANDWIDTH AND DIGITAL SIGNAL-PROCESSING - A REPORT FOR HEALTH-PROFESSIONALS BY AN AD HOC WRITING GROUP OF THE COMMITTEE ON ELECTROCARDIOGRAPHY AND CARDIAC ELECTROPHYSIOLOGY OF THE COUNCIL-ON-CLINICAL-CARDIOLOGY, AMERICAN-HEART-ASSOCIATION [J].
BAILEY, JJ ;
BERSON, AS ;
GARSON, A ;
HORAN, LG ;
MACFARLANE, PW ;
MORTARA, DW ;
ZYWIETZ, C .
CIRCULATION, 1990, 81 (02) :730-739
[5]   ST segment analysis by Holter monitoring: Methodological considerations [J].
Bjerregaard, P ;
El-Shafei, A ;
Kotar, SL ;
Labovitz, AJ .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2003, 8 (03) :200-207
[6]   FREQUENCY-RESPONSE CHARACTERISTICS OF AMBULATORY ECG MONITORING SYSTEMS AND THEIR IMPLICATIONS FOR ST SEGMENT ANALYSIS [J].
BRAGGREMSCHEL, DA ;
ANDERSON, CM ;
WINKLE, RA .
AMERICAN HEART JOURNAL, 1982, 103 (01) :20-31
[7]   AN AUTOMATIC PROCEDURE FOR PREPROCESSING ECG-VCG SIGNALS [J].
CERUTTI, S ;
GATTI, E ;
MASCIADRI, L .
INTERNATIONAL JOURNAL OF BIO-MEDICAL COMPUTING, 1982, 13 (04) :329-342
[8]   A FILTER TO SUPPRESS ECG BASE-LINE WANDER AND PRESERVE ST-SEGMENT ACCURACY IN A REAL-TIME ENVIRONMENT [J].
FRANKEL, RA ;
POTTALA, EW ;
BOWSER, RW ;
BAILEY, JJ .
JOURNAL OF ELECTROCARDIOLOGY, 1991, 24 (04) :315-323
[10]  
PIPBERGER HV, 1975, CIRCULATION, V52, P11