Beam hardening artefacts occur in dental implant scans with the NewTom® cone beam CT but not with the dental 4-row multidetector CT

被引:116
作者
Draenert, F. G.
Coppenrath, E.
Herzog, P.
Muller, S.
Mueller-Lisse, U. G.
机构
[1] Univ Munich, Clin Craniomaxillofacial Surg, Study Grp Dr Draenert, D-81545 Munich, Germany
[2] Univ Munich, Dept Clin Radiol, D-81366 Munich, Germany
关键词
computed tomography (CT); dental implant; beam hardening artefact;
D O I
10.1259/dmfr/32579161
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The aim of this study was to examine beam hardening artefacts of the NewTom((R)) 9000 cone-beam computed tomography (CBCT) device compared with the Philips MX 8000 (4-row multidetector CT (MDCT)). Methods: We modified a SawBone((R)) skull to become a standardized model for our study. The skull was used for scans with the NewTom((R)) 9000 CBCT device and a standard dental multi-detector CT (MDCT) at a comparable reconstruction resolution with a standard Straumann ITI 4.1 mm implant in four implant positions in the maxilla (first permanent premolar in the right maxilla region, second permanent molar in the right maxilla region, first permanent premolar in the left maxilla region and second permanent molar in the left maxilla region). Results were compared with construction data of the dental implant. An image quality assessment of the images from both devices was performed with four experienced physicians and statistically analysed with the two-tailed Wilcoxon test. Results: Scans with the NewTom((R)) 9000 CBCT showed strong beam hardening artefacts in the form of a radiation beam shadow in all reconstructions compared with the MDCT. These imaging artefacts became stronger with greater distance from the centre of the scanned volume. These differences in the imaging quality were proved as significant in a quality evaluation by four experienced physicians (P < 0.05). Conclusions: Visual spatial resolution of the NewTom((R)) 9000 CBCT was less accurate than the Philips((R)) MX 8000 MDCT in the imaging of metallic dental implants.
引用
收藏
页码:198 / 203
页数:6
相关论文
共 32 条
[1]   Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[2]  
Bianchi S., 2001, DENTOMAXILLOFAC RAD, V30, pS28
[3]  
Carano Aldo, 2004, Prog Orthod, V5, P212
[4]   Quantitative analysis of spiral computed tomography for craniofacial clinical applications [J].
Cavalcanti, MGP ;
Vannier, MW .
DENTOMAXILLOFACIAL RADIOLOGY, 1998, 27 (06) :344-350
[5]   CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS [J].
CHO, PS ;
JOHNSON, RH ;
GRIFFIN, TW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1863-1883
[6]   Radiation dose in dental radiology [J].
Cohnen, M ;
Kemper, J ;
Möbes, O ;
Pawelzik, J ;
Mödder, U .
EUROPEAN RADIOLOGY, 2002, 12 (03) :634-637
[7]  
Enciso R, 2003, Orthod Craniofac Res, V6 Suppl 1, P66, DOI 10.1034/j.1600-0544.2003.237.x
[8]  
Erickson Michael, 2003, J Calif Dent Assoc, V31, P843
[9]   Dreidimensionale Darstellungsmöglichkeiten des Mittelgesichts mithilfe der digitalen Volumentomographie anhand einer Kadaverstudie zur winkelstabilen Osteosynthese [J].
S. Flinzberg ;
R. Schmelzle ;
D. Schulze ;
U. Rother ;
M. Heiland .
Mund-, Kiefer- und Gesichtschirurgie, 2003, 7 (5) :289-293
[10]  
Fuhrmann A, 2003, Int J Comput Dent, V6, P129