Determining the imaging quality of intraocular lenses

被引:35
|
作者
Norrby, NES [1 ]
Grossman, LW [1 ]
Geraghty, EP [1 ]
Kreiner, CF [1 ]
Mihori, M [1 ]
Patel, AS [1 ]
Portney, V [1 ]
Silberman, DM [1 ]
机构
[1] Pharmacia & Upjohn Inc, NL-9700 AX Groningen, Netherlands
来源
关键词
D O I
10.1016/S0886-3350(98)80270-8
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To validate the proposed optical requirements of a draft international standard for intraocular lenses (IOLs). Setting: Eight optical testing laboratories in the United States, Germany, Japan, and The Netherlands. Methods: The testing laboratories performed modulation transfer function (MTF) tests on various IOLs using a model eye and visual resolution tests in air. Each laboratory performed duplicate measurements on a set of 43 lenses that was circulated among the testing laboratories. Results: The interlaboratory tests showed that the MTF measurements using a model eye had better repeatability and reproducibility than the more common industry practice of resolution testing in air with parallel light and the United States Air Force three-bar target. However, the two methods correlated well. The commonly applied criterion that an IOL resolve in air at least 60% of the Rayleigh cutoff spatial frequency corresponded to a minimum requirement of 0.43 MTF units at 100 mm(-1) in a model eye. Conclusions: Either criterion may be applied in accordance with a proposed international standard for IOLs. The model eye method can be applied over a broader range of dioptric powers and is relevant for materials that interact with aqueous. Both tests appear to have a greater ability to detect unwanted surface aberrations than resolution testing of IOLs in a water cell using parallel light, a described in the current American National Standards Institute standard.
引用
收藏
页码:703 / 714
页数:12
相关论文
共 50 条
  • [11] Quality of images with toric intraocular lenses
    Tognetto, Daniele
    Perrotta, Alberto Armando
    Bauci, Francesco
    Rinaldi, Silvia
    Antonuccio, Manlio
    Pellegrino, Felice Andrea
    Fenu, Gianfranco
    Stamatelatos, George
    Alpins, Noel
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (03): : 376 - 381
  • [12] Analysis of the optical quality of intraocular lenses
    Tognetto, D
    Sanguinetti, G
    Sirotti, P
    Cecchini, P
    Marcucci, L
    Ballone, E
    Ravalico, G
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (08) : 2682 - 2690
  • [14] Imaging performance of spherical and aspheric intraocular lenses
    Werner, W
    Roth, EH
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1999, 214 (04) : 246 - 250
  • [15] Imaging quality of bifocal piggyback intraocular lens versus ReSTOR and TECNIS multifocal lenses
    Artigas, Jose M.
    Felipe, Adelina
    Diaz-Llopis, Manuel
    Garcia-Delpech, Salvador
    Navea, Amparo
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2010, 20 (01) : 71 - 75
  • [17] Customized eye models for determining optimized intraocular lenses power
    Canovas, Carmen
    Artal, Pablo
    BIOMEDICAL OPTICS EXPRESS, 2011, 2 (06): : 1649 - 1663
  • [18] Experimental characterization of the imaging properties of multifocal intraocular lenses
    Gobbi, PG
    Fasce, F
    Bozza, S
    Brancato, R
    OPHTHALMIC TECHNOLOGIES XIII, 2003, 4951 : 104 - 111
  • [19] Effect of Decentration on the Optical Quality of Two Intraocular Lenses
    Ortiz, Carolina
    Esteve-Taboada, Jose J.
    Belda-Salmeron, Lurdes
    Monsalvez-Romin, Daniel
    Dominguez-Vicent, Alberto
    OPTOMETRY AND VISION SCIENCE, 2016, 93 (12) : 1552 - 1559
  • [20] EVALUATION OF THE DYNAMIC LOADS AND QUALITY OF INTRAOCULAR LENSES.
    Fedorov, S.N.
    Semenov, P.A.
    Anisimov, S.I.
    Yuzhelevskii, Yu.A.
    1600, (19):