Inverse-geometry volumetric CT system with multiple detector arrays for wide field-of-view imaging

被引:19
作者
Mazin, Samuel R. [1 ]
Star-Lack, Josh
Bennett, N. Robert
Pelc, Norbert J.
机构
[1] Stanford Univ, Dept Elect Engn & Radiol, Stanford, CA 94305 USA
[2] NexRay Inc, Los Gatos, CA 95030 USA
[3] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[4] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
[5] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
关键词
volumetric CT; inverse-geometry CT; collimator;
D O I
10.1118/1.2737168
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Current volumetric computed tomography (CT) methods require seconds to acquire a thick volume (> 8 cm) with high resolution. Inverse-geometry CT (IGCT) is a new system geometry under investigation that is anticipated to be able to image a thick volume in a single gantry rotation with isotropic resolution and no cone-beam artifacts. IGCT employs a large array of source spots opposite a smaller detector array. The in-plane field of view (FOV) is primarily determined by the size of the source array, in much the same way that the FOV is determined by the size of the detector array in a conventional CT system. Thus, the size of the source array can be a limitation on the achievable FOV We propose adding additional detector arrays, spaced apart laterally, to increase the in-plane FOV while still using a modestly sized source array. We determine optimal detector placement to maximize the FOV while obtaining relatively uniform sampling. We also demonstrate low wasted radiation of the proposed system through design and simulation of a pre-patient collimator. Reconstructions from simulated projection data show no artifacts when combining the data from the detector arrays. Finally, to demonstrate feasibility of the concept, an anthropomorphic thorax phantom containing a porcine heart was scanned on a prototype table-top system. The reconstructed axial images demonstrate a 45 cm in-plane FOV using a 23 cm source array. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:2133 / 2142
页数:10
相关论文
共 17 条
[1]  
ARENSON JS, 1993, Patent No. 5228069
[2]   COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]  
HELL E, 1998, Patent No. 5841831
[5]   Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[6]   SELECTION OF A CONVOLUTION FUNCTION FOR FOURIER INVERSION USING GRIDDING [J].
JACKSON, JI ;
MEYER, CH ;
NISHIMURA, DG ;
MACOVSKI, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :473-478
[7]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[8]   Dual-slice spiral versus single-slice spiral scanning: Comparison of the physical performance of two computed tomography scanners [J].
Liang, Y ;
Kruger, RA .
MEDICAL PHYSICS, 1996, 23 (02) :205-220
[9]   2D simulations of an inverse-geometry volumetric CT system with multiple detector arrays [J].
Mazin, S ;
Star-Lack, J ;
Schmidt, TG ;
Pelc, N .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :889-897
[10]   Geometry analysis of an inverse-geometry volumetric CT system with multiple detector arrays [J].
Mazin, S ;
Schmidt, TG ;
Solomon, E ;
Fahrig, R ;
Pelc, N .
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 :320-329