Future Prospects: Assessment of Intraoperative Optical Coherence Tomography in Ab Interno Glaucoma Surgery

被引:34
作者
Heindl, Ludwig M. [1 ]
Siebelmann, Sebastian [1 ]
Dietlein, Thomas [1 ]
Huettmann, Gereon [2 ]
Lankenau, Eva [3 ]
Cursiefen, Claus [1 ]
Steven, Philipp [1 ]
机构
[1] Univ Cologne, Dept Ophthalmol, Kerpener Str 62, D-50924 Cologne, Germany
[2] Univ Lubeck, Inst Biomed Opt, Lubeck, Germany
[3] OptoMed Technol GmbH, Lubeck, Germany
关键词
Ab interno glaucoma surgery; intraoperative imaging; intraoperative optical coherence tomography; iOCT; ANTERIOR; MANAGEMENT;
D O I
10.3109/02713683.2014.995311
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Ab interno glaucoma surgery, such as trabecular aspiration or ab interno trabeculotomy, has been introduced as an alternative, non-filtering procedure to lower intraocular pressure in glaucoma patients. The purpose of the present study was to assess the feasibility of intraoperative spectral domain optical coherence tomography (iOCT) for improving ab interno glaucoma surgery. Materials and Methods: Intraoperative optical coherence tomography was performed to visualize the anterior chamber angle, the aspiration canula during trabecular aspiration, and the trabectome device during ab interno trabeculotomy. A commercially available 840-nm iOCT was used in a clinical setting, a 1310-nm spectral-domain OCT in an experimental setting. Results: In the clinical setting using the 840-nm iOCT, visualization of the aspiration canula and the trabectome device was possible, however, with rather limited visibility of the chamber angle and the trabecular meshwork. In the experimental setting, the 1310-nm OCTenabled excellent visualization of the chamber angle and detailed evaluation of the instrument-trabecular meshwork relation. Conclusions: Intraoperative use of optical coherence tomography may be helpful for ab interno glaucoma surgery. Adjustments of the commercially available iOCT, including longer wavelengths and oblique scanning options, are necessary for the clinical use.
引用
收藏
页码:1288 / 1291
页数:4
相关论文
共 11 条
[1]   UBM-guided chamber angle surgery for glaucoma management: an experimental study [J].
Dietlein, TS ;
Engels, BF ;
Jacobi, PC ;
Krieglstein, GK .
EYE, 2003, 17 (03) :340-345
[2]   Intraoperative optical coherence tomography using the RESCAN 700: preliminary results from the DISCOVER study [J].
Ehlers, Justis P. ;
Kaiser, Peter K. ;
Srivastava, Sunil K. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2014, 98 (10) :1329-1332
[3]   Intraoperative 2-dimensional optical coherence tomography as a new tool for anterior segment surgery [J].
Geerling, G ;
Müller, M ;
Winter, C ;
Hoerauf, H ;
Oelckers, S ;
Laqua, H ;
Birngruber, R .
ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (02) :253-257
[4]   Intraoperative Changes in Idiopathic Macular Holes by Spectral-Domain Optical Coherence Tomography [J].
Hayashi, Atsushi ;
Yagou, Takaaki ;
Nakamura, Tomoko ;
Fujita, Kazuya ;
Oka, Miyako ;
Fuchizawa, Chiharu .
CASE REPORTS IN OPHTHALMOLOGY, 2011, 2 (02) :149-154
[5]   Bimanual trabecular aspiration in pseudoexfoliation glaucoma - An alternative in nonfiltering glaucoma surgery [J].
Jacobi, PC ;
Dietlein, TS ;
Krieglstein, GK .
OPHTHALMOLOGY, 1998, 105 (05) :886-894
[6]   Microendoscopic trabecular surgery in glaucoma management [J].
Jacobi, PC ;
Dietlein, TS ;
Krieglstein, GK .
OPHTHALMOLOGY, 1999, 106 (03) :538-544
[7]  
Lankenau EM, 2013, ADV OPTICAL TECHNOLO, V2, P1
[8]   Slit-Lamp-Adapted Fourier-Domain OCT for Anterior and Posterior Segments: Preliminary Results and Comparison to Time-Domain OCT [J].
Mueller, M. ;
Schulz-Wackerbarth, C. ;
Steven, P. ;
Lankenau, E. ;
Bonin, T. ;
Mueller, H. ;
Brueggemann, A. ;
Birngruber, R. ;
Grisanti, S. ;
Huettmann, G. .
CURRENT EYE RESEARCH, 2010, 35 (08) :722-732
[9]   Optimising deep anterior lamellar keratoplasty (DALK) using intraoperative online optical coherence tomography (iOCT) [J].
Steven, Philipp ;
Le Blanc, Carolin ;
Lankenau, Eva ;
Krug, Marc ;
Oelckers, Stefan ;
Heindl, Ludwig M. ;
Gehlsen, Uta ;
Huettmann, Gereon ;
Cursiefen, Claus .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2014, 98 (07) :900-904
[10]   Optimizing Descemet Membrane Endothelial Keratoplasty Using Intraoperative Optical Coherence Tomography [J].
Steven, Philipp ;
Le Blanc, Carolin ;
Velten, Kai ;
Lankenau, Eva ;
Krug, Marc ;
Oelckers, Stefan ;
Heindl, Ludwig M. ;
Gehlsen, Uta ;
Huettmann, Gereon ;
Cursiefen, Claus .
JAMA OPHTHALMOLOGY, 2013, 131 (09) :1135-1142