Evaluation of a cyano stationary phase for the determination of tacrolimus, sirolimus and cyclosporin A in whole blood by high-performance liquid chromatography-tandem mass spectrometry

被引:11
作者
Hatsis, P [1 ]
Volmer, DA [1 ]
机构
[1] Natl Res Council Canada, Inst Marine Biosci, Halifax, NS B3H 3Z1, Canada
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2004年 / 809卷 / 02期
关键词
immunosuppressants; cyano column; negative ions;
D O I
10.1016/j.jchromb.2004.06.034
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The potential of a cyano HPLC column for the analysis of three immunosuppressants is investigated. Tacrolimus, sirolimus and cyclosporin A, were used to probe differences in the retention and efficiency of a cyano column compared to the more widely used C 18 column. The cyano column showed comparable retention for all three compounds, whereas the C 18 column showed stronger retention, especially for cyclosporin A. Furthermore, the efficiencies at 50degreesC were up to 12 times higher on the cyano column. As a result, a baseline separation was achieved in less than three minutes with the cyano column, using an isocratic mobile phase of 52/48 (v/v) acetonitrile/water at 0.45 mL/min. The analysis of immunosuppressant drugs in human whole blood was performed with the cyano column using a selected reaction monitoring (SRM) method for each analyte with negative ion mode electrospray ionization on a triple quadrupole mass spectrometer. Detection limits were 0.05 ng/mL for sirolimus, 0.1 ng/mL for cyclosporin A and 0.2 ng/mL for tacrolimus. Calibration curves were linear over three orders of magnitude. Good agreement was obtained with the actual levels of immunosuppressant drugs in patient samples with an average error of less than 10%. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:287 / 294
页数:8
相关论文
共 39 条
[1]   Ion suppression in mass spectrometry [J].
Annesley, TM .
CLINICAL CHEMISTRY, 2003, 49 (07) :1041-1044
[2]   INVESTIGATION OF THE MECHANISM OF PEAK BROADENING OBSERVED IN THE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF CYCLOSPORINE [J].
BOWERS, LD ;
MATHEWS, SE .
JOURNAL OF CHROMATOGRAPHY, 1985, 333 (01) :231-238
[3]   Pitfalls in monitoring tacrolimus (FK 506) [J].
Braun, F ;
Schutz, E ;
Christians, U ;
Lorf, T ;
Schiffmann, JH ;
Armstrong, VW ;
Schroter, W ;
Sewing, KF ;
Oellerich, M ;
Ringe, B .
THERAPEUTIC DRUG MONITORING, 1997, 19 (06) :628-631
[4]   Automated, fast and sensitive quantification of drugs in blood by liquid chromatography-mass spectrometry with on-line extraction: immunosuppressants [J].
Christians, U ;
Jacobsen, W ;
Serkova, N ;
Benet, LZ ;
Vidal, C ;
Sewing, KF ;
Manns, MP ;
Kirchner, GI .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 748 (01) :41-53
[5]   Matrix effect in bio-analysis of illicit drugs with LC-MS/MS: Influence of ionization type, sample preparation, and biofluid [J].
Dams, R ;
Huestis, MA ;
Lambert, WE ;
Murphy, CM .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (11) :1290-1294
[6]   Liquid chromatography/mass spectrometry for therapeutic drug monitoring of immunosuppressants [J].
Deters, M ;
Kaever, V ;
Kirchner, GI .
ANALYTICA CHIMICA ACTA, 2003, 492 (1-2) :133-145
[7]   Zirconia stationary phases for extreme separations [J].
Dunlap, CJ ;
McNeff, CV ;
Stoll, D ;
Carr, PW .
ANALYTICAL CHEMISTRY, 2001, 73 (21) :598A-607A
[8]   EQUATIONS FOR CALCULATION OF CHROMATOGRAPHIC FIGURES OF MERIT FOR IDEAL AND SKEWED PEAKS [J].
FOLEY, JP ;
DORSEY, JG .
ANALYTICAL CHEMISTRY, 1983, 55 (04) :730-737
[9]  
HOGGE LR, 2000, ANN C MASS SPECTR AL
[10]   Quantitative screening and matrix effect studies of drug discovery compounds in monkey plasma using fast-gradient liquid chromatography/tandem mass spectrometry [J].
Hsieh, YS ;
Chintala, M ;
Mei, H ;
Agans, J ;
Brisson, JM ;
Ng, K ;
Korfmacher, WA .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (24) :2481-2487