High wear resistance of femoral components coated with titanium nitride: a retrieval analysis

被引:19
作者
Fabry, Christian [1 ,2 ]
Zietz, Carmen [1 ]
Baumann, Axel [2 ]
Ehall, Reinhard [3 ]
Bader, Rainer [1 ]
机构
[1] Univ Med Rostock, Dept Orthopaed, Biomech & Implant Technol Res Lab, Doberaner Str 142, D-18057 Rostock, Germany
[2] DOT GmbH, Charles Darwin Ring 1a, D-18059 Rostock, Germany
[3] Graz Ragnitz Private Hosp, Berthold Linder Weg 15, A-8047 Graz, Austria
关键词
Polyethylene; Retrieval analysis; Surface damage; Failure analysis; Ceramized surface; Wear; TiN; Coating; Total knee arthroplasty; TOTAL KNEE ARTHROPLASTY; OXIDIZED ZIRCONIUM; COBALT-CHROMIUM; POLYETHYLENE; PERFORMANCE; DAMAGE;
D O I
10.1007/s00167-017-4578-7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The objective of this study was to evaluate the in vivo wear resistance of cobalt-chromium femoral components coated with titanium nitride (TiN). Our null hypothesis was that the surface damage and the thickness of the TiN coating do not correlate with the time in vivo. Methods Twenty-five TiN-coated bicondylar femoral retrievals with a mean implantation period of 30.7 +/- 11.7 months were subjected to an objective surface damage analysis with a semi-quantitative assessment method. A visual examination of scratches, indentations, notches and coating breakthroughs of the surfaces was performed. The roughness and the coating thickness of the TiN coating were evaluated in the main articulation regions. Results Narrow scratches and indentations in the range of low flexion angles on the retrieval surfaces were the most common modes of damage. There was no evidence of delamination on the articulation surface but rather at the bottom of isolated severe indentations or notches. An analysis of three retrievals revealed a coating breakthrough in the patellofemoral joint region, resulting from patella maltracking and a dislocation. The arithmetical mean roughness of the TiN surface slightly increased with the implantation period. In contrast, the maximum peak height of the roughness profile was reduced at the condyles of the retrieved components in comparison with new, unused surfaces. No significant association between the coating thickness and implantation period was determined. Moreover, the measured values were retained in the range of the initial coating thickness even after several years of in vivo service. Conclusions As was demonstrated by the results of this study, the surface damage to the TiN coating did not deteriorate with the implantation period. The calculated damage scores and the measured coating thickness in particular both confirmed that the TiN coating provides low wear rates. Our findings support the use of wear-resistant TiN-coated components in total knee arthroplasty with the objective of reducing the risk of aseptic loosening. However, in terms of TiN-coated femoral components, particular attention should be paid to a correct patellar tracking in order to avoid wear propagation at the implant.
引用
收藏
页码:2630 / 2639
页数:10
相关论文
共 30 条
[1]  
Abicht C., 2005, THESIS
[2]  
[Anonymous], 2003, 0712200302 DIN EN
[3]   Corrosion Damage and Wear Mechanisms in Long-Term Retrieved CoCr Femoral Components for Total Knee Arthroplasty [J].
Arnholt, Christina M. ;
MacDonald, Daniel W. ;
Malkani, Arthur L. ;
Klein, Gregg R. ;
Rimnac, Clare M. ;
Kurtz, Steven M. ;
Kocagoz, Sevi B. ;
Gilbert, Jeremy L. .
JOURNAL OF ARTHROPLASTY, 2016, 31 (12) :2900-2906
[4]   Performance assessment of femoral knee components made from cobalt-chromium alloy and oxidized zirconium [J].
Brandt, J. -M. ;
Guenther, L. ;
O'Brien, S. ;
Vecherya, A. ;
Turgeon, T. R. ;
Bohm, E. R. .
KNEE, 2013, 20 (06) :388-396
[5]   Posterior Condyle Surface Damage on Retrieved Femoral Knee Components [J].
Burnell, Colin D. C. ;
Brandt, Jan-M. ;
Petrak, Martin J. ;
Bourne, Robert B. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (08) :1460-1467
[6]  
Cooper J. R., 1993, Clinical Materials, V14, P295, DOI 10.1016/0267-6605(93)90016-Z
[7]  
D'Lima D, 2007, THESIS
[8]   Analysis of third body particles generated during total knee arthroplasty: Is metal debris an issue? [J].
De Baets, T. ;
Waelput, W. ;
Bellemans, J. .
KNEE, 2008, 15 (02) :95-97
[9]   Patellofemoral instability after total knee arthroplasty [J].
Eisenhuth, Scott A. ;
Saleh, Khaled J. ;
Cui, Quanjun ;
Clark, Charles R. ;
Brown, Thomas E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (446) :149-160
[10]  
Fabry C., 2015, The Unhappy Total Knee Replacement, P135, DOI [10.1007/978-3-319-08099-4, DOI 10.1007/978-3-319-08099-4]