Effect of heart rate on the diagnostic accuracy of 256-slice computed tomography angiography in the detection of coronary artery stenosis: ROC curve analysis

被引:5
|
作者
Wang, Gang [1 ]
Wu, Yifen [2 ]
Zhang, Zhentao [1 ]
Zheng, Xiaolin [1 ]
Zhang, Yulan [1 ]
Liang, Manqiu [1 ]
Yuan, Huanchu [1 ]
Shen, Haiping [1 ]
Li, Dewei [1 ]
机构
[1] Peoples Hosp Dongguan, Dept Radiol, 3 South Wandao Rd, Dongguan 523000, Guangdong, Peoples R China
[2] Peoples Hosp Dongguan, Dept Oncol, Dongguan 523000, Guangdong, Peoples R China
关键词
tomography; coronary angiography; coronary stenosis disease; receiver operating characteristic curve; MYOCARDIAL-INFARCTION; DISEASE; CT; CALCIFICATION; PATIENT; MSCT; CAG;
D O I
10.3892/etm.2016.3150
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The aim of the present study was to investigate the effect of heart rate (HR) on the diagnostic accuracy of 256-slice computed tomography angiography (CTA) in the detection of coronary artery stenosis. Coronary imaging was performed using a Philips 256-slice spiral CT, and receiver operating characteristic (ROC) curve analysis was conducted to evaluate the diagnostic value of 256-slice CTA in coronary artery stenosis. The HR of the research subjects in the study was within a certain range (39-107 bpm). One hundred patients suspected of coronary heart disease underwent 256-slice CTA examination. The cases were divided into three groups: Low HR (HR < 75 bpm), moderate HR (75 <= HR <90 bpm) and high HR (HR >= 90 bpm). For the three groups, two observers independently assessed the image quality for all coronary segments on a four-point ordinal scale. An image quality of grades 1-3 was considered diagnostic, while grade 4 was non-diagnostic. A total of 97.76% of the images were diagnostic in the low-HR group, 96.86% in the moderate-HR group and 95.80% in the high-HR group. According to the ROC curve analysis, the specificity of CTA in diagnosing coronary artery stenosis was 98.40, 96.00 and 97.60% in the low-, moderate- and high-HR groups, respectively. In conclusion, 256-slice coronary CTA can be used to clearly show the main segments of the coronary artery and to effectively diagnose coronary artery stenosis. Within the range of HRs investigated, HR was found to have no significant effect on the diagnostic accuracy of 256-slice coronary CTA for coronary artery stenosis.
引用
收藏
页码:1937 / 1942
页数:6
相关论文
共 50 条
  • [1] Image Quality of 256-Slice Computed Tomography for Coronary Angiography
    Chao, Shu-Ping
    Leu, Jyh-Gang
    Law, Wai-Yip
    Kuo, Chu-Jen
    Shyu, Kou-Gi
    ACTA CARDIOLOGICA SINICA, 2013, 29 (05) : 444 - 450
  • [2] Diagnostic accuracy of 320-slice computed tomography angiography for detection of coronary artery stenosis: Meta-analysis
    Li, Suhua
    Ni, Qiongqiong
    Wu, Huilan
    Peng, Long
    Dong, Ruimin
    Chen, Lin
    Liu, Jinlai
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (03) : 2699 - 2705
  • [3] Diagnostic accuracy of 320-slice computed-tomography for detection of significant coronary artery stenosis in patients with various heart rates and heart rhythms compared with conventional coronary-angiography
    Uehara, Masae
    Takaoka, Hiroyuki
    Kobayashi, Yoshio
    Funabashi, Nobusada
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 167 (03) : 809 - 815
  • [4] Diagnostic Accuracy of 64-Slice Computed Tomography Coronary Angiography
    Meijboom, W. Bob
    Meijs, Matthijs F. L.
    Schuijf, Joanne D.
    Cramer, Maarten J.
    Mollet, Nico R.
    van Mieghem, Carlos A. G.
    Nieman, Koen
    van Werkhoven, Jacob M.
    Pundziute, Gabija
    Weustink, Annick C.
    de Vos, Alexander M.
    Pugliese, Francesca
    Rensing, Benno
    Jukema, J. Wouter
    Bax, Jeroen J.
    Prokop, Mathias
    Doevendans, Pieter A.
    Hunink, Myriam G. M.
    Krestin, Gabriel P.
    de Feyter, Pim J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (25) : 2135 - 2144
  • [5] Detection of significant coronary artery stenosis with 64-slice computed tomography in heart transplant recipients: a comparative study with conventional coronary angiography
    von Ziegler, Franz
    Leber, Alexander W.
    Becker, Alexander
    Kaczmarek, Ingo
    Schoenermarck, Ulf
    Raps, Christine
    Tittus, Janine
    Ueberfuhr, Peter
    Becker, Christoph R.
    Reiser, Maximilian
    Steinbeck, Gerhard
    Knez, Andreas
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2009, 25 (01) : 91 - 100
  • [6] Diagnostic performance of 64-slice computed tomography coronary angiography to detect significant coronary artery stenosis
    Bayrak, Fatih
    Guneysu, Tahsin
    Gemici, Gokmen
    Sevinc, Deniz
    Mutlu, Bulent
    Aytaclar, Semih
    Degertekin, Muzaffer
    ACTA CARDIOLOGICA, 2008, 63 (01) : 11 - 17
  • [7] Diagnostic accuracy of 64-slice computed tomography coronary angiography for the detection of in-stent restenosis: A meta-analysis
    Carrabba, Nazario
    Schuijf, Joanne D.
    de Graaf, Fleur R.
    Parodi, Guido
    Maffei, Erica
    Valenti, Renato
    Palumbo, Alessandro
    Weustink, Annick C.
    Mollet, Nico R.
    Accetta, Gabriele
    Cademartiri, Filippo
    Antoniucci, David
    Bax, Jeroen J.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (03) : 470 - 478
  • [8] Role of 320-slice multislice computed tomography coronary angiography in the assessment of coronary artery stenosis
    Youssef, Mohamed Ahmed
    Dawoud, Mahmoud Abdelaziz
    Elbarbary, Aly Aly
    Elbedewy, Mohamed Mohamed
    Elkhateeb, Hasan Mohamed
    EGYPTIAN JOURNAL OF RADIOLOGY AND NUCLEAR MEDICINE, 2014, 45 (02) : 317 - 324
  • [9] Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease
    Hou, Yang
    Ma, Yue
    Fan, Weipeng
    Wang, Yuke
    Yu, Mei
    Vembar, Mani
    Guo, Qiyong
    EUROPEAN RADIOLOGY, 2014, 24 (01) : 3 - 11
  • [10] Diagnostic accuracy of low-dose 256-slice multi-detector coronary CT angiography using iterative reconstruction in patients with suspected coronary artery disease
    Yang Hou
    Yue Ma
    Weipeng Fan
    Yuke Wang
    Mei Yu
    Mani Vembar
    Qiyong Guo
    European Radiology, 2014, 24 : 3 - 11