Coronary Flow Assessment Using 3-Dimensional Ultrafast Ultrasound Localization Microscopy

被引:46
作者
Demeulenaere, Oscar [1 ]
Sandoval, Zulma [1 ]
Mateo, Philippe [1 ]
Dizeux, Alexandre [1 ]
Villemain, Olivier [1 ]
Gallet, Romain [2 ]
Ghaleh, Bijan [2 ]
Deffieux, Thomas [1 ]
Demene, Charlie [1 ]
Tanter, Mickael [1 ]
Papadacci, Clement [1 ]
Pernot, Mathieu [1 ]
机构
[1] PSL Univ, Phys Med, Ecole Super Phys Chim Ind Paris, Inst Natl Sante & Rech Med U1273,CNRS UMR 8063, Paris, France
[2] Univ Paris Est Creteil, Ecole Natl Vet Alfort, Equipe 03, Inserm U955 IMRB, Maisons Alfort, France
基金
欧洲研究理事会;
关键词
blood flow; coronary; imaging; microcirculation; microscopy; ultrasound; volumetric imaging; DOPPLER;
D O I
10.1016/j.jcmg.2022.02.008
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Direct assessment of the coronary microcirculation has long been hampered by the limited spatial and temporal resolutions of cardiac imaging modalities. OBJECTIVES The purpose of this study was to demonstrate 3-dimensional (3D) coronary ultrasound localization microscopy (CorULM) of the whole heart beyond the acoustic diffraction limit (<20 mm resolution) at ultrafast frame rate (>1000 images/s). METHODS CorULM was performed in isolated beating rat hearts (N = 6) with ultrasound contrast agents (Sonovue, Bracco), using an ultrasonic matrix transducer connected to a high channel-count ultrafast electronics. We assessed the 3D coronary microvascular anatomy, flow velocity, and flow rate of beating hearts under normal conditions, during vasodilator adenosine infusion, and during coronary occlusion. The coronary vasculature was compared with micro-computed tomography performed on the fixed heart. In vivo transthoracic CorULM was eventually assessed on anaesthetized rats (N = 3). RESULTS CorULM enables the 3D visualization of the coronary vasculature in beating hearts at a scale down to microvascular structures (<20 mm resolution). Absolute flow velocity estimates range from 10 mm/s in tiny arterioles up to more than 300 mm/s in large arteries. Fitting to a power law, the flow rate-radius relationship provides an exponent of 2.61 (r(2) = 0.96; P < 0.001), which is consistent with theoretical predictions and experimental validations of scaling laws in vascular trees. A 2-fold increase of the microvascular coronary flow rate is found in response to adenosine, which is in good agreement with the overall perfusion flow rate measured in the aorta (control measurement) that increased from 8.80 imaging 1.03 mL/min to 16.54 imaging 2.35 mL/min (P < 0.001). The feasibility of CorULM was demonstrated in vivo for N = 3 rats. CONCLUSIONS CorULM provides unprecedented insights into the anatomy and function of coronary arteries at the microvasculature level in beating hearts. This new technology is highly translational and has the potential to become a major tool for the clinical investigation of the coronary microcirculation. (C) 2022 by the American College of Cardiology Foundation.
引用
收藏
页码:1193 / 1208
页数:16
相关论文
共 22 条
  • [1] Decreased Expression and Activity of cAMP Phosphodiesterases in Cardiac Hypertrophy and Its Impact on β-Adrenergic cAMP Signals
    Abi-Gerges, Aniella
    Richter, Wito
    Lefebvre, Florence
    Mateo, Philippe
    Varin, Audrey
    Heymes, Christophe
    Samuel, Jane-Lise
    Lugnier, Claire
    Conti, Marco
    Fischmeister, Rodolphe
    Vandecasteele, Gregoire
    [J]. CIRCULATION RESEARCH, 2009, 105 (08) : 784 - U161
  • [2] Coronary microvascular dysfunction: mechanisms and functional assessment
    Camici, Paolo G.
    d'Amati, Giulia
    Rimoldi, Ornella
    [J]. NATURE REVIEWS CARDIOLOGY, 2015, 12 (01) : 48 - 62
  • [3] PHASIC CORONARY BLOOD-FLOW VELOCITY IN INTRAMURAL AND EPICARDIAL CORONARY-ARTERIES
    CHILIAN, WM
    MARCUS, ML
    [J]. CIRCULATION RESEARCH, 1982, 50 (06) : 775 - 781
  • [4] Dynamic Myocardial Ultrasound Localization Angiography
    Cormier, Philippe
    Poree, Jonathan
    Bourquin, Chloe
    Provost, Jean
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (12) : 3379 - 3388
  • [5] Quantitative imaging of coronary flows using 3D ultrafast Doppler coronary angiography
    Correia, M.
    Maresca, D.
    Goudot, G.
    Villemain, O.
    Bize, A.
    Sambin, L.
    Tanter, M.
    Ghaleh, B.
    Pernot, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (10)
  • [6] Ultrasound Localization Microscopy and Super-Resolution: A State of the Art
    Couture, Olivier
    Hingot, Vincent
    Heiles, Baptiste
    Muleki-Seya, Pauline
    Tanter, Mickael
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (08) : 1304 - 1320
  • [7] Spatiotemporal Clutter Filtering of Ultrafast Ultrasound Data Highly Increases Doppler and fUltrasound Sensitivity
    Demene, Charlie
    Deffieux, Thomas
    Pernot, Mathieu
    Osmanski, Bruno-Felix
    Biran, Valerie
    Gennisson, Jean-Luc
    Sieu, Lim-Anna
    Bergel, Antoine
    Franqui, Stephanie
    Correas, Jean-Michel
    Cohen, Ivan
    Baud, Olivier
    Tanter, Mickael
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2015, 34 (11) : 2271 - 2285
  • [8] Ultrafast ultrasound localization microscopy for deep super-resolution vascular imaging
    Errico, Claudia
    Pierre, Juliette
    Pezet, Sophie
    Desailly, Yann
    Lenkei, Zsolt
    Couture, Olivier
    Tanter, Mickael
    [J]. NATURE, 2015, 527 (7579) : 499 - +
  • [9] 2014 ACC/AHA/AATS/PCNA/SCAI/STS Focused Update of the Guideline for the Diagnosis and Management of Patients With Stable Ischemic Heart Disease
    Fihn, Stephan D.
    Blankenship, James C.
    Naidu, Srihari S.
    Ohman, E. Magnus
    Smith, Peter K.
    Alexander, Karen P.
    Bittl, John A.
    Byrne, John G.
    Fletcher, Barbara J.
    Fonarow, Gregg C.
    Lange, Richard A.
    Levine, Glenn N.
    Maddox, Thomas M.
    Anderson, Jeffrey L.
    Halperin, Jonathan L.
    Albert, Nancy M.
    Bozkurt, Biykem
    Brindis, Ralph G.
    Curtis, Lesley H.
    DeMets, David
    Guyton, Robert A.
    Hochman, Judith S.
    Kovacs, Richard J.
    Pressler, Susan J.
    Sellke, Frank W.
    Shen, Win-Kuang
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 64 (18) : 1929 - 1949
  • [10] Anatomic Versus Physiologic Assessment of Coronary Artery Disease
    Gould, K. Lance
    Johnson, Nils P.
    Bateman, Timothy M.
    Beanlands, Rob S.
    Bengel, Frank M.
    Bober, Robert
    Camici, Paolo G.
    Cerqueira, Manuel D.
    Chow, Benjamin J. W.
    Di Carli, Marcelo F.
    Dorbala, Sharmila
    Gewirtz, Henry
    Gropler, Robert J.
    Kaufmann, Philipp A.
    Knaapen, Paul
    Knuuti, Juhani
    Merhige, Michael E.
    Rentrop, K. Peter
    Ruddy, Terrence D.
    Schelbert, Heinrich R.
    Schindler, Thomas H.
    Schwaiger, Markus
    Sdringola, Stefano
    Vitarello, John
    Williams, Kim A.
    Gordon, Donald
    Dilsizian, Vasken
    Narula, Jagat
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : 1639 - 1653