Enantioselective determination of azelnidipine in human plasma using liquid chromatography-tandem mass spectrometry

被引:16
作者
Kawabata, Kiyoshi
Samata, Naozumi
Urasaki, Yoko
Fukazawa, Ichiro
Uchida, Naoki
Uchida, Eiji
Yasuhara, Hajime
机构
[1] Sankyo Co Ltd, Drug Metab & Pharmacokinet Res Labs, Shinagawa Ku, Tokyo 1408710, Japan
[2] Showa Univ, Sch Med, Dept Pharmacol 2, Tokyo 104, Japan
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2007年 / 852卷 / 1-2期
关键词
azelnidipine; enantiomer separation; quantitative analysis; LC/APCI-MS/MS;
D O I
10.1016/j.jchromb.2007.01.050
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive and simple method was developed for determination of the enantiomers of azeinidipine, (R)-(-)-azelnidipine and (S)-(+)-azelnidipine, in human plasma using chiral liquid chromatography with positive ion atmospheric pressure chemical ionization tandem mass spectrometry. Plasma samples spiked with stable isotope-labeled azelnidipine, [H-2(6)]-azelnidipine, as an internal standard, were processed for analysis using a solid-phase extraction in a 96-well plate format. The azelnidipine enantiomers were separated on a chiral column containing alpha(1)-acid glycoprotein as a chiral selector under isocratic mobile phase conditions. Acquisition of mass spectrometric data was performed in multiple reaction monitoring mode, monitoring the transitions from m/z 583 -> 167 for (R)-(-)-azelnidipine and (S)-(+)-azelnidipine, and from m/z 589 -> 167 for [H-2(6)]-azelnidipine. The standard curve was linear over the studied range (0.05-20 ng/mL), with r(2) > 0.997 using weighted (1/x(2)) quadratic regression, and the chromatographic run time was 5.0 min/injection. The intra- and inter-assay precision (coefficient of variation), calculated from the assay data of the quality control samples, was 1.2-8.2% and 2.4-5.8% for (R)-(-)-azelnidipine and (S)-(+)-azelnidipine, respectively. The accuracy was 101.2-117.0% for (R)-(-)-azelnidipine and 100.0-107.0% for (S)-(+)-azelnidipine. The overall recoveries for (R)-(-)-azelnidipine and (S)-(+)azelnidipine were 71.4-79.7% and 71.7-84.2%, respectively. The lower limit of quantification for both enantiomers was 0.05 ng/mL using 1.0 mL of plasma. All the analytes showed acceptable short-term, long-term, auto-sampler and stock solution stability. Furthermore, the method described above was used to separately measure the concentrations of the azelnidipine enantionters in plasma samples collected from healthy subjects who had received a single oral dose of 16 mg of azelnidipine. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:389 / 397
页数:9
相关论文
共 17 条
[1]  
[Anonymous], GUID IND BIOAN METH
[2]   SIMULTANEOUS ASSESSMENT OF THE INTRAVENOUS AND ORAL DISPOSITION OF THE ENANTIOMERS OF RACEMIC NIMODIPINE BY CHIRAL STATIONARY-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND GAS-CHROMATOGRAPHY MASS-SPECTROSCOPY COMBINED WITH A STABLE ISOTOPE TECHNIQUE [J].
FISCHER, C ;
SCHONBERGER, F ;
MUCK, W ;
HEUCK, K ;
EICHELBAUM, M .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1993, 82 (03) :244-250
[3]   DETERMINATION OF THE ENANTIOMERS OF NISOLDIPINE IN HUMAN PLASMA USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON A CHIRAL STATIONARY-PHASE AND GAS-CHROMATOGRAPHY WITH MASS-SELECTIVE DETECTION [J].
HEINIG, R ;
MUSCHALEK, V ;
AHR, G .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 655 (02) :286-292
[4]   Enantio selective determination of lercanidipine in human plasma for pharmacokinetic studies by normal-phase liquid chromatography-tandem mass spectrometry [J].
Jabor, VAP ;
Coelho, EB ;
Ifa, DR ;
Bonato, PS ;
dos Santos, NAG ;
Lanchote, VL .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 796 (02) :429-437
[5]   Coupled-column chromatography on a Chiral-AGP phase for determination of amlodipine enantiomers in human plasma: An HPLC assay with electrochemical detection [J].
Josefsson, M ;
Norlander, B .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 15 (02) :267-277
[6]   Analysis of benidipine enantiomers in human plasma by liquid chromatography-mass spectrometry using a macrocyclic antibiotic (Vancomycin) chiral stationary phase column [J].
Kang, W ;
Lee, DJ ;
Liu, KH ;
Sunwoo, YE ;
Kwon, K ;
Cha, IJ ;
Shin, JG .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 814 (01) :75-81
[7]   Simultaneous determination of azelnidipine and two metabolites in human plasma using liquid chromatography-tandem mass spectrometry [J].
Kawabata, Kiyoshi ;
Urasaki, Yoko .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 844 (01) :45-52
[8]   Method development and validation for quantitative determination of methadone enantiomers in human plasma by liquid chromatography/tandem mass spectrometry [J].
Liang, HR ;
Foltz, RL ;
Meng, M ;
Bennett, P .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 806 (02) :191-198
[9]   Enantioselective determination of felodipine in human plasma by chiral normal-phase liquid chromatography and electrospray ionisation mass spectrometry [J].
Lindmark, B ;
Ahnoff, M ;
Persson, BA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 27 (3-4) :489-495
[10]   Enantioselective assay of nisoldipine in human plasma by chiral high-performance liquid chromatography combined with gas chromatographic-mass spectrometry: applications to pharmacokinetics [J].
Marques, MP ;
Santos, NAG ;
Coelho, EB ;
Bonato, PS ;
Lanchote, VL .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2001, 762 (01) :87-95